Analog - Vertical |
Analog Bandwidth @ 50Ω (-3dB) |
100 MHz |
200 MHz |
350 MHz |
500 MHz |
1 GHz |
Rise time |
3.5 ns (typical) |
1.75 ns (typical) |
1 ns (typical) |
800 ps (typical) |
430 ps (typical) |
Input Channels |
4 |
Vertical Resolution |
8-bits; up to 11-bits with enhanced resolution (ERES) |
Sensitivity |
50 Ω: 1mV/div - 1 V/div; 1 MΩ: 1 mV/div - 10 V/div |
DC Gain Accuracy |
±(1.5%) Full Scale, Offset at 0V, > 5mV/div; ±(2.5%) < 5 mV/div |
BW Limit |
20 MHz |
20 MHz, 200 MHz |
Maximum Input Voltage |
50 Ω: 5 Vrms, ±10 V Peak; 1 MΩ: 400 V max (DC + Peak AC ≤ 10 kHz) |
Input Coupling |
50 Ω: DC, GND; 1 MΩ: AC, DC, GND |
Input Impedance |
50 Ω ±2.0%, 1 MΩ ±2.0% || 16 pF |
Offset Range |
50 Ω: 1 mV - 19.8 mV: ±2 V, 20 mV - 100 mV: ±5 V, 102 mV - 198 mV: ±20 V, 200 mV - 1 V: ±50 V
1 MΩ: 1 mV - 19.8 mV: ±2 V, 20 mV - 100 mV: ±5 V, 102 mV - 198 mV: ±20 V, 200 mV - 1 V: ±50 V,
1.02 V - 1.98 V: ±200 V, 2 V - 10 V: ±400 V |
Offset Accuracy |
±(1.0% of offset value + 1.5%FS + 1 mV) |
Analog - Acquisition |
Sample Rate (Single-shot) |
1 GS/s
(2 GS/s interleaved)
|
2 GS/s
(4 GS/s interleaved) |
Sample Rate (Repetitive) |
50 GS/s |
Standard Memory ( 4 Ch / 2 Ch) |
10 Mpts / 20 Mpts |
Acquisition Modes |
Real Time, Roll, RIS (Random Interleaved Sampling),
Sequence (Segmented Memory up to 1,000 segments with 1μs minimum intersegment time) |
Real Time Timebase Range |
5 ns/div - 100 s/div |
2 ns/div - 100 s/div |
1 ns/div - 100 s/div |
500 ps/div - 100 s/div |
Roll Mode Timebase Range |
Up to 100 s/div (roll mode is user selectable at ≥ 50 ms/div) |
Timebase Accuracy |
±10 ppm measured over > 1ms interval |
Display System |
Display Size |
10.1” widescreen capacitive touch screen |
Display Resolution |
1024 x 600 |
Connectivity |
Ethernet Port |
10/100Base-T Ethernet interface (RJ-45 connector) |
Removable Storage |
(1) MicroSD Port - 16 GB micro SD card installed standard |
USB Host Ports |
(4) USB 2.0 Ports Total - (2) Front USB 2.0 Ports |
USB Device Port |
(1) USBTMC |
GPIB Port (Optional) |
Supports IEEE - 488.2 |
External Monitor Port |
Standard DB-15 connector (support resolution of 1024x600) |
Remote Control |
Via Windows Automation, or via Teledyne LeCroy Remote Command Set |
Network Communication Standard |
VICP and LXI compatible |
Power Requirements |
Voltage |
100 - 240 VAC ± 10% at 50-60 Hz +/-5%; 100 - 120 VAC ± 10% at 400 Hz +/- 5%; Automatic AC Voltage Selection |
Power Consumption (Nominal) |
80 W / 80 VA |
Power Consumption (Max) |
150 W / 150 VA (with all PC peripherals, digital leadset and active probes connected to 4 channels) |
Environmental |
Temperature |
Operating: 0 °C to 50 °C; Non-Operating: -30 °C to 70 °C |
Humidity |
Operating: 5% to 90% relat ive humidity (non-condensing) up to ≤ 30 °C, Upper limit derates to 50% relative humidity(non-condensing) at +50 °C
Non-Operating: 5% to 95% relative humidity (non-condensing) as tested per MIL-PRF-28800F |
Altitude |
Operating: 3,048 m (10,000 ft) max at ≤ 25C; Non-Operating: Up to 12,192 meters (40,000 ft) |
Physical |
Dimensions (HWD) |
10.63”H x 14.96”W x 4.92”D (270 mm x 380 mm x 125 mm) |
Weight |
4.81 kg (10.6 lbs) |
Regulatory |
CE Certification |
Low Voltage Directive 2014/35/EU; EN 61010-1:2010, EN 61010-2-030:2010
EMC Directive 2014/30/EU; EN 61326-1:2013, EN61326-2-1:2013; RoHS2 Directive 2011/65/EU |
UL and cUL Listing |
UL 61010-1, UL 61010-2-030:2010, 3rd Edition; CAN/CSA C22.2 No. 61010-1-12 |