Operation Frequency |
During R measurement, operator should select the following test frequencies: 270 Hz ± 1 Hz or 1470 Hz ± 1 Hz |
Voltmeter |
In the voltmeter function, the equipment operates as a CA VoLTMETER conventional voltmeter, making it possible to measure voltages generated by parasitic currents. |
Measurement Ranges |
Resistance: 0 - 20 kΩ (autoranging)
Resistivity: 0 - 50 kΩm (autoranging)
Voltage: 0 - 60 V |
Accuracy |
Resistance and Resistivity measurements:
R ≤ 2 kΩ: ± (2% of the measured value ± 2 digits)
R > 2 kΩ: ± (5% of the measured value ± 2 digits)
Voltage measurement: ± (3% of the measured value ± 2 digits) |
Reading Resolution |
0.01 Ω in the resistance measurement
0.01 Ωm in the resistivity measurement
0.1 V in the voltage measurement |
Output Current |
The short-circuit current is limited to less than 3.5 mA RMS (according to IEC 61557-5 - 4.5) |
Immunity to Spurious Voltage Interference |
During the R measurement, it allows the presence of spurious voltage up to 7 V. |
Earth Resistance of Auxiliary Rods |
In the R measurement it allows Raux up to 50 kΩ with error <30%. |
Battery Status Chec |
The battery charge status is verified under normal use conditions. |
Advanced Features |
Automatic detection of abnormal conditions that may cause excessive errors (low battery, too high noise interference, too high test spike resistance). |
Soil Resistivity Computing |
When performing soil resistivity measurements, the operator provides the distance between spikes and the equipment automatically computes soil resistivity using the Wenner method. |
Data Output |
USB interface |
Built-In Printer (optional) |
Print-out of measured values |
Power Supply |
Internal rechargeable battery, 12 V - 2.3 Ah |
Battery Charger |
For 100-240 V mains supply |
Safety Class |
Meets IEC 61010-1:1990,IEC 61010-1:1992 amendment 2. |
E.M.C |
In accordance with IEC 61326-1 |
Electrostatic Immunity |
In accordance with IEC 1000-4-2 |
Electro Magnetic Irradiation Immunity |
In accordance with IEC 61000-4-3 |
Environmental Protection |
IP54 (with closed lid) |
Operating Temperature |
-10°C to 50°C |
Storage Temperature |
-25°C to 65°C |
Humidity |
95% RH (without condensation) |
Dimensions (approx.) |
11” (274 mm) L x 10” (250 mm) W x 5” ( 124 mm) H |
Weight (approx.) |
10 lbs ( 3.6 kg) |