Humidity chamber HCP Specifications |
Model sizes/Description |
108 |
153 |
246 |
Monitor |
Microprocessor temperature monitor acting as overtemperature protection (protection class 3.1), with Pt100 incorporating fault diagnostics with visual and audible alarm |
Digital over- and undertemperature monitor |
Temperature monitoring band automatically linked to the setpoint (ASF) |
Relay for reliable heating cut-off in case of fault |
Mechanical temperature limiter (TB) |
Audible alarm: Over- and undertemperature, underhumidity, open door and empty water tank |
Timer functions |
Real-time/weekly programmer with group function (e.g. Monday – Friday), programme operation with up to 40 ramps for temperature and humidity (MEMoryCard XL) |
Documentation |
Internal log memory 1024 kB as ring memory for all setpoints, actual values, errors, settings with real-time and date; capacity approx. 3 months at 1 min intervals |
Parallel printer interface for printing logging files, suitable for all PCL3-compatible ink jet printers (USB available via converter) |
"Celsius" software for control and documentation of temperature and relative humidity |
Setup |
Calibration (no separate PC required), Temperature: 3-point calibration on controller, Humidity: 2-point calibration at 20 % and 90 % |
Setting of language for dialogue and display DE / EN / ES / FR / IT |
Further data |
Electrical load at 230/115 V (50/60 Hz) |
Standard accessories |
Perforated stainless steel shelves |
Works calibration certificate (measuring point chamber centre at +60 °C) |
number 2 |
Packing data |
Net weight
approx. kg 70 |
80 |
110 |
Gross weight (packed in carton)
approx. kg 95 |
106 |
132 |
Width
approx. cm 83 |
83 |
93 |
Height
approx. cm 105 |
130 |
114 |
Depth
approx. cm 80 |
80 |
93 |
Order No. Humidity Chambers |
HCP108 |
HCP153 |
HCP246 |
Options |
108 |
153 |
246 |
Voltage 115 V, 50/60 Hz |
2X |
Door hinged on the left |
B8 |
Entry port, 23 mm clear diameter, for introducing connections at the side, moisture tight, can be closed by flap and silicone stopper, standard positions left centre/centre, left centre top, right centre/centre, right centre top |
F0,F1,F2,F3 |
Entry port (silicone), 40 mm clear diameter, for introducing connections, moisture tight, can be closed by silicone stopper, at the back (please, state location) |
F7 |
Works calibration certificate for 80 % rh (measured at +50 °C) |
D00107 |
Start-up of HCP and brief training (D, A, CH only) through Memmert service |
K9 |
Stacking version for 2 units of equal size (bottom unit modification) |
G3 |
Process-dependent electromagnetic door lock |
D4 |