Frequency |
Frequency range | 100 kHz to 3 GHz |
Reference frequency | Aging: 1 ppm/year Temperature drift:0 ?C to 30 ?C, 30 ?C to 50 ?C, 2 ppm in addition 2 ppm/10 ?C |
Frequency counter | Resolution:1 Hz Counter accuracy: S/N > 25 dB, ? (frequency ? reference frequency errror) |
Frequency span | 0 Hz, 10 kHz to 3 GHz |
Spectral purity | SSB phase noise: f = 500 MHz, 20 ?C to 30 ?C 30 kHz from carrier: <85 dBc (1 Hz) 100 kHz from carrier: <100 dBc (1 Hz) 1 MHz from carrier: <120 dBc (1 Hz) |
Sweep time | span = 0 Hz: 1 ms to 100 s span > 0 Hz: 20 ms to 1000 s, min. 20 ms/600 MHz |
Bandwidths | Resolution bandwidths (-3 dB): 1145.5850.13, 1, 3, 10, 30, 100, 200, 300 kHz, 1 MHz, 1145.5850.03/.23, 1145.5850.06/.26: in addition 100 Hz, 300 Hz Tolerance: ≤300 kHz, ?5 %, nominal, 1 MHz, ?10 %, nominal Resolution bandwidths (-6 dB): with option ?FSH-K3 installed, in addition 200 Hz, 9 kHz, 120 kHz, 1 MHz Video bandwidths: 10 Hz to 1 MHz in 1, 3 steps |
Amplitude |
Display range | average noise level displayed to +20 dBm |
Maximum permissible DC voltage at RF input | 50 V/80 V |
Maximum power | 20 dBm, 30 dBm (1 W) for max. 3 minutes |
Intermodulation-free dynamic range | third-order IM products, 2 ? -20 dBm, reference level = -10 dBm, typ. 66 dB (typ. +13 dBm third-order intercept, IP3) |
Displayed average noise level | resolution bandwidth 1 kHz, video bandwidth 10 Hz, reference level ≤-30 dBm 10 MHz to 3 GHz: <-105 dBm, typ. -114 dBm |
With preamplifier | only models 1145.5850.03, 1145.5850.23, 1145.5850.06, and 1145.5850.26 10 MHz to 2.5 GHz: <-120 dBm, typ. -125 dBm 2.5 GHz to 3 GHz: <-115 dBm, typ. -120 dBm |
Inherent spurious | reference level ≤-20 dBm, f > 30 MHz, RBW ≤ 100 kHz: <-80 dBm |
Input related spurious | mixer level -40 dBm, carrier offset >1 MHz Up to 3 GHz: <-70 dBc (nominal) |
2nd harmonic | mixer level -40 dBm typ. <-60 dBc |
Level display | Reference level: -80 dBm to +20 dBm in steps of 1 dB Display range: 100 dB, 50 dB, 20 dB, 10 dB, linear Display units Logarithmic: dBm, dBμV, dBmV with transducer also dBμV/m and dBμA/m Linear: μV, mV, V, nW, μW, mW, W with transducer also V/m, mV/m and μV/m Traces: 1 trace and 1 memory trace Detectors:auto peak, maximum peak, minimum peak, sample, RMS with option ?FSH-K3 installed: in addition average and quasi-peak Level measurement error:frequency >1 MHz, at reference level down to -50 dB, 20 ?C to 30 ?C: <1.5 dB, typ. 0.5 dB |
Markers | Number of markers or delta markers: max. 6 Marker functions: peak, next peak, minimum, center = marker frequency, reference level = marker level, all markers to peak Marker displays: normal (level), noise marker, frequency counter (count) |
Trigger | free-running, video, external |
Audio demodulation | AM (video voltage without AGC) and FM |
Inputs |
RF input | N female |
Input impedance | 50Ω |
VSWR | 10 MHz to 3 GHz: typ. 1.5 |
Trigger/external reference input | BNC female, selectable |
Trigger voltage | TTL |
Reference frequency | 10 MHz |
Required level | from 50Ω:10 dBm |
Outputs |
AF output | 3.5 mm mini jack |
Output impedance Open-circuit voltage | 100Ω adjustable up to 1.5 V |
Tracking generator | only models 145.5850.13, 1145.5850.23 and 1145.5850.26 |
Frequency range | 5 MHz to 3 GHz |
Output level | model 1145.5850.13:-20 dBm (nominal) model 1145.5850.23:0 dBm/-20 dBm, selectable |
Output impedance | 50 Ω, nominal |
Interfaces |
RS-232-C optical interface | Baud rate:1200, 2400, 9600, 19200, 38400, 57600, 115200 baud Power sensor:7-contact female connector (type Binder 712) |