General Specifications: (Unless otherwise indicated all specifications are rated from a nominal 23 °C, 70 % RH for 1 year from calibration) |
Temperature Range |
-25 to 60 °C (-5 to 140 °F) |
Storage Temperature Range |
-30 to 60 °C (-22 to 140 °F) |
Relative Humidity Range |
10 % =RH =90 % (0 to 35 °C), Non-condensing
10 % =RH= 70 % (35 to 60 °C), Non-condensing |
Size |
7.00 X 3.30 X 2.21 inches (177.8 x 83.8 x 56.1mm) |
Weight |
12.0 oz (340 grams) |
Battery |
9V Alkaline
Optional 120 VAC 50/60 Hz AC adaptor available |
Miscellaneous |
Low battery indication with nominal 1 hour of operation left
Over-voltage protection to 120 Vrms (rated for 30 seconds) or 240 Vrms (rated for 15 seconds)
Bar graph display with 1% resolution of 4-20 mA signal scale
High contrast graphic liquid crystal display with 0.45 (11.4 mm) high digits |
Common Specifications for all current modes: |
Ranges |
0.000 to 24.000 mA, -25.00 to 125.00% of 4-20 mA |
Accuracy |
< ± (0.012 % of Reading + 0.004 mA) |
Temperature effect |
< ± 50 ppm/°C of Range |
Resolution(s) |
0.001 mA and 0.01 % |
Source/Power and Measure 2-Wire Transmitter Specifications: |
Cold Junction Uncertainty |
±0.25°C (0.5°F) |
Loop compliance voltage |
= 24 Volts Read |
Loop drive capability |
1200 Ω at 20 mA for entire battery life |
Miscellaneous |
Open loop or out of compliance conditions are indicated by appropriate error display
Battery Life in:
Source mode = 18 hrs at 12mA typical (HART disabled)
Power measure = 10 hrs at 12mA typical
HART protocol mode is a selectable option at turn on. HART protocol mode places a 250? resistor in series with the output
Selectable EZ-Step(s) for Source Mode/2-Wire Transmitter Simulation:
In mA mode: 0.001, 0.010, 0.100, 1.000, 4.000(default), 8.000 mA
% of 4-20 mA mode: 0.01, 0.10, 1.00, 10.00, 25.00(default), 50.00 % |
Read mA Specifications: |
Voltage burden |
< 2V at 20 mA |
Overload/Current limit protection |
24 mA nominal |
Battery life |
Typical >40 Hours |