Keysight N9950B-238 - Spectrum Analyzer Time Gating Option for N9950B
THIS OPTION REQUIRES OPTION 233 (SOLD SEPARATELY)
Option 238 is the spectrum analyzer time gating option for the FieldFox spectrum analyzer mode.
With time gating, you can measure the spectrum of a periodic signal during a specified time interval. Pulsed-RF signals are an example of a periodic signal that can be measured with time gating. For example, you can measure the pulse during the on period, not the transition or the off period. Or you can exclude interfering signals such as a periodic transient. Time gating allows you to view spectral components that would otherwise be hidden. FieldFox’s time gating method is a Gated FFT.
- Gate method: Gated FFT
- Span range: Any span
- RBW range: 1 Hz to 300 kHz (derived from gate width)
- Gate delay range: -150 ms to 10 s
- Gate width (length) range: 6 μs to 1.8 s
- Gate sources: External, RF burst, Video