Electrical supply range |
120 V a.c. ±10% |
Supply measurement |
Voltage |
±2% ±1 V |
Frequency |
±1% ± 0.1 Hz |
Bond test |
Open circuit voltage |
9 V a.c. ± 10% ± 0.1 V |
10 A bond test current |
10 A rms ±25% ± 0.5 A into 0.1 O |
25 A bond test current |
25 A rms ± 20% ± 0.5 A into 0.1 O |
Earth bond resistance accuracy |
±5% ±3 digits (0 to 0.49 O)
±5% ±5 digits 0.5 to 1.99 O) |
Earth bond resistance resolution |
10 m O (0 to 1.99 O) |
Display range |
0 to 1.99 O |
Bond test nulling |
Up to 1.00 O |
Adjustable test duration |
User selectable from 1 sec to 20 sec |
Continuity test |
Note |
The continuity test is a DC test performed automatically in both positive and negative directions. The average of the two results is shown. |
Continuity test voltage |
4.0 V d.c. -0% +10% (open circuit) |
Continuity test current n |
200 mA -0% +10% ±5 mA (into 2 O load) |
Continuity resistance accuracy n |
±5% ±3 digits (0 to 0.49 )
±5% ±5 digits 0.5 to 19.99 O) |
Continuity resistance resolution |
10 mO (1 to 19.99 O) |
Display range |
0 to 19.99 O |
Continuity test nulling |
up to 9.99 O |
Test duration |
User selectable from 1 sec to 20 sec |
Insulation test |
Insulation test voltage |
250 V d.c. –0%/+25% open circuit
500 V d.c. –0%/+25% open circuit
(500 V d.c. across 0.5 MO) |
Insulation resistance accuracy |
±2% ±5 digits (0 to 19.99 MO)
±5% ±10 digits (20 to 99.99 MO) |
Insulation resistance resolution |
0.01 MO (0.10 to 99.99 MO) |
Display range |
0 to 99.99 MO |
Test duration |
User selectable from 1 sec to 1 minute |
Substitute leakage test |
Test voltage |
40 V a.c. ±10% |
Test frequency |
Nominal mains 50/60 Hz |
Leakage current accuracy |
±5% ±5 digits |
Leakage current resolution |
0.01 mA |
Display range |
0 to 19.99 mA |
Test duration |
User selectable from 1 sec to 1 minute |
Reading corrected to 120 V + 10% a.c. |
Differential leakage current |
Test voltage |
Nominal mains 120 V a.c. |
Test frequency |
Nominal mains 60 Hz |
Accuracy |
±5% ±5 digits |
Resolution |
0.01 mA |
Display range |
0 to 19.99 mA |
Reading corrected to 120 V +10% a.c. |
Touch current test |
Test voltage |
Nominal mains 120 V a.c. |
Test frequency |
Nominal mains 60 Hz |
Touch current accuracy |
±5% ±5 digits |
Touch current resolution |
0.01 mA |
Display range |
0 to 10 mA |
Test duration |
User selectable from 1 sec to 5 sec |
Reading corrected to 120 V + 10% a.c |
Operational test |
Test voltage |
Nominal mains 120 V a.c. |
Accuracy |
±5% ±10 digits (0 to 99 VA)
±5% ±50 digits (100 VA - 999 VA)
±5% ±100 digits (1000 VA - 1800 VA) |
Resolution |
1 V A (0 to 1800 VA) |
Display range |
0 to 3.99 KVA |
Reading corrected to 120 V a.c. Results show load VA |
Extension lead test |
Test includes insulation and bond tests. |
Polarity test voltage |
12 V |
Polarity |
Lead OK
Live neutral S/C
Live neutral reversed
Live/neutral O/C |
Fuse test |
Test voltage |
3.3 V |
Warning |
Audible beep if fuse is OK |
Safety |
Meets the requirements of IEC 61010-1 |
2001 |
Test leads meet the requirements of IEC 61010-031 |
2002 |
Creepage and clearances for 300 V to Earth Category II |
Environmental |
Operating temperature |
50º to 122º F (10º to 50º C) |
Storage Temperature |
-4º to +140º F (-20º to +60º C) |
Humidity |
90% RH @ 14º to 86º F (-10º to +30º C)
75% RH @ 86º to 122º F(+30º to +50º C) |
Weight |
10 lb (4.5 kg) |
Dimensions |
Instrument only |
7 in. H x 13 in. D x 10 in. W
(175 mm H x 320 mm D x 250 mm W) |
With case |
7.5 in. W x 16 in. D x 11 in. W
(190 mm H x 400 mm D x 290 mm W) |