Oscilloscope — vertical (analog channels) |
Input channels |
2 |
Bandwidth |
25 MHz |
Rise time (calculated) |
14 ns |
Input type |
Single–ended, BNC connector |
Vertical resolution |
8 bits |
Enhanced vertical resolution |
12 bits |
Input sensitivity |
10 mV/div to 4 V/div (10 vertical divisions) |
Input ranges (full scale) |
±50 mV to ±20 V in 9 ranges |
Input coupling |
Software selectable AC/DC |
Input characteristics |
1 MΩ ±1% in parallel with 14 pF ±2 pF |
DC accuracy |
±3% of full scale |
Overvoltage protection |
±100 V (DC + AC peak) |
Oscilloscope — vertical (digital channels) |
Input channels |
16 channels (2 ports of 8 channels each) |
Input connectors |
2.54 mm pitch, 10 x 2 way connector |
Maximum input frequency |
100 MHz |
Input impedance (with TA136 cable) |
200 kΩ ±2% in parallel with 8 pF ±2 pF |
Digital threshold range |
±5 V |
Input range |
±20 V |
Overvoltage protection |
±50 V |
Threshold grouping |
Two independent threshold controls, one per port |
Threshold selection |
TTL, CMOS, ECL, PECL, user–defined |
Threshold accuracy |
±100 mV |
Minimum input voltage swing |
500 mV |
Channel–to–channel skew |
< 5 ns |
Minimum input slew rate |
10 V/µs |
Oscilloscope — horizontal |
Maximum sampling rate (real-time)
Ch A only
CH A + 1 digital port
1 or 2 digital ports
All other combinations |
200 MS/s
200 MS/s
200 MS/s
100 MS/s |
Maximum equivalent sampling rate (repetitive signals) |
4 GS/s |
Maximum sampling rate (continuous streaming mode) |
10 MS/s on all scope channels and digital ports in PicoScope 6 > 20 MS/s using supplied SDK (PC-dependent) |
Timebase ranges |
2 ns to 5000 s/div |
Buffer memory |
48 kS shared between active channels and ports |
Streaming buffer memory (PicoScope) |
100 MS |
Streaming buffer memory (SDK) |
Up to available PC memory |
Maximum buffer segments (PicoScope)
Normal triggering
Rapid triggering |
10,000 software segments
32 hardware segments |
Maximum buffer segments (SDK) |
32 hardware segments |
Timebase accuracy |
±100 ppm |
Sample jitter |
< 300 ps RMS |
Analog channels - dynamic performance (typical) |
Crosstalk |
Better than 200:1 up to full bandwidth (equal voltage ranges) |
Harmonic distortion |
< -55 dB at 100 kHz, full–scale input |
SFDR |
> 55 dB at 100 kHz, full–scale input |
Noise |
≤ 3 counts (all ranges) |
Bandwidth flatness (at scope input) |
(+0.3 dB, -3 dB) from DC to full bandwidth |
Pulse response |
< 7% overshoot |
Linearity |
≤ 1 LSB |
Triggering (all inputs) |
Trigger modes |
Free run, auto, repeat, single, rapid (segmented memory) |
Maximum pre–trigger capture |
100% of capture size |
Maximum post–trigger delay |
4 billion samples |
Trigger rearm time |
< 2 µs on fastest timebase |
Maximum trigger rate |
32 waveforms in a 100 µs burst |
Triggering (analog inputs) |
Source |
Ch A, Ch B |
Trigger types |
Edge, window, pulse width, window pulse width, dropout, window dropout, interval, runt pulse, logic |
Trigger level |
Adjustable over whole of selected voltage range |
Trigger sensitivity (real-time) |
Digital triggering provides 1 LSB accuracy up to full bandwidth of scope |
Trigger types (ETS mode) |
Rising edge, falling edge |
Trigger sensitivity (ETS mode) |
Typical 10 mV p–p at full bandwidth |
Triggering (digital inputs) |
Source |
D0 to D15 |
Trigger types |
Combined level and edge |
Trigger levels |
Uses port threshold setting |
Advanced triggers |
Data pattern (can be grouped by user) |
Function generator |
Standard output signals |
Sine, square, triangle, DC voltage, ramp, sinc, Gaussian, half–sine, white noise |
Standard signal frequency |
DC to 100 kHz |
Sweep modes |
Up, down, dual |
Frequency accuracy |
As oscilloscope timebase |
Frequency resolution |
< 0.01 Hz |
Voltage range |
±2 V |
Amplitude and offset adjustment |
1 mV steps within overall ±2 V range |
Amplitude flatness |
< 1 dB to 100 kHz typical |
DC accuracy |
±1% of full scale |
SFDR |
> 55 dB @ 1 kHz full–scale sine wave |
Connector type |
Rear panel BNC |
Output characteristics |
600 Ω |
Overvoltage protection |
±10 V |
Arbitrary waveform generator |
Update rate |
2 MS/s |
Buffer size |
8 kS |
Resolution |
12 bits |
Bandwidth |
100 kHz |
Rise time (10% to 90%) |
< 2 µs |
Math channels |
Functions |
−x, x+y, x−y, x*y, x/y, x^y, sqrt, exp, ln, log, abs, norm, sign, sin, cos, tan, arcsin, arccos, arctan, sinh, cosh, tanh, freq, derivative, integral, min, max, average, peak, delay |
Operands |
A, B (input channels), T (time), reference waveforms, constants, Pi |
Automatic measurements (analog channels only) |
Oscilloscope mode |
AC RMS, true RMS, cycle time, DC average, duty cycle, falling rate, fall time, frequency, high pulse width, low pulse width, maximum, minimum, peak to peak, rise time, rising rate. |
Spectrum mode |
Frequency at peak, amplitude at peak, average amplitude at peak, total power, THD %, THD dB, THD plus noise, SFDR, SINAD, SNR, IMD |
Statistics |
Minimum, maximum, average and standard deviation |
Serial decoding |
Protocols |
CAN, LIN, I²C, UART/RS-232, SPI, I²S, FlexRay |
Inputs |
All input channels with any mixture of protocols |
Mask limit testing |
Mask generation |
Auto generate from captured waveform, manual drawing, manual coordinate entry |
Actions |
Highlight on screen, select in buffer overview, activate alarm |
Statistics |
Pass/fail, failure count, total count |
Alarms |
Initiating events |
Capture, buffer full, mask fail |
Alarm actions |
Beep, play sound, stop/restart capture, run executable, save current buffer/all buffers, trigger signal generator |
Display |
Interpolation |
Linear or sin(x)/x |
Persistence modes |
Digital color, analog intensity, fast, custom |
Data export |
Output file formats |
BMP, CSV, GIF, JPG, MATLAB 4, PDF, PNG, PicoScope data, PicoScope settings, TXT |
Output functions |
Copy to clipboard, print |
Environmental |
Temperature range (operating) |
0 °C to 50 °C |
Temperature range (stated accuracy) |
20 °C to 30 °C |
Temperature range (storage) |
–20 °C to +60 °C |
Humidity range (operating) |
5% to 80% RH non-condensing |
Humidity range (storage) |
5% to 95% RH non-condensing |
Physical properties |
Dimensions (including connectors) |
200 x 140 x 40 mm (approx 7.9 x 5.5 x 1.6 in) |
Weight |
<500 g (approx 17.6 oz) |
Software |
Software included |
PicoScope 6 for Windows oscilloscope software |
Optional free software |
PicoScope 6 Beta for Linux and OS X. Drivers for Linux and OS X. |
Languages |
Chinese (simplified), Chinese (traditional), Czech, Danish, Dutch, English, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Norwegian, Polish, Portuguese, Romanian, Russian, Spanish, Swedish, Turkish |
General |
Additional hardware (supplied) |
USB 2.0 cable, user manuals, software CD–ROM |
PC requirements |
Microsoft Windows XP (SP3), Vista, Windows 7 or Windows 8 (not Windows RT). |
PC connectivity |
USB 2.0 (USB 1.1 and USB 3.0 compatible) |
Power requirements |
Powered from USB port |
Safety approvals |
Designed to EN 61010–1:2010 |
EMC approvals
CE
FCC |
Tested to EN61326–1:2006
Tested to part 15 subpart B |
Environmental approvals |
RoHS and WEEE compliant |
Total satisfaction guarantee |
In the event that this product does not fully meet your requirements you can return it for an exchange or refund. To claim, the product must be returned in good condition within 14 days. |
Warranty |
5 years |