Rohde & Schwarz HM8118 LCR Bridge/Meter 200 kHz
Basic Accuracy 0.05 %
Binning Interface HO118 (optional) for automatic Sorting of Components
Parallel and Serial Mode
Kelvin Cable and 4-Terminal SMD Test Adapter included
The HM8118 LCR bridge offers a wide measuring frequency range (20 Hz to 200 kHz) and variable measuring voltage (0.10 to 1.50 Vrms). Additionally, it has a bias function for voltage (0 to 5.00 V and 0 to 40 V ext.) and current (0 to 200 mA). The multi-slope conversion technology and the sophisticated generation of the measuring signal allow stable capacitor measurements down to 1 femto farad (resolution 0.01 fF) and inductors down to 0.1 nH (resolution 0.01 nH). All settings and measurements can be operated manually by the front panel controls and may also be remote controlled via the built-in, galvanically isolated, dual interface (USB and RS232) or with the optional IEEE-488 (GPIB) interface (HO880).
The HM8118 LCR measuring bridge offers measurements of non ideal characteristics of real components. At the push of one button it will measure the Rs of the series equivalent circuit or the Rp of the parallel equivalent circuit of a capacitor or a inductor. The HM8118 also calculates the derived parameters, like the quality factor Q, dissipation factor D, phase angle Theta, as well as the complex values of the impedance Z and the admittance Y.
In addition to the characterization of passive components like L, C and R the HM8118 can determine the specific parameters of transformers — not only on the fixed frequencies 50/60 Hz and 100/120 Hz, but in the whole range of 20 Hz to 200 kHz.
Control Elements
- POWER (Pushbutton):Turns the instrument on/off.
- DISPLAY (LCD)
Display of measurement results and units, ranges, frequencies, level, equivalent circuit, functions, parameters,instrument states and messages.
MENU
- SELECT (Pushbutton)
Opening of the menu with the submenus SETUP, CORR, SYST and BIN (with optional Binning Interface HO118 installed)
- ENTER (Pushbutton)
Enters an input value to the HM8118.
- ESC (Pushbutton)
Deletes an input value by pushing the key
- Rotary knob(Knob/Pushbutton)
Selection and adjustment of functions and parameters
- Arrow buttons (Pushbuttons)
Pushbuttons for selection and adjustment of functions and parameters.
SET
- FREQ (Pushbutton)
Setting of the test signal frequency with the rotary knob (6) or the arrow buttons (7).
- LEVEL (Pushbutton)
Sets the test signal level with the rotary knob (6) and the Cursor position with the arrow buttons (7)
- BIAS (Pushbutton)
Sets the bias voltage or current (depending on the measure-
ment function) with the rotary knob (6) and the Cursor
position with the arrow buttons (7)
ZERO
- OPEN (Pushbutton): Executes an OPEN correction
- SHORT (Pushbutton): Executes a SHORT correction
- LOAD (Pushbutton): Executes a LOAD correction
MODE
- AUTO/MODE (Pushbutton)
Activates the automatic selection of the equivalent circuit
model of the component being measured (series, parallel)
|
- SER (Pushbutton)
Activates the series equivalent circuit model of the compo-
nent being measured
- PAR (Pushbutton)
Activates the parallel equivalent circuit model of the component being measured
RANGE
- AUTO/HOLD/ (Pushbutton)
Toggles the measurement range mode between AUTO/HOLD
- UP (Pushbutton): Switching to a higher range
- DOWN (Pushbutton): Switching to a lower range
SOCKETS
- L CUR (BNC socket)
Low Current; signal output for series measurements (signal
generator)
- L POT (BNC socket)
Low Potential; signal input for parallel measurement (voltage measurements)
- H POT (BNC socket)
High Potential; signal input / output for parallel measure-
ments (measurement bridge)
- H CUR (BNC socket)
High Current; signal input for series measurements (current
measurements)
- BIAS MODE/ESC (Pushbutton)
Selectes the bias mode from internal or external; leaving the menu without acceptance of the entered value
- TRIG MODE/ENTER (Pushbutton)
Selects the trigger mode from internal, manual, or external; enters an input value and terminates the key entry for the
current input function
- BIAS /(Pushbutton)
Enables or disables DC Bias output; erases the last cha-
racter entered when entering numeric values
|