General Specifications |
Oscilloscope - vertical |
Input channels |
2 |
Bandwidth
±20 mV range
±10 mV range |
5 MHz
4 MHz
3 MHz |
Hardware bandwidth limiter |
200 kHz, switchable |
Rise time (calculated)
±20 mV range
±10 mV range |
70 ns
88 ns
117 ns |
Input type |
Single-ended, BNC connector |
Vertical resolution |
16 bits |
Enhanced vertical resolution |
20 bits |
Input sensitivity |
2 mV/div to 4 V/div (10 vertical divisions) |
Input ranges (full scale) |
±10 mV to ±20 V in 11 ranges |
Input coupling |
Software-selectable AC/DC |
Input characteristics |
1 MΩ ±2% || 15 pF ±2 pF |
DC accuracy (% of full scale)
±50 mV range
±20 mV range
±10 mV range |
±0.25%
±0.5%
±1%
±2% |
Overvoltage protection |
±50 V (DC + AC peak) |
Oscilloscope — horizontal |
Maximum sampling rate (real-time) |
10 MS/s |
Maximum sampling rate
(continuous streaming mode)
Using PicoScope
Using the SDK |
10 MS/s
10 MS/s one channel (PC-dependent)
6.7 MS/s two channels (PC-dependent) |
Timebase ranges |
1 µs/div to 5000 s/div |
Buffer memory |
16 MS shared between active channels |
Streaming buffer memory
(PicoScope) |
100 MS (shared) |
Streaming buffer memory (SDK) |
Up to available PC memory |
Maximum buffer segments
((PicoScope) |
10,000 |
Maximum buffer segments (SDK) |
32,768 |
Maximum waveforms per second |
9,100 (typical) |
Timebase accuracy |
±50 ppm |
Sample jitter |
< 10 ps RMS |
Dynamic performance (typical) |
Crosstalk |
> 50,000:1 |
Harmonic distortion |
-95 dB typical @ 10 kHz, -1 dBfs input |
SFDR |
102 dB typical @ 10 kHz, -1 dBfs input |
Noise |
8.5 µV RMS (on most sensitive range) |
Bandwidth flatness |
±0.2 dB (DC to full bandwidth) |
Pulse response |
< 1% overshoot on all ranges |
Triggering (general) |
Trigger modes |
Free run, auto, repeat, single, rapid (segmented memory) |
Maximum pre-trigger capture |
Up to 100% of capture size |
Maximum post-trigger delay |
Up to 4 billion samples |
Trigger rearm time |
< 10 µs on fastest timebase |
Triggering (main inputs) |
Source |
Ch A, Ch B |
Trigger types |
Rising, falling, adjustable hysteresis, window, pulse width, window
pulse width, dropout, window dropout, interval, logic, runt pulse |
Trigger level |
Adjustable over whole of selected voltage range |
Trigger sensitivity |
Digital triggering provides 1 LSB accuracy up to full bandwidth of scope |
Maximum pre-trigger capture |
Up to 100% of capture size |
Maximum post-trigger delay |
Up to 4 billion samples |
Trigger rearm time |
< 10 µs on fastest timebase |
External trigger input |
Trigger types |
Edge, pulse width, dropout, interval, logic, delayed |
Input characteristics |
1 MΩ ±2% || 15 pF ±2 pF |
Input type |
Rear-panel BNC connector |
Threshold range |
±5 V and ±500 mV, DC coupled |
Sensitivity |
25 mV p-p at 1 MHz, typical |
DC accuracy |
±1% of full scale |
Bandwidth |
5 MHz |
Overvoltage protection |
±50 V |
Signal generator |
Standard output signals |
Sine, square, triangle, DC voltage, ramp, sinc, Gaussian, half-sine,
white noise, PRBS |
Standard signal frequency |
DC to 20 kHz |
Sweep modes |
Up, down, dual |
Frequency accuracy |
±50 ppm |
Frequency resolution |
< 0.01 Hz |
Voltage range |
±1 V (no load) |
Amplitude and offset adjustment |
100 µV steps (within overall ±1 V range) |
Amplitude flatness |
< 0.1 dB to 20 kHz, typical |
DC accuracy |
±0.5% of full scale |
SFDR |
102 dB typical @ 10 kHz, -1 dBfs input |
Connector type |
Front-panel BNC |
Output characteristics |
600 Ω |
Overvoltage protection |
±10 V |
Arbitrary Waveform Generator |
Update rate |
192 kS/s |
Buffer size |
4,096 kS |
Resolution |
16 bits |
Bandwidth |
20 kHz |
Rise time (10% to 90%) |
11 µs, typical |
Spectrum analyzer |
Frequency range |
DC to 5 MHz |
Display modes |
Magnitude, peak hold, average |
Windowing functions |
Rectangular, Gaussian, triangular, Blackman, Blackman-Harris,
Hamming, Hann, flat-top |
Number of FFT points |
Selectable from 128 to 1 million in powers of 2 |
Scale / units |
X axis : linear or log 10
Y axis : logarithmic (dbV, dBu, dBm, arbitrary) or linear (volts) |
Math channels |
Functions |
−x, x+y, x−y, x*y, x/y, x^y, sqrt, exp, ln, log, abs, norm, sign, sin, cos,
tan, arcsin, arccos, arctan, sinh, cosh, tanh, freq, derivative, integral,
min, max, average, peak, delay |
Operands |
A, B (input channels), T (time), reference waveforms, constants, Pi |
Automatic measurements |
Oscilloscope mode |
AC RMS, true RMS, cycle time, DC average, duty cycle, falling rate, fall
time, frequency, high pulse width, low pulse width, maximum, minimum,
peak to peak, rise time, rising rate. |
Spectrum mode |
Frequency at peak, amplitude at peak, average amplitude at peak,
total power, THD %, THD dB, THD plus noise, SFDR, SINAD, SNR, IMD |
Statistics |
Minimum, maximum, average and standard deviation |
Serial decoding |
Protocols |
CAN, LIN, I²C, UART/RS-232, SPI, I²S, FlexRay |
Inputs |
All input channels with any mixture of protocols |
Mask limit testing |
Mask generation |
Auto generate from captured waveform, manual drawing, manual
coordinate entry |
Actions |
Highlight on screen, select in buffer overview, activate alarm |
Statistics |
Pass/fail, failure count, total count |
Alarms |
Initiating events |
Capture, buffer full, mask fail |
Alarm actions |
Beep, play sound, stop/restart capture, run executable, save current
buffer/all buffers, trigger signal generator |
Display |
Interpolation |
Linear or sin(x)/x |
Persistence modes |
Digital color, analog intensity, fast, custom |
Data export |
Output file formats |
BMP, CSV, GIF, JPG, MATLAB 4, PDF, PNG, PicoScope data, PicoScope
settings, TXT |
Output functions |
Copy to clipboard, print |
Environmental |
Temperature range (operating) |
0 °C to 45 °C |
Temperature range (stated
accuracy) |
20 °C to 30 °C |
Temperature range (storage) |
-20 °C to +60 °C |
Humidity range (operating) |
5% to 80% RH non-condensing |
Humidity range (storage) |
5% to 95% RH non-condensing |
Physical properties |
Dimensions (including connectors) |
210 x 135 x 40 mm (approx 8.2 x 5.3 x 1.5 in) |
Weight |
< 500 g (approx 1.1 lb) |
Software |
Software included |
PicoScope 6 for Windows oscilloscope software |
Optional free software |
PicoScope 6 Beta for Linux and OS X. Drivers for Linux and OS X. |
Languages |
Chinese (simplified), Chinese (traditional), Czech, Danish, Dutch,
English, Finnish, French, German, Greek, Hungarian, Italian, Japanese,
Korean, Norwegian, Polish, Portuguese, Romanian, Russian, Spanish,
Swedish, Turkish |
General |
Additional hardware (supplied) |
2 x MI007 probes, USB 2.0 cable, user manuals, software CD-ROM |
PC requirements |
Microsoft Windows XP (SP3), Vista, Windows 7 or Windows 8 (not
Windows RT). |
PC connectivity |
USB 2.0 (compatible with USB 1.1 and USB 3.0; USB 1.1 not
recommended) |
Power requirements |
Powered from USB port |
Safety approvals |
European LVD standard |
EMC approvals |
European EMC standard |
Environmental approvals |
RoHS compliant |
Total satisfaction guarantee |
In the event that this product does not fully meet your requirements
you can return it for an exchange or refund. To claim, the product must
be returned in good condition within 14 days. |
Warranty |
5 years |