| Setting temperature range in °C |
min. 10°C above ambient up to 250°C |
| resolution of display for actual values |
0.1°C |
| Temperature |
2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
| ControlCOCKPIT |
TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays. |
| Language setting |
German, English, Spanish, French, Polish, Czech, Hungarian |
| Timer |
Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
| Function HeatBALANCE |
adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to 50 % |
| Function SetpointWAIT |
the process time does not start until the set temperature is reached |
| Calibration |
three freely selectable temperature values |
| Fan |
forced air circulation by quite air turbine, adjustable in 10 % steps for each segment individually |
| Fresh air admixture |
adjustment of pre-heated fresh air admixture by air flap control in 10 % steps for each segment individually |
| Vent |
vent connection with restrictor flap |
| Documentation |
programme stored in case of power failure |
| Programming |
AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
| Temperature control |
independently working, digitally adjustable electronic micro-processor overtemperature monitor TWW, protection class 3.1 (max-value for overtemperature, min-value for undertemperature) |
| Temperature control |
mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature |
| AutoSAFETY |
additionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature |
| Autodiagnostic system |
for fault analysis |
| Alarm |
visual and acoustic |
| Internals |
2 stainless steel grid(s), electropolished |
| Door |
fully insulated stainless steel doors on two sides |
| Installation |
with feet |
| Interior |
easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides |
| Volume |
749 l |
| Dimensions W x H x D in mm |
w(A) x h(B) x d(C): 1040 x 1200 x 600 mm |
| Max. number of internals |
14 |
| Max. loading of chamber |
300 kg |
| Max. loading per internal |
30 kg |
| Dimensions |
w(D) x h(E) x d(F): 1224 x 1714 x 782 mm |
| Altitude of installation |
max. 2,000 m above sea level |
| Ambient temperature |
5 °C to 40 °C |
| Humidity rh |
max. 80 %, non-condensing |
| Overvoltage header |
II |
| Pollution degree |
2 |
| Transport information |
The appliances must be transported upright |
| Customs tariff number |
8419 8998 |
| Country of origin |
Federal Republic of Germany |
| WEEE-Reg.-No. |
DE 66812464 |
| Dimensions approx incl. carton |
B x H x T: 1330 x 1920 x 1050 mm |
| Net weight |
approx. 260 kg |
| Gross weight carton |
approx. 331 kg |