General Specifications |
Record |
Sample Rates: User configured 1 every 2 seconds to 1 every 24 hours
No. of Samples: 130,000 (depends on setup) |
Display |
Two line, 6 digit 7-segment LCD and 5 digit alpha numeric display with custom icons, multi-segment bargraph for 0-100% indication, backlit |
Record Trigger |
Two independent triggers. Multiple trigger modes: Instantaneous, button control, on alarm, time and date (start and stop), day of week |
Record Mode |
Fill to end of memory or cyclic, number of samples and time duration |
Alarms |
2 user programmable alarms. High or Low |
Communication |
Mini USB connection |
Software |
Track-It™ Software - Program device, view data (historic or real time), export to Excel™. Simple and advanced modes |
Battery |
Lithium 1/2AA 1.2Ah
Life: Up to 3 years typical @ 1 minute sample rate |
Enclosure |
Material: Anodized Aluminum and 316L Stainless Steel
Dimensions: 2.94” dia x 1.82” deep x 4.78” tall (including port threads) |
Measurement |
Pressure Ranges |
0-3, 0-35, 0-150, 0-350, 0-550, 0-2000, 0-5800, 0-8700, or 0-14,500 PSI |
Vacuum Ranges |
0 to -380 and 0 to -760 torr, 760 to 0 torr and 760 to -760 torr |
Accuracy |
± 0.25% |
Resolution |
0.01% |
Overpressure Rating |
1.5 Times Max |
Port Connection |
1/4” NPT Male (1/2" NPT Male for 8700 and 14,500 ranges) |
Temperature Range |
-20 to +85°C / 0 to 185°F |
Accuracy |
±1°C 0 to 50°C / 32 to 122°F
±2°C –20 to 85°C / 0 to 185 F |
Resolution |
0.06°C / 0.11°F |