Auxiliary Output Option for 6813C (26V, 0.1A)
Option 026 consists of an auxiliary 26 Vrms output located on the rear panel of the AC source. This is a programmable output with a maximum output current of 100 mArms. The output frequency tracks the frequency of the main output of the AC source. Option 026 adds the following capabilities:
- AC voltage and phase are programmable, in immediate mode only
- Software calibration is available for both voltage and phase. Calibration is performed at a single frequency (400 Hz). Note that there is no compensation for frequency-dependent voltage variation or phase shift
- The auxiliary output waveform and frequency are always the same as the main output
- For a sine wave output, the maximum programmed rms voltage value is 26.02 V. For any waveform, the programmed rms voltage value may not result in a peak voltage of more than 36.8 V
- Current limit is not programmable
- DC voltage is not available