Range |
Optimal Resistance Range from 0.1 (1.0E-1) Ohms to 1 Tera ohms (1.0E+12 ohms). |
TEST VOLTAGES |
Automatic mode Default |
0.01 volts to 10 volts Variable 1.0E-1 to 1.0E+4 Ohms |
Constant Voltage |
10 volts: ± <0.2 volts 1.0E+4 to <1.0E+6 Ohms |
At 10 & 100 Volts |
100 Volts ± <2.0 Volts 1.0E+6 to 2.0E+14 Ohms |
Manual Mode |
0.01 volts to 10 volts Variable 1.0E-1 to <1.0E+5 Ohms
10 volts: ± <0.2 volts 1.0E+2 to 1.0E+9 Ohms
100 volts: ± <2.0 volts 2.0E+5 to >1.0E+12 Ohms |
ACCURACY |
Overall |
± < 5% at ambient conditions (at 23o C and 30% Rh). |
Specific Range Tolerances |
1.0E-1 to 1.0E+1 ohms: ±5% Corrected for Test Lead Resistance
1.0E+1 to 1.0E+12 ohms: ±2.0% with 10-foot Test Leads
1.0E+12 to 2.0E+14 ohms: <±40% |
Display |
Multi-function 2-5/8” x 1-5/8” Liquid Crystal Display with 0.5" digit height.
Displays 3-1/2 digits in Ohms, or 1.0EXX in exponential format.
Ohms Display indicators: Ω, KΩ, MΩ, GΩ and TΩ. Includes 19-segment analog scale (1-10 with 0.5 indication) with x1, x10, & x100 multipliers.
Number of Data Points in Memory (0 – 80).Automatic Electrification Time (seconds), or Time required to Manually obtain steady state measurement.
Displays data HOLD, BATTERY status, MIN, MAX, AVG, REC and Test Voltage (<10, 10, or 100 V) |
LED Indicators |
14 Color LED’s from <103 to >1012 ohms.
Colors (RED, GREEN, YELLOW/ORANGE, or Blank/OFF) |
Timer |
Times measurements in Seconds up to 99 seconds (Displayed on LCD) |
Memory |
Register stores up to 80 data points (MEM # Displayed after RESET) |
Response & Electrification |
Response from >0.1 to <1.0E06 Ohms: <2.0 seconds Average Measurement Period from 0.1 ohms to 1.0E12 Ohms 2.5 Seconds. Calculated Electrification Period per ESD S11.11, 7.5 seconds 0.1 ohms to 10E+12 Ohms. Programmed Electrification 1.0E+12 to >2.0E14 Ohms: 15.0+ seconds |
Power |
Two 9-VDC alkaline batteries. Nominal battery life 25 hours in Automatic
Greater than 35 hours in Automatic Manual |
Dimensions (W x L x D) |
4.0" wide x 6.0" long x 2.0" deep. |
Weight |
22 ounces, with batteries |
Open Circuit Current (I) |
<4 ma @ 100 Volts |
Resistance Range Select |
2 Triangular Arrow Buttons: UP and DOWN ¯.
Select Resistance Range in single decades in Manual and Automatic/Manual modes. |
Test Volts |
Manual Selection of <10, 10 or 100 volts in Manual Mode |
Record/Recall |
Turns Memory Register ON if OFF. Provides access to all data in Memory Register.
Calculates and Displays Minimum, Maximum and Average of data stored in Memory Register. |
Clear |
Erases all data in Memory Register; if in HOLD mode, discards the most recent Held Value |
ON/OFF |
Power-up, perform functional & Battery tests. Power down if ON. |
Battery Test |
Displays GOOD on LCD if acceptable voltage or Lo if unacceptable |
Reset |
Enters (saves) data into Memory Register, Clears HOLD and Display. |
Test |
Begins measurement sequence. |
Battery Buss Cut Off |
ON/OFF Switch isolates batteries from instrument circuits for storage & transport |
USEFUL MODES OF OPERATIONS |
Auto Mode |
The instrument automatically selects and adjusts test voltage, resistance range, electrification period, then displays and Holds the
measurement. The displayed measurement is the average of eight consecutive measurements, all within ±5% of each other. The measurement is “saved” in the Memory Register by pressing
RESET. The instrument is now ready for the next measurement. |
Manual Mode |
Allows operator to select resistance decade, test voltage and electrification period (EP). |
Automatic Manual |
Same as Auto Mode with following exceptions:
- Allows operator to select starting resistance decade
- Always starts measurement from the last measurement value without resetting to zero. This extends battery life & speeds up measurement sequence. |
Data Log & Calculation |
In RECORD mode, the PRS-812 stores up to 80 measurements, and on demand will calculate and display the Minimum, Maximum and Average measurements stored in the register. |
Usage Recommendation |
Designed for Intermittent use. Not intended for continuous use or production applications. |