Fast |
0.01 |
|
80 dB |
Medium |
0.1 |
|
80 dB |
Slow |
1 |
60 dB |
80 dB |
Load Impedance |
Stable into 20,000pF typical |
Common Mode Voltage |
250V DC (40V DC for Model 2440) |
Common Mode Isolation |
>10 9 Ω, <1000pF |
Overrange |
105% of range, source and measure |
Max. Voltage Drop Between Input/Output and sense Terminals |
5V |
Max. Sense Lead Resistance |
1MΩ for rated accuracy |
Sense Input Impedance |
>10 Ω |
Guard Offset Voltage |
<150µV, typical (300µV for Models 2430, 2440) |
Source Output Modes |
Pulse (Model 2430 only)
Fixed DC level
Memory List (mixed function)
Stair (linear and log) |
Memory Buffer |
5,000 readings @ 5 digits (two 2,500 point buffers). Includes selected measured value(s) and time stamp. Lithium battery backup (3 yr+ battery life) |
Source Memory List |
100 points max |
Programmability |
IEEE-488 (SCPI-1995.0), RS-232, 5 user-definable power-up states plus factory default and *RST |
Interlock |
Active low input |
Handler Interface |
Start of test, end of test, 3 category bits. +5V@ 300mA supply. |
Digital I/O |
1 trigger input, 4 TTL/Relay Drive outputs (33V @ 500mA, diode clamped) |
Power Supply |
100V to 240V rms, 50–60Hz (automatically detected at power up). Model 2400, 2401: 190VA. Model 2410: 210VA. Model 2420: 220VA. Model 2425, 2430: 250VA. Model 2440: 240VA. |
Cooling |
Convection. Model 2410, 2420, 2425, 2430, 2440: Forced air, variable speed. |
EMC |
Conforms to European Union Directive 89/336/EEC, EN 61326-1. |
Safety: UL listed to UL 61010B-1:2003 |
Conforms to European Union Low Voltage Directive. |
Vibration |
MIL-PRF-28800F Class 3 Random |
Warm-up |
1 hour to rated accuracies |
Dimensions |
89mm high × 213mm wide × 370mm deep (3½ in × 8 3 / 8 in × 14 9 / 16 in). Bench Configuration (with handle and feet):104mm high × 238mm wide × 370mm deep (4 1 / 8 in × 9 3 / 8 in × 14 9 / 16 in) |
Dimensions |
89mm high × 213mm wide × 370mm deep (3½ in × 8 3 / 8 in × 14 9 / 16 in). Bench Configuration (with handle and feet):104mm high × 238mm wide × 370mm deep (4 1 / 8 in × 9 3 / 8 in × 14 9 / 16 in) |
Weight |
3.21kg (7.08 lbs) (Model 2425, 2430, 2440: 4.1kg, 9.0 lbs) |
Operating |
0°–50°C, 70% R.H. up to 35°C. Derate 3% R.H./°C, 35°–50°C. |
Storage |
–25°C to 65°C |
2400,
2401 |
200.000mV |
5µV |
0.02%+600µV |
1µV |
0.012%+300µV |
|
±21V@ ±1.05A
±210V@
±105mA* |
2.00000V |
50µV |
0.02%+600µV |
10µV |
0.012%+300µV |
|
20.00000V |
500µV |
0.02%+2.4mV |
100µV |
0.015%+1.5mV |
0.08V/µs |
200.000V* |
5mV |
0.02%+24mV |
1mV |
0.015%+10mV |
0.5 V/µs |
2410,
2410 |
200.000mV |
5µV |
0.02%+600µV |
1µV |
0.012%+300µV |
|
±21V@±1.05A
±1100V@±21mA |
2.00000V |
50µV |
0.02%+600µV |
10µV |
0.012%+300µV |
|
20.0000V |
500µV |
0.02%+2.4mV |
100µV |
0.015%+1mV |
0.15V/µs |
1000.00V |
50mV |
0.02%+100mV |
10mV |
0.015%+50mV |
0.5V/µs |
2420,
2420 |
200.000mV |
5µV |
0.02%+600 µV |
1µV |
0.012%+300µV |
|
±21V@±3.15A
±63V@±1.05A |
2.00000V |
50µV |
0.02%+600µV |
10µV |
0.012%+300µV |
|
20.0000V |
500µV |
0.02%+2.4mV |
100µV |
0.015%+1mV |
0.08V/µs |
60.0000V |
1.5mV |
0.02%+7.2mV |
1mV |
0.015%+3mV |
0.14V/µs |
2425,
2425 |
200.000mV |
5µV |
0.02%+600µV |
1µV |
0.012%+300µV |
|
±21V@±3.15A
±105V@±1.05A |
2.00000V |
50µV |
0.02%+600µV |
10µV |
0.012%+300µV |
|
20.0000V |
500µV |
0.02%+2.4mV |
100µV |
0.015%+1mV |
0.08V/µs |
100.0000V |
2.5mV |
0.02%+12mV |
1mV |
0.015%+5mV |
0.25V/µs |
2430,
2430 |
200.000mV |
5µV |
0.02%+600µV |
1µV |
0.012%+300µV |
|
±105V@±1.05A
±105V@±10.5A
(pulse mode only) |
2.00000V |
50µV |
0.02%+600µV |
10µV |
0.012%+300µV |
|
20.0000V |
500µV |
0.02%+2.4mV |
100µV |
0.015%+1mV |
0.08V/µs |
100.0000V |
2.5mV |
0.02%+12mV |
1mV |
0.015%+5mV |
0.25V/µs |
2440,
2440 |
200.000mV |
5µV |
0.02%+600µV |
1µV |
0.012%+300µV |
|
±10.5V@±5.25A
±42V@±1.05A |
2.00000V |
50µV |
0.02%+600µV |
10µV |
0.012%+300µV |
|
20.0000V |
500µV |
0.02%+1.2mV |
100µV |
0.015%+750µV |
0.08V/µs |
100.0000V |
5mV |
0.02%+4.8mV |
1mV |
0.015%+3mV |
0.25V/µs |
*Not available on Model 2401. |
Temperature Coefficient
(0°–18°C and 28°–50°C) |
±(0.15 × accuracy specification)/°C |
Line |
0.01% of range |
Load |
0.01% of range + 100µV |
Over Voltage Protection |
User selectable values, 5% tolerance. Factory default = none |
Current Limit |
Bipolar current limit (compliance) set with single value. Min. 0.1% of range |
Overshoot |
<0.1% typical (full scale step, resistive load, 10mA range). |
Notes:
1. 2400, 2401, 2410 Only: Specifications valid for continuous output currents below 105 mA. For operation above 105 mA continuous for >1 minute, derate accuracy 10%/35 mA above 105 mA.
2. Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200 mV, 1 A, 10 A ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset specifications, except 200 mV, 1 A, 10 A ranges, add 0.5%.
3. Accuracies apply to 2- or 4-wire mode when properly zeroed.
4. In pulse mode, limited to 0.1 PLC measurement. |
Transient Response Time |
30µs minimum for the output to recover to its spec. following a step change in load |
Command Processing Time |
Maximum time required for the output to begin to change following the receipt of :SOURce:VOLTage|CURRent < nrf > command. |
Autorange On |
10ms. |
Autorange Off |
7ms. |
Output Settling Time |
Time required to reach 0.1% of final value after command is processed. 100µs typical. Resistive load. 10µA to 100mA range |
DC Floating Voltage |
Output can be floated up to ±250VDC (Model 2440 ±40VDC) from chassis ground |
Remote Sense |
Up to 1V drop per load lead |
Compliance Accuracy |
Add 0.3% of range and ±0.02% of reading to base specification. |
Over Temperature Protection |
Internally sensed temperature overload puts unit in standby mode |
Range Change Overshoot |
Overshoot into a fully resistive 100 kΩ load, 10 Hz to 1 MHz BW, adjacent ranges: 100 mV typical, except 20 V/200 V (20 V/60 V on 2420), range boundary, and 2440. |
Minimum Compliance Value |
0.1% of range. |
Model |
Range |
Programming
Resolution |
Source
Accuracy (1 Year)
23°C ±5°C
±(% rdg. + volts) |
Default
Measurement
Resolution |
Measurement
Accuracy (1 Year)
23°C±5°C
±(% rdg. + volts) |
Source/
Sink Limit |
2400,
2401 |
1.00000µA |
50pA |
0.035%+600pA |
10pA |
0.029%+300pA |
±1.05A@±21V
±105mA@±210V |
10.0000µA |
500pA |
0.033%+2nA |
100pA |
0.027%+700pA |
100.000µA |
5nA |
0.031%+20nA |
1nA |
0.025%+6nA |
1.00000mA |
50nA |
0.034%+200nA |
10nA |
0.027%+60nA |
10.00000mA |
500nA |
0.045%+2µA |
100nA |
0.035%+600nA |
100.00000mA |
5µA |
0.066%+20µA |
1µA |
0.055%+6µA |
1.00000A2 |
50µA |
0.27%+900µA |
10µA |
0.22%+570µA |
2410,
2410 |
1.00000µA |
50pA |
0.035%+600pA |
10pA |
0.029%+300pA |
±1.05A@±21V
±21mA@±1100V |
10.0000µA |
500pA |
0.033%+2nA |
100pA |
0.027%+700pA |
100.000µA |
5nA |
0.031%+20nA |
1nA |
0.025%+6nA |
1.00000mA |
50nA |
0.034%+200nA |
10nA |
0.027%+60nA |
20.0000mA |
500nA |
0.045%+4µA |
100nA |
0.035%+1.2µA |
100.00000mA |
5µA |
0.066%+20µA |
1µA |
0.055%+6µA |
1.00000A2 |
50µA |
0.27%+900µA |
10µA |
0.22%+570µA |
2420,
2420 |
10.0000µA |
500pA |
0.033%+2nA |
100pA |
0.027%+700pA |
±3.15A@±21V
±1.05A@±63V |
100.000µA |
5nA |
0.031%+20nA |
1nA |
0.025%+6nA |
1.00000mA |
50nA |
0.034%+200nA |
10nA |
0.027%+60nA |
10.0000mA |
500nA |
0.045%+2µA |
100nA |
0.035%+600nA |
100.000mA |
5µA |
0.066%+20µA |
1µA |
0.055%+6µA |
1.00000A2 |
50µA |
0.067%+900µA |
10µA |
0.066%+570µA |
3.00000A2 |
50µA |
0.059%+2.7mA |
10µA |
0.052%+1.71mA |
2425,
2425 |
10.0000µA |
500pA |
0.033%+2nA |
100pA |
0.027%+700pA |
±3.15A@±21V
±1.05A@±105V |
100.000µA |
5nA |
0.031%+20nA |
1nA |
0.025%+6nA |
1.00000mA |
50nA |
0.034%+200nA |
10nA |
0.027%+60nA |
10.0000mA |
500nA |
0.045%+2µA |
100nA |
0.035%+600nA |
100.000mA |
5µA |
0.066%+20µA |
1µA |
0.055%+6µA |
1.00000A2 |
50µA |
0.067%+900µA |
10µA |
0.066%+570µA |
3.00000A2 |
50µA |
0.059%+2.8mA |
10µA |
0.052%+1.71mA |
2430,
2430 |
10.0000µA |
500pA |
0.033%+2nA |
100pA |
0.027%+700pA |
±1.05A@±105V
±10.5A@±105V
(pulse mode only) |
100.000µA |
5nA |
0.031%+20nA |
1nA |
0.025%+6nA |
1.00000mA |
50nA |
0.034%+200nA |
10nA |
0.027%+60nA |
10.0000mA |
500nA |
0.045%+2µA |
100nA |
0.035%+600nA |
100.000mA |
5µA |
0.066%+20µA |
1µA |
0.055%+6µA |
1.00000A2 |
50µA |
0.067%+900µA |
10µA |
0.066%+570µA |
3.00000A2 |
500µA |
0.059%+2.8mA |
10µA |
0.052%+1.71mA |
10.00000A4 |
500µA |
0.089%+5.9mA |
10µA |
0.082%+1.71mA |
2440,
2440 |
10.0000µA |
500pA |
0.033%+2nA |
100pA |
0.027%+700pA |
±5.25A@±10.5V
±1.05A@±42V |
100.000µA |
5nA |
0.031%+20nA |
1nA |
0.025%+6nA |
1.00000mA |
50nA |
0.034%+200nA |
10nA |
0.027%+60nA |
10.0000mA |
500nA |
0.045%+2µA |
100nA |
0.035%+600nA |
100.000mA |
5µA |
0.066%+20µA |
1µA |
0.055%+6µA |
1.00000A |
50µA |
0.067%+900µA |
10µA |
0.066%+570µA |
5.00000A |
50µA |
0.10%+5.4mA |
10µA |
0.10%+3.42mA |
Temperature Coefficient
(0°–18°C and 28°–50°C) |
±(0.15 × accuracy specification)/°C |
Line |
0.01% of range |
Load |
0.01% of range (except Model 2440 5A range 0.05%) + 100pA |
Voltage Limit |
Bipolar voltage limit (compliance) set with single value. Min. 0.1% of range |
Overshoot |
<0.1% typical (1mA step, RL = 10kO, 20V range for Model 2400, 2401, 2410, 2420, 2425, 2430), (10V range for Model 2440) |
NOTES
1. 2400, 2401, 2410 Only: Specifications valid for continuous output currents below 105 mA. For operation above 105 mA continuous for >1 minute, derate accuracy 10%/35 mA above
105 mA.
2. Full operation (1 A) regardless of load to 30°C (50°C for 2420 and 2440). Above 30°C (50°C for 2420 and 2440) ambient, derate 35 mA/°C and prorate 35 mA/Ω load. 4-wire mode. For current sink operation on 1 A, 3 A, or 5 A ranges, maximum continuous power is limited to approximately 1/2 rated power or less, depending on current, up to 30°C ambient. See power equations in the User’s Manual to calculate allowable duty cycle for specific conditions.
3. For sink mode, 1 µA to 100 mA range, accuracy is: 2400, 2401: ±(0.15% + offset*4). 2410, 2420, 2440: ±(0.5% + offset*3). For 1A range, accuracy is: 2400, 2401: ±(1.5% + offset*8). 2410, 2420, 2440: ±(1.5% + offset*3).
4. 10 A range only in pulse mode. Limited to 2.5 ms pulse width maximum. 10% duty cycle maximum.
5. Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200 mV, 1 A, 10 A ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset specifications, except 200 mV, 1 A, 10 A ranges, add 0.5%.
6. Accuracies apply to 2- or 4-wire mode when properly zeroed.
7. In pulse mode, limited to 0.1 PLC measurement.
8. 2400 only. |
Range
txt |
Default Resolution
txt |
Default Test Current 2400, 2401, 2410
txt |
Default Test Current 2420, 2440
txt |
Normal Accuracy (23°C ±5°C)
1 Year, ±(% rdg. + ohms)txttxt |
Enhanced Accuracy (23°C ±5°C)4 1 Year, ±(% rdg. + ohms) |
2400, 2401 |
2410 |
2420, 2440 |
2400, 2401 |
<0.20000 Ω3 |
- |
- |
- |
Source IACC +
Meas. VACC |
Source IACC +
Meas. VACC |
Source IACC +
Meas. VACC |
Source IACC +
Meas. VACC |
2.00000 Ω3 |
10 µΩ |
- |
1 A |
Source IACC +
Meas. VACC |
Source IACC +
Meas. VACC |
0.17% + 0.0003 Ω |
Source IACC +
Meas. VACC |
20.0000 Ω |
100 µΩ |
100 mA |
100 mA |
0.10% + 0.003 Ω |
0.11% + 0.006 Ω |
0.10% + 0.003 Ω |
0.07% + 0.001 Ω |
200.000 Ω |
1 mΩ |
10 mA |
10 mA |
0.08% + 0.03 Ω |
0.09% + 0.1 Ω |
0.08% + 0.03 Ω |
0.05% + 0.01 Ω |
2.00000 kΩ |
10 mΩ |
1 mA |
1 mA |
0.07% + 0.3 Ω |
0.08% + 0.6 Ω |
0.07% + 0.3 Ω |
0.05% + 0.1 Ω |
20.0000 kΩ |
100 mΩ |
100 µA |
100 µA |
0.06% + 3 Ω |
0.07% + 6 Ω |
0.06% + 3 Ω |
0.04% + 1 Ω |
200.000 kΩ |
1 Ω |
10 µA |
10 µA |
0.07% + 30 Ω |
0.07% + 60 Ω |
0.07% + 30 Ω |
0.05% + 10 Ω |
2.00000 MΩ6 |
10 Ω |
1 µA |
1 µA |
0.11% + 300 Ω |
0.12% + 600 Ω |
0.11% + 300 Ω |
0.05% + 100 Ω |
20.0000 MΩ7 |
100 Ω |
1 µA |
1 µA |
0.11% + 1 kΩ |
0.12% + 2.4 kΩ |
0.11% + 1 kΩ |
0.05% + 500 Ω |
200.000 MΩ3 |
1 kΩ |
100 nA |
- |
0.66% + 10 kΩ |
0.66% + 24 kΩ |
Source IACC +
Meas. VACC |
0.35% + 5 kΩ |
>200.000 MΩ3 |
- |
- |
- |
Source IACC +
Meas. VACC |
Source IACC +
Meas. VACC |
Source IACC +
Meas. VACC |
Source IACC +
Meas. VACC |
Temperature Coefficient (0°–18°C and 28°–50°C) |
±(0.15 × accuracy specification)/°C. |
Source I Mode, Manual Ohms |
Total uncertainty = I source accuracy + V measure accuracy (4-wire remote sense). |
Source V Mode, Manual Ohms |
Total uncertainty = V source accuracy + I measure accuracy (4-wire remote sense) |
6-wire Ohms Mode |
Available using active ohms guard and guard sense. Max. Guard Output Current: 50 mA (except 1 A range). Accuracy is load dependent. Refer to White Paper below for calculation formula: “Obtaining More Accurate Resistance Measurements Using the 6-wire Ohms Measurement Technique.” |
Guard Output Impedance |
<0.1Ω in ohms mode |
NOTES
1. Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200 mV, 1 A, 10 A ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset specifications, except 200 mV, 1 A, 10 A ranges, add 0.5%.
2. Accuracies apply to 2- or 4-wire mode when properly zeroed.
3. Manual ohms only – except 2420, 2440 for 2 Ω range and 2400, 2401, or 2410 for 200 MΩ range.
4. Source readback enabled, offset compensation ON. Also available on 2410, 2420, and 2440 with similar accuracy enhancement.
5. In pulse mode, limited to 0.1 PLC measurement.
6. Except 2440; default test current is 5 µA.
7. Except 2440; default test current is 0.5 µA. |