Electrical supply range |
120 V ± 10% |
Bond test (10 A and 25 A) |
Open circuit voltage |
9 V a.c. ± 10% ± 0.1 V
(Supply: 120 V 60 Hz) |
Earth bond resistance accuracy |
±5% ±3 digits (0 to 0.49 Ω)
±5% ±5 digits 0.5 to .99 Ω) |
Earth bond resistance resolution |
10 mO (0 to 1.99 Ω) |
Display range |
0 to 1.99 O |
Continuity test (200 mA) |
Continuity test voltage |
4.0 V d.c. -0% +10% |
Continuity resistance accuracy |
±5% ±3 digits (0 to 0.49 Ω)
±5% ±5 digits (0.5 to .99 Ω) |
Continuity resistance resolution |
10 mO (1 to 19.99 Ω) |
Display range |
0 to 19.99 O |
Insulation resistance test (250 V and 500 V) |
Accuracy (120 V) |
±2% ±5 digits (0 to 19.99 MΩ) |
±5% ±10 digits (20 to 99.99 MΩ) |
Resolution |
0.01 MO |
Display range |
0 to 99.99 MO |
Insulation resistance test (250 V and 500 V) |
Accuracy (120 V) |
±2% ±5 digits (0 to 19.99 MΩ)
±5% ±10 digits (20 to 99.99 MΩ) |
Resolution |
0.01 MO |
Display range |
0 to 99.99 MO |
Substitute leakage test |
Test voltage and frequency |
40 V ac ±10% nominal mains frequency |
Leakage current accuracy |
±5% ±5 digits |
Leakage current resolution |
0.01 mA |
Display range |
0 to 19.99 mA |
Differential leakage current |
Test voltage and frequency |
Nominal mains supply |
Accuracy |
±5% ±5 digits |
Resolution |
0.01 mA |
Display range |
0 to 19.99 mA |
Touch current test |
Test voltage and frequency |
Nominal mains supply |
Accuracy |
±5% ±5 digits |
Resolution |
0.01 mA |
Display range |
0 to 10 mA |
Operational test |
Test voltage and frequency |
Nominal mains supply |
VA rating |
110 V 1800 VA |
VA accuracy |
±5% ±10 digits (0 - 99 VA)
±5% ±50 digits (100 VA - 999 VA)
±5% ±100 digits (1000 VA - 1800 VA) |
Resolution |
1 VA (0 to 1800 VA) |
Display range |
0 to 1800 VA |
Extension lead test |
Tests performed |
Bond, insulation and polarity |
Polarity test voltage |
12 V |
Polarity checks |
Lead OK
Live neutral S/C
Live neutral reversed
Live/neutral O/C |
Hi-pot/Flash test |
Voltage |
1500 V ac. nominal for Class 1
3000 V ac nominal for Class 2 |
Current |
<3.5 mA short circuit @ 253 V primary supply voltage |
Breakdown current accuracy |
±5% ±5 digits |
Breakdown current resolution |
0.01 mA |
Display range |
0 to 3.0 mA |
Fuse test |
Test voltage |
3.3 V |
Warning |
Audible beep if fuse is OK |
Environmental |
Operating temperature |
-10 to +50 ºC (+14 ºF to +122 ºF) |
Storage temperature |
-20 to +60 ºC (-4 ºF to +140 ºF) |
Humidity |
90% RH @ -10 to +30 ºC (14 ºF to 86 ºF) 75% RH @ +30 to +50 ºC (86 ºF to 122 ºF) |
Maximum altitude |
2000 m |
Ingress protection |
IP40 |
Weight |
5.0 kg (11 lb) |
Dimesions |
Instrument |
180 (H) x 255 (W) x 320 (L) mm
7 (H) x 10 (W) x 12.5 (L) in. |
Packaging |
210 (H) x 280 (W) x 390 (L) mm
8.2 (H) x 11 (W) x 15.3 (L) in. |
Safety |
IEC 61010-1: CAT II 300 V |
EMC |
IEC 61326-1 |
2006 |
IEC 61326-2-2 |
2005 |