| Rohde & Schwarz NGA101 Specifications |
| Electrical specifications |
| Outputs |
The channel outputs are galvanically isolated and not connected to ground. |
| Number of output channels |
R&S NGA101, R&S NGA141 |
1 |
| Maximum total output power |
R&S NGA101, R&S NGA141 |
40 W |
| Maximum output power per channel |
|
40 W |
| Output voltage per channel |
R&S NGA101, R&S NGA102 |
0 V to 35 V |
| Maximum output current per channel |
R&S NGA101, R&S NGA102 |
6 A |
| Voltage ripple and noise |
20 Hz to 20 MHz |
|
| R&S NGA101, R&S NGA102 |
< 0.5 mV (RMS), < 10 mV (peak-to-peak) (meas.) |
| Current ripple and noise |
20 Hz to 20 MHz |
< 500 µA (RMS) (meas.) |
| Load regulation |
load change: 10 % to 90 % |
| Voltage |
±(% of output + offset) |
| R&S NGA101, R&S NGA102 |
< 0.01 % + 5 mV |
| Current |
±(% of output + offset) |
< 0.01 % + 5 mA |
| Load recovery time |
10 % to 90 % load change to within 0.2 % of rated voltage |
| R&S NGA101, R&S NGA102 |
< 100 µs (meas.) |
| Rise time |
10 % to 90 % of rated output voltage, resistive load |
| R&S NGA101, R&S NGA102 |
< 50 ms |
| Fall time |
90 % to 10 % of rated output voltage, resistive load |
full load: < 10 ms,
no load: < 50 ms |
| Programming resolution |
| Voltage |
|
1 mV |
| Current |
|
1 mA |
| Output measurements |
| Measurement functions |
|
voltage, current, power |
| Readback resolution |
| Voltage |
|
1 mV |
| Current |
|
100 µA |
| Low-current measurement range |
≤ 200 mA output current |
1 µA |
| Readback accuracy |
| Voltage |
±(% of output + offset) |
| R&S NGA101, R&S NGA102 |
< 0.02 % + 5 mV |
| Current |
±(% of output + offset) |
< 0.03 % + 500 µA |
| Low-current measurement range |
R&S NGA101, R&S NGA102 |
< 0.15 % + 40 µA |
| Temperature coefficient (per °C) |
+5 °C to +20 °C and +30 °C to +40 °C |
| Voltage |
±(% of output + offset) |
< 0.0075 % + 0.75 mV |
| Current |
±(% of output + offset) |
< 0.015 % + 3 mA |
| Low-current measurement range |
|
< 0.023 % + 5 µA |
| Remote sensing |
| Maximum sense compensation |
R&S NGA101, R&S NGA102 |
0.5 V (meas.) |
| Ratings |
| Maximum voltage to ground |
|
250 V DC |
| Maximum counter voltage |
voltage with the same polarity connected to the outputs |
| R&S NGA101, R&S NGA102 |
36 V |
| Maximum reverse voltage |
voltage with opposite polarity connected to the outputs |
0.4 V |
| Maximum reverse current |
for 5 min max. |
6 A |
| Remote control |
| Command processing time |
|
< 100 ms (typ.) |
| Protection functions |
| Overvoltage protection |
|
adjustable for each channel |
| Programming resolution |
R&S NGA101, R&S NGA102 |
1 mV |
| Overpower protection |
|
adjustable for each channel |
| Overcurrent protection (electronic fuse) |
|
adjustable for each channel |
| Programming resolution |
|
1 mA |
| Response time |
(Iload > Iresp × 2) at Iload ≥ 2 A |
< 1 ms |
| Fuse linking (FuseLink function) |
R&S NGA101, R&S NGA102 |
yes |
| Response time for linked channels |
|
< 5 ms |
| Fuse delay time |
adjustable for each channel |
10 ms to 10 s (10 ms increments) |
| Overtemperature protection |
independent for each channel |
yes |
| General data |
| Environmental conditions |
| Temperature |
operating temperature range |
+5 °C to +40 °C |
| storage temperature range |
-20 °C to +70 °C |
| Humidity |
noncondensing |
5 % to 95 % |
| Power rating |
| Mains nominal voltage |
|
100 V/115 V/230 V (±10 %) |
| Mains frequency |
|
50 Hz to 60 Hz |
| Maximum power consumption |
|
230 W |
| Main fuses |
100 V/115 V AC power source |
5 A, 250 V IEC 60127-2/5 T |
| 230 V AC power source |
2.5 A, 250 V IEC 60127-2/5 T |
| Product conformity |
| Electromagnetic compatibility |
EU: in line with Radio Equipment Directive 2014/53/EU |
applied standards:
- ETSI EN 300328 V2.2.2
- EN 61326-1
- EN 61326-2-1
- EN 55011 (Class A)
- EN 55032 (Class A)
- ETSI EN 301489-1 V2.1.1
- ETSI EN 301489-17 V3.1.1
|
| |
Korea |
KC mark |
| |
USA, Canada |
FCC47 CFR Part 15B, ICES-003 Issue 6 |
| Electrical safety |
EU: in line with Low Voltage Directive 2014/35/EU |
applied harmonized standard: EN 61010-1 |
| |
USA, Canada |
UL61010-1, CAN/CSA-C22.2 No. 61010-1 |
| WLAN approvals |
Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark,
Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland,
Italy, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey, United Kingdom |
CE |
| |
Singapore |
IMDA standards DB102020 |
| |
USA, Canada |
FCC, IC |
| RoHS |
in line with EU Directive 2011/65/EU |
EN 50581 |
| Mechanical resistance |
| Vibration |
sinusoidal |
5 Hz to 55 Hz, 0.3 mm (peak-to-peak), 55 Hz to 150 Hz, 0.5 g const., in line with EN 60068-2-6 |
| random |
8 Hz to 500 Hz, acceleration: 1.2 g (RMS), in line with EN 60068-2-64 |
| Shock |
|
40 g shock spectrum, in line with MIL-STD-810E, method 516.4, procedure I |
| Mechanical data |
| Dimensions |
W × H × D |
222 mm × 97 mm × 448 mm
(8.74 in × 3.82 in × 17.64 in) |
| Weight |
R&S NGA101 |
6.6 kg (14.5 lb) |
| Rack installation |
R&S HZN96, space for two instruments |
19", 2 HU |
| Recommended calibration interval |
operation 40 h/week over entire range of specified environmental conditions |
1 year |