Specifications |
MODEL |
VERTICAL |
Input channels |
4, BNC connectors, single-ended |
Analog bandwidth (-3 dB) * |
250 MHz (200 MHz on ±50 mV range) |
350 MHz (250 MHz on ±50 mV range) |
500 MHz |
Bandwidth limiter |
20 MHz, switchable |
20 MHz, switchable |
25 MHz, switchable |
Rise time (10% to 90%, calculated) |
1.4 ns (50 mV range 1.8 ns) |
1.0 ns (50 mV range 1.4 ns) |
0.7 ns (all ranges) |
Input ranges (full scale) |
±50 mV to ±20V, in 9 ranges (1 MΩ input), ±50 mV to ±5V, in 7 ranges (50 Ω input) |
Input sensitivity |
10 mV/div to 4 V/div at x1 zoom (1 MΩ input), 10 mV/div to 1 V/div at x1 zoom (50 Ω input) |
Input coupling |
AC (1 MΩ) or DC (1 MΩ or 50 O) |
Input characteristics |
1 MΩ ± 1% || 15 pF, or 50 Ω ± 2% |
1 MΩ ± 1% || 10 pF, or 50 Ω ± 2% |
Analog offset range |
±50V to ±200 mV input ranges: ±0.5 V
±500V input range: ±2.5 V
±1V input range: ±2.5 V
±2V input range: ±2.5 V
±5V input range: ±20 V (50 O: ±0.5 V)
±10V input range: ±20 V
±20V input range: ±20 V |
±2 V
±10 V (50 &OMega;: ±5 V)
±10 V (50 &OMega;: ±4.5 V)
±10 V (50 &OMega;: ±3.5 V)
±10 V (50 &OMega;: ±0.5 V)
±30 V
±20 V |
DC accuracy |
3% of full scale |
Overvoltage protection |
±100 V to ground (1 MΩ inputs), 5.5 V RMS (50 Ω inputs) |
* Stated bandwidth is with supplied probes or at BNC when 50 Ω impedance selected |
DYNAMIC PERFORMANCE |
Noise |
200 µV RMS (50 mV range) |
320 µV RMS (50 mV range) |
THD |
–55 dB typical |
–54 dB typical |
SFDR |
60 dB typical |
55 dB typical |
Crosstalk |
17 000:1 typical at 20 MHz
1000:1 typical at full bandwidth |
5600:1 typical at 20 MHz
560:1 typical at full bandwidth |
HORIZONTAL (TIMEBASE) |
Timebase ranges |
1 ns/div to 5000 s/div (real-time sampling)
50 ps/div to 100 ns/div (equivalent-time sampling / ETS) |
Timebase accuracy |
±2 ppm |
Timebase ageing |
1 ppm per year |
ACQUISITION |
ADC resolution |
8 bits (up to 12 bits using software resolution enhancement) |
Maximum real-time sampling rate |
1 channel 5 GS/s
2 channels 2.5 GS/s **
4 channels 1.25 GS/s |
Maximum ETS rate < |
50 GS/s (any number of channels) |
Maximum streaming data rate (PicoScope 6) |
10 MS/s |
Maximum streaming data rate (SDK) |
Data transfer > 150 MS/s, streaming to SSD hard drive 78 MS/s
(USB 3.0, PC-dependent, subject to application loadings) |
Buffer size (shared between active channels) |
256 MS |
512 MS |
512 MS |
1 GS |
1 GS |
2 GS |
Buffer size (streaming Mode) |
100 MS in PicoScope software. Up to available PC memory when using SDK. |
Max. buffer segments (using PicoScope 6) |
10 000 |
Max. buffer segments (using SDK) |
250 000 |
500 000 |
500 000 |
1 000 000 |
1 000 000 |
2 000 000 |
** To achieve 2.5 GS/s sampling rate in 2-channel Mode, use channel A or B and channel C or D. |
TRIGGERING |
Sources |
Channels A to D, AUX |
Trigger Modes |
None, single, repeat, auto, rapid (segmented memory), ETS |
Advanced trigger types (real-time Mode) |
Edge, pulse width, window, window pulse width, dropout, window dropout, level, interval, logic level, runt pulse |
Trigger types (ETS Mode) |
Rising edge, falling edge |
Trigger sensitivity |
1 LSB accuracy up to full bandwidth of scope |
Trigger level |
Adjustable over whole of selected voltage range |
Maximum pre-trigger capture |
100% of capture size |
Maximum post-trigger delay |
4 billion samples |
Re-arm time |
Less than 1 µs on fastest timebase |
Maximum trigger rate |
Up to 10,000 waveforms in a 10 ms burst |
Trigger timing resolution |
1 sample period |
AUX TRIGGER INPUT |
AUX trigger connector type |
Rear panel BNC, shared with reference clock input |
Trigger types |
Edge, pulse width, dropout, interval, logic |
Input characteristics |
50 Ω ±1%, DC coupled |
Bandwidth |
25 MHz |
Threshold range |
±1 V |
Overvoltage protection |
±5 V (DC + AC peak) |
REFERENCE CLOCK INPUT (SDK ONLY) |
Clock input characteristics |
50 Ω, BNC, ±1 V, DC coupled |
Frequency range |
5, 10, 20, 25 MHz, user-selectable |
Connector |
Rear panel BNC, shared with AUX trigger |
Level |
Adjustable threshold, ±1 V |
Overvoltage protection |
±5 V |
Standard signal frequency |
DC to 20 MHz |
Standard output signals |
All models |
Sine, square, triangle, DC |
D models only |
Ramp, sinc, Gaussian, half-sine, white noise, PRBS |
Output frequency accuracy |
Same as scope timebase accuracy |
Output frequency resolution |
< 0.05 Hz |
Output voltage adjustment |
Amplitude adjustment: ±2 V (4 V max. p-p)
Offset adjustment: ±1 V
Maximum combined output voltage: ±2.5 V |
DC accuracy |
±1% of full scale |
Connector type |
Rear panel BNC |
Output impedance |
50 Ω |
Overvoltage protection |
±5 V |
Sweep modes |
Up, down, or dual, with selectable start/stop frequencies and increments |
Signal generator triggering |
Scope, manual, or AUX input; programmable number of cycles from 1 to 1 billion |
ARBITRARY WAVEFORM GENERATOR (AWG) |
Buffer size |
– |
64 kS |
– |
64 kS |
– |
64 kS |
Sample rate |
200 MS/s |
200 MS/s 2 |
200 MS/s |
Resolution |
12 bits |
12 bits |
12 bits |
Bandwidth |
20 MHz |
20 MHz |
20 MHz |
PROBE COMPENSATION OUTPUT |
Impedance |
600 Ω |
Frequency |
1 kHz square wave |
Level |
2 V pk-pk |
Overvoltage protection |
±5 V (DC + AC peak) |
SPECTRUM ANALYZER |
Frequency range |
DC to 250 MHz |
DC to 350 MHz |
DC to 500 MHz |
Display modes |
Magnitude, average, peak hold |
Windowing functions |
Rectangular, Gaussian, triangular, Blackman, Blackman-Harris, Hamming, Hann, flat-top |
Number of FFT points |
Selectable power of 2 from 128 to 1 048 576 |
MATH CHANNELS |
Functions |
-x, x+y, x-y, x*y, x/y, x^y, sqrt, exp, ln, log, abs, norm, sign, sin, cos, tan, arcsin, arccos, arctan, sinh, cosh, tanh, freq, derivative, integral, min, max, average, peak, delay |
Operands |
Input channels A to D, reference waveforms, time, p |
AUTOMATIC MEASUREMENTS |
Scope mode |
AC RMS, true RMS, cycle time, DC average, duty cycle, falling rate, fall time, frequency, high pulse width, low pulse width, maximum, minimum, peak-to-peak, rise time and rising rate |
Spectrum mode |
Frequency at peak, amplitude at peak, average amplitude at peak, total power, THD %, THD dB, THD+N, SFDR, SINAD, SNR and IMD |
Statistics |
Minimum, maximum, average, and standard deviation |
SERIAL BUS DECODING |
Data formats |
CAN, LIN, I 2 C, I 2 S, UART/RS-232, SPI, FlexRay |
MASK LIMIT TESTING |
Statisticss |
Pass/fail, failure count, total count |
DISPLAY |
Interpolation |
Linear or sin(x)/x |
Persistence modes |
Digital color, analog intensity, custom, or none |
GENERAL SPECIFICATIONS |
PC connectivity |
USB 3.0 (USB 2.0 compatible) |
Exported data formats |
Comma-separated values, tab-delimited, BMP, GIF, PNG, MATLAB 4 format |
Power requirement |
12 V DC, 4 A max. AC adaptor and cable supplied |
Dimensions (inc. connectors & end caps) |
170 x 255 x 40 mm |
170 x 285 x 40 mm |
Weight |
1 kg (approx. 2 lb 3 oz) |
1.3 kg (approx. 2 lb 14 oz) |
Temperature range |
Operating: 0 °C to 40 °C (20 °C to 30 °C for stated accuracy). Storage: -20 °C to +60 °C. |
Humidity range |
Operating: 5% to 80% RH non-condensing. Storage: 5% to 95% RH non-condensing. |
Compliance |
EU: EMC, LVD, RoHS, WEEE. USA: FCC Part 15 Subpart B |
Safety approvals |
Designed to EN 61010-1:2010 |
PC requirements |
Microsoft Windows XP, Windows Vista, Windows 7, or Windows 8 (not Windows RT) |
Software included |
PicoScope 6, Windows SDK and example programs |
Languages supported (software) |
Simplified Chinese, Traditional Chinese, Czech, Danish, Dutch, English, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Norwegian, Polish, Portuguese, Romanian, Spanish, Swedish, Turkish |
Languages supported (help) |
English, French, German, Italian, Spanish |