| Meriam MGF16BN General Specifications |
| ACCURACY |
Standard |
±0.25% of full scale ±1 least significant digit includes linearity, hysteresis, repeatability |
| Optional |
-HA ±0.1% FS ±1 LSD
-NC NIST traceable test report and calibration data |
| DISPLAY |
3 readings per second nominal display update rate
4½ digit LCD, 0.5" H, 5 character 0.25" H alphanumeric lower display |
| CONTROLS & FUNCTIONS |
Front button turns gauge on or off, zeros gauge, and cycles through min/max functions Internal push buttons for selection of engineering units, selection of auto shutoff times, and calibration |
| CALIBRATION |
Pass code protected calibration
Non-interactive zero, span, and linearity, ±10% of range |
| AUTOMATiC SHUTOFF |
5 minute default
User selectable times ranging from 1 minute to 8 hours or front button on/off |
| POWER |
2 AA alkaline batteries
Approximately 2000 hours battery life
Low battery symbol on display when batteries must be replaced |
| WEIGHT |
Gauge |
9 ounces (approximately) |
| Shipping weight |
1 pound (approximately) |
| MATERIAL |
ABS/polycarbonate NEMA 4X case, rear gasket, polycarbonate label |
| CONNECTION SIZE |
1/4" NPT male, 316L SS |
| MATERIAL, MEDIA COMPATIBILITY |
All wetted parts are 316L SS, compatible with most liquids and gases |
| OVERPRESSURE |
3000 psig range |
5000 psig |
| 5000 psig range |
7500 psig |
| All others |
2 x pressure range |
| 112.5% FS out-of-range display |
i – – – or i –.–.–.– depending on model |
| BURST PRESSURE |
4 times sensor pressure rating, or 10,000 psi, whichever is less |
| ENVIRONMENTAL |
Storage temperature |
–40 to 203°F (–40 to 95°C) |
| Operating temperature |
–4 to 185°F (–20 to 85°C) |
| Compensated temperature |
32 to 158°F (0 to 70°C) |