Specifications are valid under the following conditions: the instrument is within the calibration period, is stored for at least two hours at 0°C
to 50 °C temperature, and is warmed up for 40 minutes. Unless otherwise noted, the specifications in this manual include the measurement
uncertainty.
Typical (typ.): characteristic performance, which 80 percent of the measurement results will meet at room temperature (approximately 25°C ).
This data is not warranted and does not include the measurement uncertainty.
Nominal (nom.): the expected mean or average performance or a designed attribute (such as the 50O connector). This data is not warranted
and is measured at room temperature (approximately 25°C ).
Measured (meas.): an attribute measured during the design phase which can be compared to the expected performance, such as the
amplitude drift variation with time. This data is not warranted and is measured at room temperature (approximately 25°C ).
NOTE: All charts in this manual are the measurement results of multiple instruments at room temperature unless otherwise noted. The
specifications (except the TG specifications) listed in this manual are those when the tracking generator is off. |
Frequency |
|
DSA832E |
Frequency range |
9 kHz to 3.2 GHz |
Frequency resolution |
1 Hz |
Reference frequency |
10 MHz |
Accuracy |
±[ (time since last calibration × aging rate) + temperature stability + calibration accuracy] |
Initial calibration accuracy |
<1 ppm |
Temperature stability |
0°C to 50°C , reference to 25°C |
<1 ppm |
Aging rate |
<2 ppm/year |
Marker resolution |
span/ (number of sweep points - 1) |
Marker uncertainty |
±(frequency indication × reference frequency accuracy + 1% × span + 10% × resolution bandwidth +
marker resolution) |
Resolution |
1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz |
Uncertainty |
±(frequency indication × reference frequency accuracy + counter resolution) |
Range |
0 Hz, 100 Hz to maximum frequency of instrument |
Uncertainty |
±span/ (number of sweep points - 1) |
|
20°C to 30°C , f c = 1 GHz |
Carrier offset |
10 kHz offset |
<-90 dBc/Hz |
|
20°C to 30°C , RBW = VBW = 1 kHz |
Residual FM |
<20 Hz (nom.) |
|
Set "Auto SWT" to "Accy" |
Resolution bandwidth (-3 dB) |
10 Hz to 1 MHz, in 1-3-10 sequence |
RBW uncertainty |
<5% (nom.) |
Resolution filter shape factor (60 dB : 3 dB |
<5 (nom.) |
Video bandwidth (-3 dB) |
1 Hz to 3 MHz, in 1-3-10 sequence |
Resolution bandwidth (-6 dB) (EMI-DSA800 option) |
200 Hz, 9 kHz, 120 kHz |
Amplitude |
Range |
fc = 10 MHz |
DANL to +20 dBm |
DC voltage |
50 V |
CW RF power |
attenuation = 30 dB |
+20 dBm (100 mW) |
Max. damage level [1] |
+30 dBm (1 W) |
NOTE: [1] When f c = 10 MHz, input level > +25 dBm and PA is Off, the protection switch will be on. |
|
attenuation = 0 dB, RBW = VBW = 10 Hz, sample detector, trace average = 50,
tracking generator off, 20°C to 30°C , input impendence = 50 Ω |
PA off |
9 kHz to 100 kHz |
<-110 dBm (typ.) |
100 kHz to 5 MHz |
<-122 dBm, <-125 dBm (typ.) |
5 MHz to 3.2 GHz |
<-127 dBm, <-130 dBm (typ.) |
PA on |
100 kHz to 1 MHz |
<-142 dBm (typ.) |
1 MHz to 5 MHz |
<-140 dBm, <-143 dBm (typ.) |
5 MHz to 3.2 GHz |
<-145 dBm, <-148 dBm (typ.) |
Logarithmic level axis |
1 dB to 200 dB |
Linear level axis |
0 to reference level |
Number of display points |
601 |
Number of traces |
3 + math trace |
Trace detectors |
normal, positive-peak, negative-peak, sample, RMS, voltage average |
quasi-peak (with EMI-DSA800 option) |
Trace functions |
clear write, max hold, min hold, average, view, blank |
Units of level axis |
dBm, dBmV, dBµV, nV, µV, mV, V, nW, µW, mW, W |
|
f c = 100 kHz, attenuation = 10 dB, relative to 50 MHz, 20°C to 30°C |
PA off |
100 kHz to 3.2 GHz |
<0.7 dB |
|
f c = 1MHz, attenuation = 10 dB, relative to 50 MHz, 20°C to 30°C |
PA on |
100 kHz to 3.2 GHz |
<1.0 dB |
Setting range |
0 dB to 30 dB, in 1 dB step |
Switching uncertainty |
f c = 50 MHz, relative to 10 dB, 20°C to 30°C |
<0.3 dB |
Uncertainty |
f c = 50 MHz, peak detector, preamplifier off, attenuation = 10 dB, input signal level = -10dBm, 20°C to 30°C |
<0.3 dB |
Uncertainty |
relative to 1 kHz RBW |
<0.1 dB |
Range |
-100 dBm to +20 dBm, in 1 dB step |
Resolution |
log scale |
0.01 dB |
linear scale |
4 digits |
|
PA-DSA832 (Standard - As of 4/1/2017) |
Gain |
100 kHz to 3.2 GHz |
17 dB (nom.) |
|
95% confidence level, S/N > 20 dB, RBW = VBW = 1 kHz, preamplifier off, attenuation = 10 dB, -50 dBm <
input level = 0 dBm, f c > 10 MHz, 20°C to 30°C |
Level measurement
uncertainty |
<1.0 dB (nom.) |
|
attenuation = 10 dB |
VSWR |
300 kHz to 3.2 GHz |
<1.5 (nom.) |
Distortion |
Second harmonic intercept
(SHI) |
f c = 50 MHz, input signal level = -20 dBm, attenuation = 10 dB |
+40 dBm |
Third-order intercept
(TOI) |
f c = 50 MHz, two -20 dBm tones at input mixer spaced by 200 kHz, attenuation = 10 dB |
+7 dBm |
1dB compression of input
mixer (P 1dB ) |
f c = 50 MHz, attenuation = 0 dB |
>0 dBm |
Spurious response, inherent |
input terminated 50 Ω, attenuation = 0 dB, 20°C to 30°C |
<-90 dBm [2] , <-100 dBm (typ.) |
Intermediate frequency |
<-60 dBc |
System related sidebands |
referenced to local oscillators, referenced to A/D conversion, referenced to subharmonic of first LO,
referenced to harmonic of first LO |
<-60 dBc |
Input related spurious |
mixer level = -30dBm |
<-60 dBc |
NOTE: [2] Except the internal local oscillator (1820 MHz) and its harmonics. |
Sweep |
Sweep time |
span = 100 Hz |
1 ms to 3200 s |
zero span |
20 µs to 3200 s |
Sweep time uncertainty |
span = 100 Hz |
5% (nom.) |
zero span (sweep time setting
value > 1 ms) |
5% (nom.) |
Sweep mode |
continuous, single |
Tracking Generator (Option) |
Frequency range |
100 kHz to 3.2 GHz |
Output level range |
-40 dBm to 0 dBm |
Output level resolution |
1 dB |
Output flatness |
relative to 50 MHz |
±3 dB (nom.) |
Trigger |
Trigger source |
Trigger source |
External trigger level |
External trigger level |
Input /Output |
RF input |
impedance |
50 Ω (nom.) |
connector |
N female |
Tracking generator output |
impedance |
50 Ω (nom.) < |
connector |
N female |
Internal reference |
frequency |
10 MHz |
output level |
+3 dBm to +10 dBm, +8 dBm (typ.) |
impedance |
50 Ω (nom.) |
connector |
BNC female |
External reference |
frequency |
10 MHz ± 5 ppm |
input level |
0 dBm to +10 dBm |
impedance |
50 Ω (nom.) |
connector |
BNC female |
External trigger input |
impedance |
1 kΩ (nom.) |
connector |
BNC female |
USB host |
connector |
A plug |
protocol |
version2.0 |
USB device |
connector |
B plug |
protocol |
version2.0 |
LAN |
LXI core 2011 device |
10/100Base, RJ-45 |
IEC/IEEE (GPIB) bus (USB-GPIB option) |
IEEE488.2 |
General Specification |
Type |
TFT LCD |
Resolution |
800 x 480 pixels |
Size |
8 inch |
Colors |
64k |
Protocol |
PictBridge |
Mass memory |
flash disk (internal),
USB storage device (not supplied) |
Input voltage range, AC |
100 V to 240 V (nom.) |
AC supply frequency |
45 Hz to 440 Hz |
Power consumption |
35 W (typ.), max. 50 W with all options |
Temperature |
operating temperature range |
0°C to 50°C |
storage temperature range |
-20°C to 70°C |
Humidity |
0°C to 30°C |
= 95% rel. humidity |
30°C to 40°C |
= 75% rel. humidity |
Altitude |
operating height |
up to 3,000m |
EMC |
in line with EN61326-1:2006 |
IEC 61000-4-2:2001 |
±4.0 kV (contact discharge), ±4.0 kV (air discharge) |
IEC 61000-4-3:2002 |
3 V/m (80 MHz to 1 GHz), 3 V/m (1.4 GHz to 2 GHz), 1 V/m (2.0 GHz
to 2.7 GHz) |
IEC 61000-4-4:2004 |
1 kV power lines |
IEC 61000-4-5:2001 |
0.5 kV (phase to neutral), 0.5 kV (phase to PE), 1 kV (neutral to PE) |
IEC 61000-4-6:2003 |
3 V, 0.15 to 80 MHz |
IEC 61000-4-11:2004 |
voltage dip: 0% UT during half cycle, 0% UT during 1 cycle, 70% UT
during 25 cycles
short interruption: 0% UT during 250 cycles |
Electrical safety |
in line with
UL 61010-1:2012, CAN/CSA-C22.2 No. 61010-1-12, EN 61010-1:2010 |
(W x H x D) |
361.6 mm × 178.8 mm × 128 mm
(14.2 in × 7.0 in × 5.0 in) |
Standard |
4.55 kg (10.0 lb) |
With tracking generator |
5.15 kg (11.4 lb) |
Recommended calibration interval |
1 year |