Controls |
ON/OFF Battery Bus Cut-off |
Slide Switch ON SIDE OF CASE |
ON: Energizes instrument and LCD displays Software Version then GO A with AUTO indication (Default) |
OFF: Isolates battery and DC voltage from instrument controls and circuit. Use for transport and storage |
TEST Pad |
With Test Leads installed: Press once & Release |
Measurements performed automatically |
In conjunction with MODE changes from Auto operation to Continuous |
MODE Pad |
Steps through Auto/Voltage ranges with each press of MODE Pad: |
AUTO: Default automatic resistance range & Voltage control |
<10V: Manual Operation 0.01 to = 10V Test Voltage; |
Resistance Range: 0.01E-1 to 9.99E9 O |
10V: Manual Operation at 10V (Nominal, Semi-Constant) |
Resistance Range: 0.98E4 to 9.99E10 O |
100V: Manual Operation at 100V (±<5V) Constant Voltage |
Resistance Range: 0.98E6 to 9.99E12 O |
When data in HOLD, press MODE to terminate HOLD without changing operational mode |
Press MODE with TEST to change from Auto to Continuous |
Used to measure lead resistance (NULL) |
Used by Prostat Authorized Technicians for Voltage Calibration |
CONNECTIONS |
Positive (+): Test Voltage (RED) – Supplies test voltage to electrodes/fixtures |
Negative (-): Current sensing (BLACK) – Conducts current from electrode, fixtures, or Ground to instrument |
Chassis: Shielding connection to chassis (WHITE) – Reduces transmitted energy and effect on current measurement |
Note: ¾ inch spacing between RED and BLACK accommodates dual banana BNC adapters |
INDICATION |
Colored LED’s indicate instrument operational MODE and Test Voltage (YELLOW) used for the current measurement, as well as measurement completion (GREEN LED) |
AUTO: ON during automatic operations; OFF in Manual mode |
<10V |
10V |
100V |
GREEN LED: Measurement hold |
TYPICAL TOLERANCES |
1Nominal performance by range: |
0.01E- 1 - 1.0E 0 O <±10%, ±2 Counts |
1.0E 0 - 9.99E10 O <±2%, ±2 Counts |
1.0E11 - 9.99E12 O <± 10%, ±4 Counts |
TEST VOLTAGES |
Automatically selected and controlled in AUTO mode. May be manually selected by pressing MODE pad in the following order: |
<10V Variable based on resistance load: |
Provides 0.001 to =10V |
Functional in AUTO within range 0.01 O - <1.0E+4 O |
Manual Operation Resistance Range: 0.01E-1 to 9.99E+9 O |
10V Constant Voltage (Under Load): |
Test Voltage Varies ± <5% in AUTO or Manual Operations |
Functional in AUTO within range 1.0E+4O - <1.0E+6 O |
Manual Operation Resistance Range: 0.9E+4 to 9.99E+10 O |
100V Constant Voltage (Under Load): |
Test Voltage Varies ± <2% in AUTO or Manual Operations |
Functional in AUTO within range 1.0E+6 O - <1.0E+13 O |
Manual Operation Resistance Range: 0.90E6 to 9.99E12 O |
Returns to Auto Mode |
MODES OF OPERATIONS |
GO A: Performs all automatic and manual voltage select functions, stops and displays measurements after Electrification Period (EP) is complete, then displays data with Green LED ON for 10 seconds |
GO C: Performs all automatic and manual voltage select functions, continuously displays measurements until the operator presses MODE or TEST pads. GO C disables electrification period controls |
POWER |
One 9V alkaline transistor battery; battery life > 50 hours |
DIMENSIONS |
7.25 in. (184 mm) x 4.3 in, (109 mm) x 1.4 in. |
WEIGHT |
13.2 oz. (375 gram) with battery installed |
NOTES |
1Tolerance and accuracy measurements were conducted under controlled conditions with close tolerance reference resistors |
Do not attempt to open instrument case: |
No internal operational parts or adjustments |
Provides access to 100DCV power supply |
Damage to instrument and environmental seals may occur |
Will void warranty |
Store instrument in case and dry environment |