NGE100 Power Supply Series Specifications |
Number of output channels |
R&S NGE102 |
2 |
R&S NGE103 |
3 |
Total output power |
R&S NGE102 |
66 W |
R&S NGE103 |
100 W |
Maximum output power |
per channel |
33.6W |
Output voltage |
per channel |
0 V to 32 V |
Output current |
per channel |
0 A to 3 A |
Line and load regulation |
Constant voltage mode |
± (% of output + offset) |
< 0.1 % + 20 mV |
Constant current mode |
± (% of output + offset) |
< 0.1% + 5 mA |
Voltage ripple |
20 Hz to 20 MHz; at V = 16 V, I = I max/2 |
typ. < 1.5 mV (RMS) |
typ. < 20 mV (V pp) |
Current ripple |
20 Hz to 20 MHz; at V = 16 V, I = I max/2 |
typ. < 2 mA (RMS) |
Recovery time |
10 % to 90 % load change within a band of ± 20 mV of set voltage |
200 μs |
Setting accuracy (+23 °C -3 °C/+7 °C) |
Voltage |
± (% of output + offset) |
< 0.1% + 30 mV |
Current |
± (% of output + offset) |
< 0.1% + 5 mA |
Readback accuracy (+23 °C -3 °C/+7 °C) |
Voltage |
± (% of output + offset) |
< 0.1% + 20 mV |
Current |
± (% of output + offset) |
< 0.1% + 5 mA |
Resolution |
Voltage |
|
10 mV |
Current |
|
1 mA |
Maximum voltage to earth |
|
150 V DC |
Maximum reverse voltage |
|
33 V |
Maximum inverse voltage |
|
0.4 V |
Maximum permitted current in case of inverse voltage |
|
3 A |
Temperature coefficient (per °C) |
± (% of output + offset) |
voltage: < 0.02% + 5 mV |
current: < 0.02% + 3 mA |
Output voltage overshoot when a channel is active and the mains power is turned off |
|
< 100 mV |
Command processing time |
|
≤ 30 ms |
Programming time (within 1% of final value) |
Positive voltage change |
no load |
10 ms + command processing time |
full load (resistive) |
10 ms + command processing time |
Negative voltage change |
no load |
500 ms + command processing time |
full load (resistive) |
10 ms + command processing time |
Overvoltage protection |
|
adjustable for each channel |
Overpower protection |
|
adjustable for each channel |
Electronic fuse |
|
adjustable for each channel |
Response time |
(Iload > Iresp x 2) |
< 10 ms |
Response time of linked channels |
|
< 100 μs + response time of linked channel |
Fuse delay time |
adjustable for each channel |
10 ms to 10 s (10 ms steps) |
Over-temperature protection |
|
yes |
Energy meter |
|
yes |
Output ramp function (EasyRamp) |
|
yes |
EasyRamp time |
|
10 ms to 10 s |
Trigger input |
Minimum trigger interval |
|
10 ms |
Trigger response time |
|
< 100 ms |
Trigger level |
|
TTL |
Edge direction |
|
rising, falling |
Arbitrary function (EasyArb) |
Parameter |
|
voltage, current, time |
Maximum number of points |
|
128 |
Dwell time |
|
10 ms to 600 s |
Repetition mode |
|
continuous or burst mode with 1 to 255 repetitions |
Trigger |
|
manually, remote control or via optional trigger input (R&S NGE-K103 option) |
Interfaces |
Front panel connectors |
channel outputs |
4 mm screw-type sockets |
Remote interfaces |
standard |
USB-TMC, USB-CDC (virtual COM) |
optional |
LAN (LXI), R&S NGE-K101 |
optional |
Wireless LAN, R&S NGE-K102 |
Trigger and control interface |
optional |
Digital I/O, R&S NGE-K103 |
Mains nominal voltage |
AC |
115/230 V (±10%) 50/6 |
Power consumption |
maximum input power |
180 W |
Mains fuses |
115 V AC |
IEC 60127-2/5 T 5 A 250 V |
230 V AC |
60127-2/5 T 2.5 A 250 V |
Temperature |
operating temperature range |
0 °C to +40 °C |
storage temperature range |
-20 °C to +70 °C |
Humidity |
non-condensing |
5 % to 80 % |
3.5", QVGA |
Display |
Rackmount capability |
½ 19" |
R&S HZC95 option |
Dimensions |
W x H x D |
222 mm × 88 mm × 280 mm (8.74 in × 3.46 in × 11.02 in) |
Weight |
R&S NGE102 |
4.4 kg (9.70 lb) |
R&S NGE103 |
4.5 kg (9.92 lb) |
Recommended calibration interval |
operation 40 h/week in the full range of the specified environmental conditions |
1 year |