Bandwidth |
500 MHz (guaranteed, without leads)
500 MHz (High performance solder-in and browser)
500 MHz (2" Y-lead/solder-in)
500 MHz (5" Y-lead)
30 MHz (Hi-Temp lead) |
Rise Time (10-90%) |
700 ps 350 ps |
Differential Voltage Range |
80 V (DC + peak AC) from 200 mV/div to 20 V/div |
Common Mode Voltage Range |
±60 V (DC + peak AC) |
Maximum Input Voltage to Earth |
80 V (DC + peak AC), max 60 V DC (either input to ground) |
Maximum Safe Input Voltage |
For Hand-held use: 28.28 Vrms or 60 V DC (referenced to ground) per IEC/EN 61010-031:2015 |
Sensitivity |
200 mV/div to 1 V/div (7.8x) |
DC Gain Accuracy |
±0.5% (guaranteed) |
DC Gain Drift |
≤ 0.075%/°C, can be calibrated out with precision gain cal |
Frequency Response Flatness |
DC to 100MHz: 0.1 dB |
Offset Range |
±60V |
Attenuation |
7.8x / 17.5x / 70x |
Input Impedance |
200 kΩ || 0.6 pF (between inputs), 100 kΩ || 1 pF (either input to ground) |
Input/Output Coupling |
DC |
Output Termination |
50 Ω |
Interface |
ProBus |
Cable Length |
1.42 m from probe sockets to oscilloscope connection |
Noise and Rejection |
CMRR |
DC - 10 kHz: 80 dB
100 kHz: 70 dB
1 MHz: 55 dB
100 MHz: 50 dB
500 MHz: 35 dB |
Noise (Probe) |
200 mV/div to 1 V/div: 3.25mV rms
1.02 V/div to 2.5 V/div: 4.5mV rms
2.55 V/div to 20 V/div: 14.5mV rms |
Environmental |
Temperature |
0°C to 50°C (Operating), -40°C to 70°C (Non-Operating) |
Humidity (Operating) |
5% to 90% RH (Non-Condensing) up to 30°C, decreasing linearly to 45% RH at 50°C |
Humidity (Non-Operating) |
5% to 95% RH (Non-Condensing), 75% RH above 40°C, 45% RH above 50°C |
Altitude (Operating) |
Up to 3000 m (9842 ft) |
Certifications |
CE Declaration of Conformity |
Conforms to EN61010-031:2015, EN61326-1:2013, and EN50581:2012 |