Meriam M2110 General Specifications |
TYPE & RANGE |
DN |
Differential Non-Isolated |
CI |
Compound Isolated |
GI |
Gauge Isolated |
AI |
Absolute Isolated |
DI |
Wet/Wet Differential Isolated |
MEDIA COMPATIBILITY |
Non-isolated sensor for clean, dry, non-corrosive gases. Isolated sensor for fluids compatible with 316SS. |
ACCURACY |
Indication |
± 0.05% of Full Scale (includes combined effects of linearity, repeatability, hysteresis and temperature). |
Analog output |
± 0.05% of Full Scale + 3 counts. AC-powered gauges require 15 minute warm-up. NIST certification supplied. |
ZERO DRIFT |
± 0.015% of full scale. Zeroing prior to measurement eliminates this effect. |
TEMPERATURE |
Storage |
-40 °F to 140 °F (-40 °C to 60 °C) |
Operating |
-4 °F to 122 °F ( -20 °C to 50 °C) |
PRESSURE LIMITS |
Twice range for AI and GI sensors. Twice range for DN sensors when pressurized on high side only; 150 PSI (10.5 Kg/cm²) static when applied to both sides of the sensor simultaneously. |
DISPLAY |
4 1/2 digit, 0.6 inch (15.24 mm) LCD. |
PRESSURE CONNECTIONS |
1/8" female NPT. Brass on non-isolated, 316SS on isolated sensors. |
INPUT POWER |
SetPoint model is standard with selectable 110V AC 50/60 Hz, 240V AC 50/60 Hz, or 24V DC power. Battery models are powered by two Lithium “D” Cells. Battery Life is approximately 100 days at a 100% duty cycle. |
I/O |
Included are an RS 232C interface and two programmable SPDT relays, rated 1.0 amp resistive @ 24V DC, 0.5 amp resistive @ 115V AC. |
Option 1 |
Lithium Battery model with replaceable battery board with automatic shut-off (programmable). No outputs available. |
Option 2 |
Current Loop model with 4-20mA output, 24V DC (50mA) power and RS 232C. |
ENCLOSURE |
4 lbs. 1/4 DIN (3.8" X 3.8" X 6.5") aluminum enclosure with epoxy finish. Standard model is NEMA 4X on front panel only. Battery and Current Loop models are NEMA 4X throughout. |
MOUNTING |
1/4 DIN panel mounting standard. Options: 2" pipe mount bracket; or portable handle. |