| General Specifications |
| Operating Temperature Range |
-25 to 60 °C (-10 to 140 °F) |
| Storage Temp. Range |
-30 to 60 °C (-22 to 140 °F) |
| Relative Humidity Range |
10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing |
| 10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing |
| Temperature Drift |
± 0.01% of span outside of 23°C ±10 °C (73°C ±18 °F) |
| Size |
5.63 x 3.00 x 1.60 in, 143 x 76 x 41 mm (L x W x H) |
| Weight |
12.1 ounces, 0.34 kg (including boot & batteries) |
| Batteries |
Four “AA” Alkaline 1.5V (LR6) |
| Battery Life (Typical) |
25 Hours Thermocouple or RTD, 8 Hours milliamp |
| Isolation |
Voltage |
60V rms between all milliamp functions/Read V DC and Source V DC/Thermocouple/Ohms/RTD |
| Normal Mode Rejection |
50/60 Hz, 50 dB |
| Common Mode Rejection |
50/60 Hz, 120 dB |
| Optional NiMh Rechargeable battery kit |
Charger, four NiMh batteries [Part # 020-0103] |
| Low Battery |
Low battery indication with nominal 1 hour of operation left |
| Protection against misconnection |
Over-voltage protection to 60 V dc (rated for 30 seconds) |
| Display |
High contrast graphic liquid crystal display. LED backlighting for use in low lit areas. |
| Read mA |
| Ranges and Resolution |
0.000 to 24.000 mA or -25.00 to 125.00% of 4-20 mA |
| Accuracy |
≤ ± (0.02 % of Reading + 0.003 mA) |
| Voltage burden |
≤ 2V at 24 mA |
| Overload/Current limit protection |
25 mA nominal |
| Source mA / Power & Measure Two Wire Transmitters & PWRM LEAK |
| Ranges and Resolution |
0.000 to 24.000 mA or -25.00 to 125.00% of 4-20 mA |
| Accuracy |
≤ ± (0.02 % of Reading + 0.003 mA) |
| Loop compliance voltage |
≥ 24 DCV @ 20.00mA |
| Loop drive capability |
1200 Ω at 20 mA
950 Ω with Hart Resistor or leak detection running |
| mA 2-Wire Transmitter Simulation |
| Accuracy |
Same as Source/Power & Measure |
| Voltage burden |
≤ 2V at 20 mA |
| Overload/Current limit protection |
24 mA nominal |
| Loop voltage limits |
2 to 60 VDC (fuse-less protected from reverse polarity connections) |
| Source Thermocouple |
| Accuracy |
±(0.008% of Reading + 0.006 mV) |
| Cold Junction Compensation |
± 0.09°F (±0.05 °C) - Thermistor traceable to NIST for 11 years |
| Millivolt Range |
-13.000 to 80.000 mV |
| Output Impedance |
< 0.3 Ohms |
| Source Current |
> 20 mA (drives 80 mV into 10 Ohms) |
| Noise |
≤ 4 microvolts p-p for frequencies of 10 Hz or below |
| Read Thermocouple |
| Accuracy |
±(0.008% of Reading + 0.006 mV) |
| Cold Junction Compensation |
± 0.09°F (±0.05 °C) - Thermistor traceable to NIST for 11 years |
| Millivolt Range |
-13.000 to 80.000 mV |
| Input Impedance |
> 10 Megohms |
| Open Thermocouple |
Threshold |
18,000 Ohms nominal |
| Pulse |
< 10 microamp pulse for 400 milliseconds |
| Source Ohms & RTD |
| 3 & 4 Wire Accuracy |
From 1 to 10.2 mA External Excitation Current |
±(0.015% of Reading + 0.05 Ohms) |
| Below 1 mA of External Excitation Current |
|
| 2 Wire Accuracy* |
Add 0.1 Ohms to 3 Wire & 4 Wire Accuracy |
| Resistance Ranges |
400 Ohm Range: 0.00 to 401.00 & 0.000 to 401.000
4000 Ohm Range: 0.0 to 4010.0 & 0.00 to 4010.00 |
| MS Noise |
00 Ohm Range: ≤ ± 0.005 Ohms from 0.1 to 10 Hz
4000 Ohm Range: ≤ ± 0.05 Ohms from 0.1 to 10 Hz |
| Allowable Excitation Current Range |
0 to 400 Ohm: 10.2 mA max; steady or pulsed/intermittent
401 to 4000 Ohms: 1 mA max; steady or pulsed/intermittent |
| Pulsed Excitation Current Compatibility |
DC to 0.01 second pulse width |
| Read Ohms & RTD |
| 3 Wire & 4 Wire Accuracy |
±(0.015% of Reading + 0.05 Ohms) |
| 2 Wire Accuracy* |
(0.015% of Reading + 0.15 Ohms) |
| Resistance Ranges |
Same as Source Ohms & RTD |
| Excitation Current |
0.9 mA to 401 Ohms, 0.4 mA to 4010 Ohms (nominal) |
| * 2 Wire specifications do not include resistance of the lead wires |