Model |
PicoScope 2104 |
PicoScope 2105 |
Input channels |
1 |
Input type |
Built-in oscilloscope probe |
Bandwidth |
10 MHz |
25 MHz |
Rise time (calculated) |
35 ns |
14 ns |
Vertical resolution |
8 bits |
Enhanced vertical resolution |
12 bits |
Input sensitivity |
20 mV/div to 4 V/div (10 vertical divisions) |
Input ranges (full scale) |
±100 mV to ±20 V in 8 ranges |
Input coupling |
Software selectable AC/DC |
Input characteristics |
1 MΩ in parallel with 20 pF |
DC accuracy |
±3% |
Linearity |
< 1 LSB at 25 °C |
Overvoltage protection |
±50 V (DC + AC peak) |
Maximum sampling rate (real-time) |
50 MS/s |
100 MS/s |
Maximum sampling rate (repetitive
signals, ETS mode) |
1 GS/s |
2 GS/s |
Maximum sampling rate (continuous
streaming mode) |
10 MS/s in PicoScope 6 |
Timebase ranges |
200 ns/div to 5000 s/div
(10 ns/div to 5000 s/div with ETS) |
100 ns/div to 5000 s/div
(5 ns/div to 5000 s/div with ETS) |
Buffer memory# |
8 kS |
24 kS |
Maximum buffer segments |
10,000 |
Maximum waveforms per second |
180 wfm/s (PC-dependent) |
# Shared between active channels |
Source |
Main input |
Trigger modes |
None, auto, repeat, single, rapid (segmented memory) |
Digital triggers |
Rising edge, falling edge |
Maximum pre–trigger capture |
Up to 100% of capture size |
Maximum post–trigger delay |
Up to 4 billion samples (selectable in 1 sample steps) |
Maximum trigger rate |
150 wfm/s |
Trigger level |
Adjustable over whole of selected voltage range |
Trigger sensitivity |
Digital triggering provides 1 LSB accuracy up to full bandwidth of scope |
Frequency range |
DC to maximum bandwidth of scope |
Display modes |
Magnitude, peak hold, average |
Windowing functions |
Rectangular, Gaussian, triangular, Blackman, Blackman–Harris,
Hamming, Hann, flat-top |
Number of FFT points |
Selectable from 128 to 1 million in powers of 2 |
X axis |
Linear or log 10 |
Y axis |
Logarithmic (dbV, dBu, dBm, arbitrary) or linear (volts) |
General functions |
−x, x+y, x−y, x*y, x/y, x^y, sqrt, exp, ln, log, abs, norm, sign, sin, cos,
tan, arcsin, arccos, arctan, sinh, cosh, tanh, derivative, integral, delay |
Filter functions |
Low pass, high pass, band stop, band pass |
Graphing functions |
Frequency, duty cycle |
Multi-waveform functions |
Min, max, average, peak |
Operands |
Input channel, reference waveforms, time, constants, pi |
Oscilloscope mode |
AC RMS, true RMS, cycle time, DC average, duty cycle, falling rate, fall
time, frequency, high pulse width, low pulse width, maximum, minimum,
peak to peak, rise time, rising rate |
Spectrum mode |
Frequency at peak, amplitude at peak, average amplitude at peak, total
power, THD %, THD dB, THD+N, SFDR, SINAD, SNR, IMD |
Statistics |
Minimum, maximum, average, standard deviation |
Mask generation |
Auto generate from captured waveform, manual drawing, manual
coordinate entry |
Actions |
Highlight on screen, select in buffer overview, activate alarm |
Statistics |
Pass/fail, failure count, total count |
Initiating events |
Capture, buffer full, mask fail |
Alarm actions |
Beep, play sound, stop/restart capture, run executable, save current
buffer/all buffers, trigger signal generator |
Interpolation |
Linear or sin(x)/x |
Persistence modes |
Digital color, analog intensity, fast, advanced |
Output file formats |
BMP, CSV, GIF, JPG, MATLAB 4, PDF, PNG, PicoScope data, PicoScope
settings, TXT |
Output functions |
Copy to clipboard, print |
Temperature range (operating) |
0 °C to 45 °C |
Temperature range (stated
accuracy)
|
20 °C to 30 °C |
Dimensions
Probe
Probe and cable
Diameter |
220 mm (8.6 in) long
3 m (9.8 ft) long
32 mm (1.2 in) |
Software included |
PicoScope 6 for Windows oscilloscope software
PicoLog for Windows data logging software |
Optional free software |
PicoScope 6 Beta for Linux and OS X. Drivers for Linux and OS X. |
Languages |
Chinese (simplified), Chinese (traditional), Czech, Danish, Dutch,
English, Finnish, French, German, Greek, Hungarian, Italian, Japanese,
Korean, Norwegian, Polish, Portuguese, Romanian, Russian, Spanish,
Swedish, Turkish |
Additional items supplied |
Quick start guide, software CD-ROM, accessories kit (ground clip, spring
hook, BNC adaptor, insulating cap, IC probing cap, replacement probe
tip) |
PC requirements |
Microsoft Windows XP (SP3), Vista, Windows 7 or Windows 8 (not
Windows RT). |
PC connectivity |
USB 2.0 port (compatible with USB 1.1 and USB 3.0; USB 1.1 not
recommended) |
Power requirements |
Powered from USB port |
Compliance |
European EMC and LVD standards
FCC Rules Part 15 Class A |
Total satisfaction guarantee |
In the event that this product does not fully meet your requirements
you can return it for an exchange or refund. To claim, the product must
be returned in good condition within 14 days. |
Warranty |
5 years |