R&S®HMC8043,R&S®HM8042, R&S®HMC8041 Power Supplies |
The specifications are based on a 30 min warm-up period. |
Electrical Specifications |
Total power output |
100W |
Maximum power per Channel |
R&S®HMC8043 |
33W |
R&S®HMC8042 50W |
R&S®HMC8041 |
100W |
Voltage Output |
0-32V |
Current Output |
R&S®HMC8043 |
3A max (power limit) |
R&S®HMC8042 |
5A max |
R&S®HMC8041 |
10A max |
Number of outputs |
R&S®HMC8043 |
3 |
R&S®HMC8042 |
2 |
R&S®HMC8041 |
1 |
Line & load regulation (Sense connected) |
Constant voltage |
R&S®HMC8043 |
<0.02% + 3mV |
R&S®HMC8042 |
<0.03% + 5mV |
R&S®HMC8041 |
<0.03% + 5mV |
Constant Current |
R&S®HMC8043 |
<0.03% 200μA |
R&S®HMC8042 |
<0.03% 200μA |
R&S®HMC8041 |
<0.03% 200μA |
Voltage ripple 20Hz to 20MHz(Front connector) |
450μV rms/ 4mVpp |
Current ripple 20Hz to 20Mhz |
typ. <1mArms |
Response time(10%...90% load change) |
1ms (±20mV) |
Remote Sense max. voltage |
1V |
Programming accuracy (23° C ±5° C) |
voltage |
all models: <0.05% +2mV |
current |
R&S®HMC8043 |
0.05% +2mA |
R&S®HMC8042/41 |
0.1% +5mA |
Readback accuracy (23° C ± 5° C) |
voltage all models |
<0.05%+2mV |
current |
R&S®HMC8043 |
0.05% +2mA |
R&S®HMC8042 |
0.05% +7mA |
R&S®HMC8041 |
0.05% +4mA |
Resolution |
voltage |
1mV |
current |
0.1mA (I<1A)
1mA (I>=1A) |
Voltage to earth |
250VDC |
Reverse Voltage |
33V max. |
Inverse Voltage |
0.4V max. |
Max. current allowed in case of inverse voltage |
3A |
Supplemental characteristics |
Front connectors |
m4 mm saftey sockets |
Rear connectors |
Wago male connector (713-1428/037-000),
8x2-pole, pin spacing 3.5 mm / 0.138 in |
Temperature coefficient for 12 months (per K)±(% of output + offset) |
voltage: >0,02% +3mV
current: >0,02%+3mA |
|
|
Output voltage overshoot during turn-off of AC power and channel output on |
100mV |
Over temperature protection |
Yes |
Voltage programming speed (within 1 % of total excursion) |
Positive voltage change |
no load |
10ms + μC-time |
with resistive load |
10ms + μC-time |
Negative voltage change |
no load |
500ms + μC-time |
with resistive load |
10ms + μC-time |
Command processing time |
<30ms |
Over Voltage Protection |
Yes |
Over Power Protection |
Yes |
Energiemeter |
Yes |
EasyRamp |
Yes |
EasyRamp time |
10ms ... 10s |
Electronic Fuse |
Fuse trip time |
<100us |
Fuse linking |
<100us + trip time of linked channel |
Fuse delay |
10ms ... 10s |
Analog Interface |
Shunt resistance 4...20mA |
250 Ohm |
Input resistance 0...10V |
>10 kOhm |
Update rate V/I interface |
10 changes/sec |
Response time V/I interface |
<150ms |
Trigger level |
TTL |
Trigger response time |
<1ms |
Resolution |
14 bit |
Arbitrary (EasyARB) |
Parameter |
Voltage, current, time and interpolation mode |
Number of Points |
512 |
Dwell time |
10ms ... 10min |
Repetition rate |
continous or burst mode with 1...255 repetitions |
Trigger |
manually, interface or trigger input |
Logging |
Sampling speed |
1000,100,10,1...3600 Sa/s |
Resolution ¸HMC8043 |
1mV / 0.1mA (<100Sa/s);
10mV / 1mA (1000Sa/s) |
Resolution ¸HMC8042/41 |
1mV / 1mA (<100Sa/s);
10mV / 10mA (1000Sa/s) |
Memory |
Internal memory and External memory(USB-Stick) |
Maximum number of Points |
limited by memory |
Sequencing |
Synchronicity |
<100us |
Delay per channel |
1ms ... 60s |
Remote interfaces |
USB-TMC, USB-CDC (Virtual COM), LAN(LXI), GPIB (optional) |
Miscellaneous |
Input power option |
100-240 VAC +/-10% 50/60 Hz |
Maximum input power |
200W |
Fuse |
T3, 15L 250V |
Operating temperature |
+0°C ..+40°C |
Storage temperature |
-20°C..+70°C |
Humidity |
5...80% |
Display |
3,5” / QVGA |
Dimensions (H x W x D) |
222 x 88 x 280 mm |
Rack mount capability 1/2 19“ |
Yes |
Weight |
2,6kg |