| Specifications for M2001(1 year at 23°C ±5°C; % of reading, unless otherwise noted) |
| Output Voltage |
| Range | 0 to 100.000 mV 0 to 1.00000 V 0 to 10.0000 V 0 to 100.000 V |
| Resolution |
| 0 to 100 mV | Range 1 µV |
| 0 to 1 V Range | 10 µV |
| 0 to 10 V Range | 100 µV |
| 0 to 100 V Range | 1 mV |
| Accuracy (% of reading) |
| 0 to 100 mV Range | ±0.003% (30ppm) ± 3 µV |
| 0 to 1 V Range | ±0.003% (30ppm) ± 20 µV |
| 0 to 10 V Range | ±0.003% (30ppm) ± 200 µV |
| 0 to 100 V Range | ±0.003% (30ppm) ± 2 mV |
| Maximum Burden (= 1 Ohm output impedance) |
| 0 to 100 mV Range | 10 mA |
| 0 to 1 V Range | 10 mA |
| 0 to 10 V Range | 10 mA |
| 0 to 100 V Range | 1 mA |
| Output Current |
| Range | 0 to 100.000 mA |
| Resolution | 1 µA |
| Accuracy (% of reading) | ±0.01% ± 2 Counts |
| Maximum Burden | 10 V |
| Thermocouples |
| Output |
| Types | J, K, T, E, R, S, N, B, L, U, C |
| Range | mV |
| Resolution | 0.1 °C/°F |
| Accuracy | 0.14 °C; Type J, typical |
| Input |
| Types | J, K, T, E, R, S, N, B, L, U, C |
| Range | mV |
| Resolution | 0.01 °C/°F |
| Accuracy | 0.14 °C; Type J, typical |
| RTD |
| Output |
| Range | PT385 (100, 200, 500, 1000),PT392, PT3916 (JIS), Ni120, Cu10, YSI400 |
| Resolution | 0.1 °C/°F; PT385-100, typical |
| Accuracy | ±0.1 °C; PT385-100, typical |
| Input (All Ω/RTD inputs are 4 wire) |
| Range | PT385 (100, 200, 500, 1000),PT392,PT3916 (JIS), Ni120,Cu10,YSI400, 25 Ohm SPRT |
| Resolution | 0.001 °C/°F; PT385-100, typical |
| Accuracy | ±0.020 °C; PT385-100, typical |
| Ohms |
| Output |
| Range | 5 to 4000.0 Ohms |
| Resolution |
| 5 to 400.00 | 0.01 |
| 5 to 4000.0 | 0.1 |
| Accuracy |
| 5 to 400.00 | ±0.03 |
| 5 to 4000.0Ω ±0.3% |
| Input |
| Range | 0 to 4000.00 Ohms |
| Resolution |
| 0 to 400.00 | 0.001 |
| 0 to 4000.0 | 0.01 |
| Accuracy |
| 0 to 400.00 | 40 PPM ±0.002% |
| 0 to 4000.0 | 40 PPM ±0.02% |
| Pressure |
| Range | 0 to 1 inch H2O; to 10,000 psi |
| Compatibility | All Fluke 700 Series Pressure Modules |
| Stability |
| Warm-up Time | 30 minutes to rated accuracy |
| Temp Co. (=18°C/>28°C) | 10% of accuracy spec/°C |
| Environmental |
| Operating Temperature | 0°C to +50°C |
| Storage Temperature | -20°C to +70°C |
| Humidity |
| Operating | <80% to 30°C <70% to 40°C <40% to 50°C |
| Storage | <95%, non-condensing |
| Power Requirements |
| VoltageRange | 90 to 240 VAC <15 VA |
| Mechanical |
| Dimensions | 11.5" x 4.7" x 8.75"(29.21 cm x 11.93 cm x22.00cm) |
| Weight | Approximately 5 pounds(2.27 kg) |
| Display | 16 x 2 large character alphanumeric, backlit LCD |