Outputs |
The channel outputs are galvanically isolated and not connected to ground. |
Number of output channels |
1 |
Total output power |
max. 60 W |
Output voltage |
-20 V to +20 V |
Maximum output current per channel |
≤ 6 V output voltage |
8 A |
> 6 output voltage |
3 A |
Adjustable output impedance |
not available |
Voltage ripple and noise |
20 Hz to 20 MHz |
< 500 µV (RMS),
< 2 mV (peek-to-peek)(meas.) |
Current ripple and noise |
20 Hz to 20 MHz |
<1 mA (RMS) (meas.) |
Electronic load |
Sink voltage range |
- 20 V to + 20 V |
Maximum sink power |
60 W |
Maximum sink current |
Voltage: ≤ 6 V |
8 A |
Voltage: > 6 V |
3 A |
Sink modes |
constant voltage, constant current |
Load regulation in voltage priority mode |
load change: 10% to 90% |
Voltage |
±(% of output + offset) |
< 0.01% + 0.5 mV |
Voltage load recovery time in standard mode |
regulation to within a band of ±20 mV of the set voltage |
< 30 µs (meas.) |
Voltage load recovery time in high-capacity mode |
regulation to within a band of ±20 mV of the set voltage |
10 µF to 50 µF (mode low): < 30 µs (meas.)
> 50 µF to 470 µF (mode high): < 100 µs (meas.) |
Load regulation in current priority mode |
load change: 10% to 90% |
Current |
±(% of output + offset) |
< 0.01% + 0.1 mA |
Current load recovery time |
regulation to within a band of ±20 mA of the set current |
< 50 µs (meas.) |
Rise time |
10% to 90% of rated output voltage, resistive load |
full load: < 100 µs
no load: < 100 µs |
Fall time |
90% to 10% of rated output voltage, resistive load |
full load: < 100 µs
no load: < 100 µs |
Minimum pulse width |
100 µs |
Programming resolution |
Voltage |
20 V range: 200 μV
6 V range: 50 μV |
Current |
8 A range: 50 μA
3 A range: 25 μA
100 mA range: 1 μA
10 mA range: 100 nA |
Programming accuracy |
Voltage |
±(% of setting + offset) |
20 V range: < 0.02% + 2 mV
6 V range: < 0.02% + 1 mV |
Current |
±(% of setting + offset) |
8 A range: < 0.05% + 2 mA
3 A range: < 0.025% + 500 μA
100 mA range: < 0.025% + 25 μA
10 mA range: < 0.025% + 10 μA |