| DC Current (mA) Measure and Source Modes |
| Specification |
Measure Mode |
Source Mode (No External Power) |
| Resolution |
.001 mA |
.001 mA |
| Range |
±100.000 mA (55 Vdc compliance) |
0 - 22.000 mA (See Note 3 below) |
| Accuracy |
± (0.01% of reading + 0.015% FS) |
± (0.01% of reading + 0.015% FS) |
| Open Circuit Voltage |
|
24 VDC |
| Output Drive |
|
15 VDC minimum @ 24 mA, Resistive load |
| Temperature Coefficient |
± (0.001% Rdg +0.002% FS) / °C |
± (0.003% Rdg +0.003% FS) / °C |
| Volts DC Measure and Source Modes |
| Specification |
Measure Mode |
Source Mode (No External Power) |
| Resolution |
1/100,000 counts: .001 mV; .001 V |
1/100,000 |
| Range |
500 mV; 1, 2, 4, 8, 15, 30, 55 V |
0 - 24.000 VDC |
| Accuracy |
± (0.025% of reading + 0.005% FS) |
± (0.01% of reading + 0.05% FS) |
| Open Circuit Voltage |
|
24 VDC |
| Output Drive |
|
15 VDC minimum @ 24 mA, Resistive load |
| Temperature Coefficient |
± (0.001% Rdg +0.0015% FS) / °C |
± (0.0025% Rdg +0.0035% FS) / °C |
| 2-Wire Transmitter Simulation Mode |
| Specification |
Simulation (External Power) |
|
| Resolution |
.001 mA |
- |
| Range |
0 - 24.000 mA |
- |
| Accuracy |
± (0.01% of reading + 0.015% FS) |
- |
| Loop Voltage Limits |
1 VDC min., 55 VDC max. |
- |
| Temperature Coefficient |
± (0.003% Rdg +0.003% FS) / °C |
- |
| Loop Power Mode |
| Specification |
Regulated Power Source |
|
| Range |
24 VDC |
- |
| Open Circuit Voltage |
24 VDC |
- |
| Output Drive |
15 VDC Min. @ 24mA, Resistive load |
- |