Pico CC017 Oscilloscope Calibration (PS2000, 3000, 4000)
The CC017 is a calibration service offered by Pico on a selection of PicoScope oscilloscopes as listed below:
- PicoScope 2000 Series
- PicoScope 3000 Series
- PicoScope 4000 Series except PicoScope 4824
- PicoScope 5203 and 5204
All our products are tested to ensure they meet their specifications. This means that you can be confident that all measurements you make with your Pico product will be within the accuracy that we specify.
If you plan to use your product in an industrial or scientific environment where it is necessary to have documented traceability, or you just want your Pico product to be ‘fine-tuned’ to give the best possible performance, then we can provide you with a calibration service referenced to National Standards. This means that you will be given documented proof of the accuracy of your Pico product at the time it is despatched to you. This documented proof is in the form of a calibration certificate that has the errors recorded. For example, the product may have a published accuracy of ±3%, and after calibration the error might be –1%. You can then allow for this –1% error when taking measurements.
For best accuracy, we suggest that a calibrated product is recalibrated yearly. For our recalibration service on your existing instrument, please contact our technical support team.
Pico products can also be calibrated by any accredited calibration provider.