Optical resolution |
640 x 480 pixel |
Detector |
FPA, uncooled (17 µm x 17 µm) |
Spectral range |
8 - 14 µm |
Temperature ranges |
-20 ... 100 °C (-4 ... 212 °F)
0 ... 250 °C (32 ... 482 °F)
(20) 150 ... 900 °C ([68] 302 ... 1652 °F)1)
optional temperature range: 200 ... 1500 °C (392 ... 2732 °F) |
Frame rate |
32 Hz / 125 Hz @ 640 x 120 pixels |
Optics (FOV) |
60° x 45° FOV/f=10.5 mm (f=0.4 in) |
Thermal sensitivity (NETD) |
75 mK |
Accuracy |
±2 °C or ±2 % (±3.6 °F or ±2 %), whichever is greater |
PC interface |
USB 2.0 / optional USB to Gig" (PoE) interface |
Process interface (PIF), standard |
0 - 10 V input, digital input (max. 24 V), 0 - 10 V output |
Process interface (PIF), industrial |
2x 0 - 10 V input, digital input (max. 24 V),
3x 0/4 - 20 mA outputs, 3x relay (0 - 30 V/ 400 mA), fail-safe relay |
Cable length (USB) |
1 (standard) / 5 m / 10 m / 20 m (3.3 / 16.4 / 32.8 / 65.6 ft)
5 and 10 m (16.4 and 32.8 ft) also as HT cable (180 or 250 °C [356 or 482 °F]) |
Ambient temperature |
0 °C ... 50 °C (32 °F ... 122 °F) |
Storage temperature |
- 40 ... 70 °C (-40 ... 158 °F) |
Relative humidity |
20 - 80 %, non-condensing |
Enclosure (size / rating) |
46 x 56 x 76 - 100 mm (1.8 x 2.2 x 3.0 - 3.9 in) (depending on lens + focus position) / IP 67 (NEMA 4) |
Weight |
269 - 340 g (10.59 - 13.38 oz), depending on lens |
Shock / Vibration 2) |
IEC 60068-2-27 (25G and 50G) /
IEC 60068-2-6 (sinus shaped), IEC 60068-2-64 (broadband noise |
Tripod mount |
1/4- 20 UNC |
Power supply |
USB powered |
1) Accuracy effective starting at 150° C (302 °F)
2) For more details see operator's manual |