VMA0055 Voltage, Current Measure & Simulate and 24 V DC
Isolated Loop Power for MFT 4000 Series Calibrator
The VMA0055 is a DC voltage and current measurement and simulation module for use with the Meriam MFT 4000 Series Multi-Function Tester. The module features the functionality of a DVM and loop calibrator for terrific versatility. When used with other MFT functions, the VMA / MFT combination provides valuable functionality not available in other devices. Measurement ranges include ±500 mV, ±55 V, and ±100 mA. Sourcing capabilities include ±1000 mV, 0 - 24 V and 0 - 22 mA. The VMA module will also simulate transmitter outputs when connected to a loop or it can be used to temporarily supply 24V DC loop power to unpowered devices during commissioning, outages or calibration activities.
VMA0055 goes far beyond the ±50 volts / mA measurement capability of the MFT base unit and has its own isolated power supply so common ground problems are not an issue. Consider the following advantages for MFT users:
- Power a transmitter directly from VMA
Use MFT 4000 and 4010 to calibrate transmitters anywhere, anytime
Concurrently use MFT base unit V / mA meter to simultaneously check transmitter output
Use MFT 4010 HART communications to commission or reconfigure unpowered transmitters
- Measure mV, V or mA as you would with a DVM
Reduce the number of tools the technician carries
Resolution to 1/100,000 counts
Select desired V range for best possible accuracy
- Source precise voltage or mA signals to check calibration or operation of receiving devices or step a positioner
Key in exact output value desired
Increase or Decrease output value by predefined increments
Auto-step tests are user defined for start and stop values, number of steps, dwell time, and manual or automatic advance
- Read volts or mA from MFT meter while sourcing volts or mA from VMA module
Test input isolation buffers with one MFT; apply mA value to buffer input with VMA and measure mA value output of buffer
Select mA or V measurement on MFT meter from the VMA output display without interruption of output
- Simulate transmitter output in externally powered loop VMA modulates loop current to any selected value
Increase or Decrease simulated value by predefined increments
Auto-step tests are user defined for start and stop values, number of steps, dwell time, and manual or automatic advance
- I/P Device and switch testing
Combined with a pressure module, perform I/P testing
Use VMA power source to wet switch contacts and use pressure and temperature modules for pressure and temperature switch testing
- Available certifications include:
CE Mark
NIST traceable - standard
Intrinsically Safe - MET Laboratories per CSA C22.2 & UL 913 Class I Division I, Groups A, B, C & D
- Approvals
Intrinsically Safe Certification from MET Laboratories. MET Laboratories is a Nationally Recognized Test Lab (NRTL). The module is certified by MET Laboratories to comply with standards CSA C22.2 No.1010.1, CSA C22.2 No. 157-92, and UL913 Fifth Edition Rev 2/21/97 for Class I, Division I, Groups A, B, C, and D, T5 hazardous areas. Restrictions apply - see the Intrinsically Safe Control Drawing shipped with each MFT Modular Calibrator for more details.
- Resolution, Range, and Accuracy Specification for VMA0055 Module
Ambient temperature limits: -10° C to +50° C
Accuracy statements are for ambient temperatures of 18° C to 28° C
Apply the Temperature Coefficient for ambient temperatures below 18° C and above 28° C
|