Frequency |
Frequency |
Model |
DSA815
DSA815-TG (with preamplifier and tracking generator) |
DSA832 (with preamplifier as of 4/1/2017) |
DSA875 (with preamplifier as of 4/1/2017) |
Frequency range |
9 kHz to 1.5 GHz |
9 kHz to 3.2 GHz |
9 kHz to 7.5 GHz |
Frequency range |
1 Hz |
Internal Reference Frequency |
Reference frequency |
10MHz |
Accuracy |
±[(time since last adjustment × aging rate)+temperature stability + calibration accuracy] |
Initial calibration accuracy |
<1ppm |
Initial calibration accuracy |
0ºC to 50ºC , reference to 25ºC |
< 2ppm |
< 0.5ppm |
Aging rate |
< 2ppm/year |
< 1ppm/year |
Frequency Readout Accuracy |
Marker resolution |
span/ (number of sweep points - 1) |
Marker uncertainty |
±(frequency indication × frequency reference uncertainty + 1% × span + 10% × resolution bandwidth + marker resolution) |
Frequency Counter |
Resolution |
1 Hz, 10 Hz, 100 Hz, 1 kHz, 10kHz, 100kHz |
Uncertainty |
±(frequency indication × reference frequency accuracy + counter resolution) |
Frequency Span |
Range |
0Hz, 100Hz to maximum frequency of instrument |
Uncertainty |
±span/ (number of sweep points - 1) |
SSB Phase Noise |
|
20ºC to 30ºC , f c =1 GHz |
Carrier offset |
DSA815 |
DSA832 |
DSA875 |
10 kHz |
<-80 dBc/Hz |
<-98 dBc/Hz |
|
|
Residual FM |
|
20ºC to 30ºC , RBW = VBW = 1 kHz |
|
DSA815 |
DSA832 |
DSA875 |
Residual FM |
< 50 Hz (nom.) |
< 20Hz (nom.) |
Bandwidths |
|
DSA815 |
DSA832 |
DSA875 |
Resolution bandwidth (-3dB) |
100 Hz to 1 MHz, in 1-3-10 sequence |
10 Hz to 1 MHz, in 1-3-10 sequence |
RBW uncertainty |
<5% (nom.) |
Resolution filter shape factor
(60dB: 3dB) |
<5 (nom.) |
Video bandwidth (-3dB) |
1 Hz to 3 MHz, in 1-3-10 sequence |
Resolution bandwidth (-6dB)
(EMI-DSA800 option) |
200 Hz, 9 kHz, 120 kHz |
Amplitude |
Measurement Range |
Range |
f c ≥10MHz |
DANL to +20 dBm |
Maximum Input Level |
DC voltage |
50 V |
CW RF power |
attenuation = 30 dB |
+20 dBm (100 mW) |
Max. damage level [1] |
+30 dBm (1 W) |
NOTE: [1] When f c ≥ 10 MHz, input level > +25 dBm and PA is Off, the protection switch will be on. |
|
|
Displayed Average Noise Level (DANL) |
|
DSA815 |
Frequency |
attenuation = 0 dB, RBW = VBW = 100 Hz, sample detector, trace average ≥ 50, tracking generator off, 20ºC to 30ºC , input impendence = 50 Ω |
PA off |
100 kHz to 1 MHz |
<-90 dBm, <-110 dBm (typ.) |
1MHz to 1.5 GHz |
<-110 dBm+6×(f/1GHz)dB, <-115 dBm (typ.) |
PA on |
100 kHz to 1 MHz |
<-110 dBm, <-130 dBm (typ.) |
1MHz to 1.5 GHz |
<-130 dBm+6×(f/1GHz)dB, <-135 dBm (typ.) |
Displayed Average Noise Level (DANL) |
|
DSA832 |
DSA875 |
Frequency |
attenuation = 0 dB, RBW = VBW = 10 Hz, sample detector, trace average ≥ 50, tracking generator off, 20ºC to 30ºC , input impendence = 50 Ω |
PA off |
9 kHz to 100 kHz |
<-110 dBm (typ.) |
<-110 dBm (typ.) |
100 kHz to 5 MHz |
<-125 dBm, <-128 dBm (typ.) |
<-125 dBm, <-128 dBm (typ.) |
5 MHz to 3.2 GHz |
<-130 dBm, <-134 dBm (typ.) |
<-130 dBm, <-134 dBm (typ.) |
3.2 GHz to 6 GHz |
|
<-126 dBm, <-130 dBm (typ.) |
6 GHz to 7.5 GHz |
|
<-121 dBm, <-125 dBm (typ.) |
PA on |
100 kHz to 5 MHz |
<-142 dBm, <-145 dBm (typ.) |
<-142 dBm, <-145 dBm (typ.) |
5 MHz to 3.2 GHz |
<-147 dBm, <-151 dBm (typ.) |
<-147 dBm, <-151 dBm (typ.) |
3.2 GHz to 6 GHz |
|
<-143 dBm, <-147 dBm (typ.) |
6 GHz to 7.5 GHz |
|
<-138 dBm, <-142 dBm (typ.) |
Displayed Average Noise Level (DANL) (Normalized to 1Hz) |
|
DSA815 |
DSA832 |
DSA875 |
Frequency |
attenuation = 0 dB, RBW = VBW = 100 Hz, sample detector, trace average ≥ 50, tracking
generator off, normalized to 1Hz, 20ºC to 30ºC , input impendence = 50 Ω |
PA off |
9 kHz to 100 kHz |
|
<-120 dBm (typ.) |
<-120 dBm (typ.) |
100 kHz to 1 MHz |
<-110 dBm,
<-130 dBm (typ.) |
<-135 dBm,
<-138 dBm (typ.) |
<-135 dBm,
<-138 dBm (typ.) |
1MHz to 5MHz |
<-130 dBm+6×(f/1GHz)
dB,
<-135 dBm (typ.) |
5 MHz to 1.5 GHz |
<-140 dBm,
<-144 dBm (typ.) |
<-140 dBm,
<-144 dBm (typ.) |
1.5 GHz to 3.2 GHz |
|
3.2 GHz to 6 GHz |
|
|
<-136 dBm,
<-140 dBm (typ.) |
6 GHz to 7.5 GHz |
|
|
<-131 dBm,
<-135 dBm (typ.) |
PA on |
100 kHz to 1 MHz |
<-130 dBm,
<-150 dBm (typ.) |
<-152 dBm,
<-155 dBm (typ.) |
<-152 dBm,
<-155 dBm (typ.) |
1MHz to 5MHz |
<-150dBm + 6×
(f/1GHz)dB,
<-155 dBm (typ.) |
5 MHz to 1.5 GHz |
<-157 dBm,
<-161 dBm (typ.) |
<-157 dBm,
<-161 dBm (typ.) |
1.5 GHz to 3.2 GHz |
|
3.2 GHz to 6 GHz |
|
|
<-153 dBm,
<-157 dBm (typ.) |
6 GHz to 7.5 GHz |
|
|
<-148 dBm,
<-152 dBm (typ.) |
|
|
Level Display |
Logarithmic level axis |
1 dB to 200 dB |
Linear level axis |
0 to Reference Level |
Number of display points |
601 |
Number of traces |
3 + Math Trace |
Trace detectors |
normal, positive-peak, negative-peak, sample, RMS, voltage average |
quasi-peak (with EMI-DSA800 option) |
Trace functions |
clear write, max hold, min hold, average, view, blank |
Units of level axis |
dBm,dBmV, dBμV, nV, μV, mV, V, nW, μW, mW, W |
Frequency Response |
|
DSA815 |
DSA832 |
DSA875 |
Frequency response |
f c ≥100kHz, attenuation = 10 dB, relative to 50 MHz, 20ºC to 30 ºC |
PA off |
100kHz to 1.5GHz |
<0.7 dB |
<0.5 dB, <0.3 dB(typ.) |
1.5GHz to 3.2GHz |
|
3.2GHz to 7.5GHz |
|
<0.7 dB, <0.3 dB(typ.) |
|
f c ≥1MHz, attenuation = 10 dB, relative to 50 MHz, 20ºC to 30 ºC |
PA on |
100kHz to 1.5GHz |
<1.0 dB |
<0.7 dB, <0.3 dB(typ.) |
1.5GHz to 3.2GHz |
|
3.2GHz to 7.5GHz |
<1.0 dB |
<0.9 dB, <0.3 dB(typ.) |
|
|
Input Attenuation Switching Uncertainty |
|
DSA815 |
DSA832 |
DSA875 |
Setting range |
0 to 30 dB, in 1 dB step |
Switching uncertainty |
f c =50 MHz, relative to 10 dB, 20 ºC to 30 ºC |
< 0.5 dB |
< 0.3 dB |
Absolute Amplitude Uncertainty |
|
DSA815 |
DSA832 |
DSA875 |
Uncertainty |
fc = 50 MHz, peak detector, preamplifier off, attenuation = 10 dB, input signal level = -10 dBm,
20 ºC to 30 ºC |
<0.4dB |
<0.3 dB |
RBW Switching Uncertainty |
Uncertainty |
relative to 1 kHz RBW |
<0.1 dB |
Reference Level |
Range |
-100 dBm to +20 dBm, in 1 dB step |
Resolution log Scale |
0.01 dB |
Resolution linear Scale |
4 digits |
Preamplifier |
|
DSA815 (standard) |
PA-DSA832 (standard - as of 4/1/2017) |
PA-DSA875 (standard - as of 4/1/2017) |
Gain |
100kHz to 1.5 GHz |
20 dB(nom.) |
17 dB(nom.) |
17 dB(nom.) |
1.5GHz to 3.2 GHz |
|
3.2GHz to 7.5 GHz |
|
|
Level Measurement Uncertainty |
|
DSA815 |
DSA832 |
DSA875 |
|
95% confidence level, S/N > 20 dB, RBW = VBW = 1 kHz,
preamplifier off, attenuation = 10 dB, -50 dBm < input level ≤ 0 dBm, fc > 10 MHz, 20 ºC to 30 ºC |
Level measurement uncertainty |
<1.5 dB(nom.) |
<0.8 dB (nom.) |
|
|
RF Input VSWR |
|
DSA815 |
DSA832 |
DSA875 |
attenuation ≥ 10 dB |
VSWR |
300kHz to 1.5GHz |
<1.5(nom.) |
<1.5(nom.) |
<1.5(nom.) |
1.5GHz to 3.2GHz |
|
3.2GHz to 7.5GHz |
|
|
<1.8(nom.) |
|
|
Distortion |
Second Harmonic Intercept |
|
DSA815 |
DSA832 |
DSA875 |
Second harmonic Intercept (SHI) |
f c ≥ 50 MHz, input signal level = -20 dBm, attenuation = 10 dB |
+40dBm |
+45dBm |
Third-order Intercept |
|
DSA815 |
DSA832 |
DSA875 |
Third-order intercept
(TOI) |
f c ≥ 50 MHz, two -20dBm tones at input mixer spaced by 200kHz, attenuation = 10 dB |
+10dBm |
+11 dBm, +15 dBm (typ.) |
1dB Gain Compression |
1dB compression of input mixer
(P 1dB ) |
fc ≥ 50 MHz, attenuation = 0 dB |
>0dBm |
|
|
Spurious Responses |
Spurious response |
DSA815 |
DSA832 |
DSA875 |
input terminated 50 Ω, attenuation = 0 dB, 20ºC to 30ºC |
< -88dBm (typ.) |
< -90dBm,
< -100dBm (typ.) |
Intermediate frequency |
< -60 dBc |
System related sidebands |
referenced to local oscillators, referenced to A/D conversion, referenced to subharmonic of first LO, referenced to harmonic of first LO |
< -60 dBc |
Input Related Spurious |
mixer level = -30 dBm |
< -60 dBc |
Sweep |
Sweep |
|
DSA815 |
DSA832 |
DSA875 |
Sweep time |
Span≥100 Hz |
10ms to 1500s |
1ms to 1500s |
1ms to 1500s |
zero span |
20 μs to 1500 s |
20 μs to 3200 s |
20 μs to 7500 s |
Sweep time uncertainty |
zero span (sweep time setting value > 1 ms) |
5%(nom.) |
Sweep mode |
continuous, single |
Tracking Generator (Option) |
TG Output |
|
DSA815 |
DSA832 |
DSA875 |
Frequency range |
100 kHz to 1.5 GHz |
100 kHz to 3.2 GHz |
100 kHz to 7.5 GHz |
Output level range |
-20 dBm to 0 dBm |
-40 dBm to 0 dBm |
Output level resolution |
1 dB |
Output flatness |
relative to 50 MHz |
±3 dB (nom.) |
|
|
Trigger Functions |
Trigger |
Trigger source |
Free run, video, external |
External trigger level |
5 V TTL level |
Input /Output |
Front Panel Connectors |
RF input |
impedance |
50 Ω (nom.) |
connector |
N female |
Tracking generator output |
impedance |
50 Ω (nom.) |
connector |
N female |
Internal/ External Reference |
Internal reference |
frequency |
10 MHz |
output level |
+3 dBm to +10 dBm, +8 dBm (typ.) |
impedance |
50Ω (nom.) |
connector |
BNC female |
External reference |
frequency |
10 MHz±5 ppm |
output level |
0 dBm to +10 dBm |
impedance |
50Ω (nom.) |
connector |
BNC female |
External Trigger Input |
External trigger input |
impedance |
1 kΩ (nom.) |
connector |
BNC female |
Communication Interface |
USB host |
connector |
A plug |
protocol |
Version2.0 |
USB device |
connector |
B plug |
protocol |
Version2.0 |
LAN |
LXI core 2011 device |
10/100Base, RJ-45 |
IEC/IEEE(GPIB) bus(USB-GPIB option) |
IEEE488.2 |
General Specifications |
Display |
Type |
TFT LCD |
Resolution |
800 x 480 pixels |
Size |
8 inch |
Colors |
64 k |
Printer Supported |
Protocol |
PictBridge |
Mass Memory |
Mass Memory |
Flash Disk (internal), USB Disk (not supplied) |
Power Supply |
Input Voltage Range, AC |
100 V to 240 V (nom.) |
AC Supply Frequency |
45 Hz to 440 Hz |
Power Consumption |
35 W (typ.),max. 50 W with all options |
Environmental |
Temperature |
Operating temperature range |
0ºC to 50ºC |
Storage temperature range |
-20ºC to 70ºC |
Humidity |
0ºC to 30ºC |
≤95% rel. humidity |
30ºC to 40ºC |
≤75% rel. humidity |
Altitude |
operating height |
up to 3,000m |
Electromagnetic Compatibility and Safety |
EMC |
In line with EN61326-1:2006 |
IEC 61000-4-2:2001 |
±4.0kV (contact discharge), ±4.0kV (air discharge) |
IEC 61000-4-3:2002 |
3V/m (80MHz to 1GHz)
3V/m (1.4GHz to 2GHz)
1V/m (2.0GHz to 2.7GHz) |
IEC 61000-4-4:2004 |
1kV power lines |
IEC 61000-4-5:2001 |
0.5kV (Phase to Neutral)
0.5kV (Phase to PE)
1kV (Neutral to PE) |
IEC 61000-4-6:2003 |
3V,0.15-80MHz |
IEC 61000-4-11:2004 |
Voltage dip:
0% UT during half cycle
0% UT during 1 cycle
70% UT during 25 cycles
Short interruption:0% UT during 250 cycles |
Electrical Safety |
In line with
UL 61010-1:2012, CAN/CSA-C22.2 No. 61010-1-12, EN 61010-1:2010 |
Dimensions |
(W x H x D) |
361.6 mm × 178.8 mm × 128 mm
(14.2 in × 7.0 in × 5.0 in) |
Weight |
|
DSA815 |
DSA832 |
DSA875 |
Standard |
4.25 kg (9.4 lb) |
4.55 kg (10.0 lb) |
With tracking generator |
5.15 kg (11.4 lb) |