General |
Loop Power |
Voltage | 24 V or 28 V, selectable |
Maximum Current | 22 mA, short-circuit protected |
Input Voltage | 30 VDC maximum |
Accuracy | 5% |
Note | A 250Ù series resistance is automatically supplied whenever looppower is enabled. |
Ramp Functions |
Source Functions | Voltage, current, Ohms, frequency, temperature |
Rate | 4 steps/second |
Trip Detect | continuity or voltage; continuity detection not available when sourcing current |
Data Log Function |
Measure Functions | Voltage, current, Ohms, frequency, temperature, pressure |
Reading Rate | 1, 2, 5, 10, 20, 30, or 60 readings/minute |
Maximum Record Length | 8,000 readings 7,980 for 30 or 60 readings/minute |
Environmental |
Operating Temperature | -20 °C to +50 °C -10 °C to +50 °C for in spec frequency and AC read |
Storage Temperature | -20 °C to +60 °C |
Altitude | 9,186 ft (2800 m) above mean sea level |
Stability (90 day) | Typical 90-day measurement/read and source/simulate accuracy can be estimated by dividing the one year “% of reading” or “% of output” specifications by 2. Floor specifications, expressed as “% of FS” or “counts” or “Ohms” remain constant |
Power Requirements | 7.2 VDC |
Battery | NiMH rechargeable, 3500 mAH; included |
Battery Life | >8 hours, typical usage |
Physical |
Dimensions | 9.3"H x 5.1"W x 2.4"D(236 x 130 x 61 mm) |
Weight | 3 lbs, 1 oz. (1.4 kg) |
Connectors/Ports | Pressure module connector RS232 serial External power in |
Included Accessories | Test lead set (4 leads, test clips, test probes) NimMH battery pack Battery pack charger Deluxe carrying case BPPA100 Pressure Module Adapter Serial port cable HART Communications cable NIST-traceable calibration certificate, with data Multi-lingual instruction manual on CD Warranty registration card |
Optional Accessories | Barcode wand |