POWER |
Voltage |
90–134V and 180–250V, universal self-selecting. |
Frequency |
50Hz, 60Hz, or 400Hz self-identifying. |
Consumption |
<55VA |
ENVIRONMENTAL |
Operating Temperature |
0° to 50°C |
Storage Temperature |
–40° to 70°C |
Humidity |
80% R.H., 0° to 35°C, per MIL-T-28800E 1 Para 4.5.5.1.2. |
PHYSICAL |
Case Dimensions |
90mm high × 214mm wide × 369mm deep (3½ in. × 8½ in. × 14½ in.) |
Net Weight |
<4.2kg (<9.2 lbs.) |
Shipping Weight |
<9.1kg (<20 lbs.) |
STANDARDS |
EMI/RFI |
Conforms to VDE 0871B (per Vfg 1046/1984), IEC 801-2. Meets FCC part 15 Class B, CISPR-22 (EN55022) |
Safety |
Conforms to IEC348, CAN/CSA-C22.2. No. 231, MIL-T-28800E 1 . Designed to UL1244 |
Note 1 |
For MIL-T-28800E, applies to Type III, Class 5, Style E |
DC VOLTS
DCV INPUT CHARACTERISTICS AND ACCURACY |
Range |
Full
Scale |
Resolution |
Default
Resolution |
Input
Resistance |
Accuracy ±(ppm of reading + ppm of range) |
5
Minutes |
24
Hours |
90
Days |
1
Year |
2
Years |
200mV |
±210.00000mV |
10nV |
100nV |
>10GΩ |
3+3 |
10+6 |
25+6 |
37+6 |
50+6 |
2V |
±2.1000000V |
100nV |
1µV |
>10GΩ |
2+1.5 |
7+2 |
18+2 |
25+2 |
32+2 |
20V |
±21.000000V |
1µV |
10µV |
>10GΩ |
2+1.5 |
7+4 |
18+4 |
24+4 |
32+4 |
200V |
±210.00000V |
10µV |
100µV |
10MΩ±1% |
2+1.5 |
13+3 |
27+3 |
38+3 |
52+3 |
1000V |
±1100.0000V |
100µV |
1mV |
10MΩ±1% |
10+1.5 |
17+6 |
31+6 |
41+6 |
55+6 |
DC VOLTS NOTES |
- For TCAL ±1°C, following 55-minute warm-up. TCAL is ambient temperature at calibration, which is 23°C from factory
- For TCAL ±5°C, following 55-minute warm-up. Specifications include factory traceability to US NIST
- When properly zeroed using REL function
- DCV Transfer Stability typical applications are standard cell comparisons and relative accuracy measurements. Specs apply for 10 power line cycles, 20-reading digital filter, autozero on with type synchronous, fixed range following 2-hour warm-up at full scale to 10% of full scale, at TREF ±1°C (TREF is the initial ambient temperature). Specifications on the 1000V range are for measurements within 5% of the initial measurement value and following measurement settling
|
Range |
Normal Mode RMS
90 Days, ±2°C from last AC self-cal for 1% to 100% of range
±(% of reading + % of range) |
20–50Hz |
50–100Hz |
0.1–2kHz |
2–10kHz |
10–30kHz |
200 mV |
0.25 + 0.015 |
0.07 + 0.015 |
0.03 + 0.015 |
0.03 + 0.015 |
0.035 + 0.015 |
2 V |
0.25 + 0.015 |
0.07 + 0.015 |
0.03 + 0.015 |
0.03 + 0.015 |
0.035 + 0.015 |
20 V |
0.25 + 0.015 |
0.07 + 0.015 |
0.04 + 0.015 |
0.06 + 0.015 |
0.08 + 0.015 |
200 V |
0.25 + 0.015 |
0.07 + 0.015 |
0.04 + 0.015 |
0.06 + 0.015 |
0.08 + 0.015 |
750 V |
0.25 + 0.015 |
0.1 + 0.015 |
0.08 + 0.015 |
0.09 + 0.015 |
0.12 + 0.015 |
AC VOLTS NOTES |
- Specifications apply for sinewave input, AC + DC coupling, 1 power line cycle, digital filter off, following 55 minute warm-up
- For 1% to 5% of range below 750V range, and for 1% to 7% of 750V range, add 0.01% to range uncertainty. For inputs from 200kHz to 2MHz, specifications apply above 10% of range
- Add 0.001% of reading × (VIN/100V) 2 additional uncertainty above 100V rms
- Typical values
|
Range |
Normal Mode RMS
90 Days, ±2°C from last AC self-cal for 1% to 100% of range
±(% of reading + % of range) |
30–50kHz |
50–100kHz |
100–200kHz |
0.2–1MHz |
1–2MHz |
200 mV |
0.05 + 0.015 |
0.3 + 0.015 |
0.75 + 0.025 |
2 + 0.1 |
5 + 0.2 |
2 V |
0.05 + 0.015 |
0.3 + 0.015 |
0.75 + 0.025 |
2 + 0.1 |
5 + 0.2 |
20 V |
0.1 + 0.015 |
0.3 + 0.015 |
0.75 + 0.025 |
4 + 0.2 |
7 + 0.2 |
200 V |
0.1 + 0.015 |
0.3 + 0.015 |
0.75 + 0.025 |
4 + 0.2 |
|
750 V |
0.15 + 0.015 |
0.5 + 0.015 |
|
|
|
AC VOLTS NOTES |
- Specifications apply for sinewave input, AC + DC coupling, 1 power line cycle, digital filter off, following 55 minute warm-up
- For 1% to 5% of range below 750V range, and for 1% to 7% of 750V range, add 0.01% to range uncertainty. For inputs from 200kHz to 2MHz, specifications apply above 10% of range
- Add 0.001% of reading × (VIN/100V) 2 additional uncertainty above 100V rms
- Typical values
|
Range |
Full
Scale |
Resolution |
Default
Resolution |
Current
Source |
Resistance Accuracy ±(ppm of reading + ppm of range) |
24
Hours |
90
Days |
1
Year |
2
Years |
20 Ω |
21.000000Ω |
1 µΩ |
10 µΩ |
9.2 mA |
29+7 |
52+7 |
72+7 |
110+7 |
200 Ω |
210.00000Ω |
10 µΩ |
100 µΩ |
0.98 mA |
24+7 |
36+7 |
56+7 |
90+7 |
2 kΩ |
2100.0000kΩ |
100 µΩ |
1 mΩ |
0.98 mA |
22+4 |
33+4 |
50+4 |
80+4.5 |
20 kΩ |
21.000000kΩ |
1 mΩ |
10 mΩ |
89µA |
19+4 |
32+4 |
50+4 |
80+4.5 |
200 kΩ |
210.00000kΩ |
10 mΩ |
100 mΩ |
7µA |
20+4.5 |
72+4.5 |
90+4.5 |
130+5 |
2 MΩ |
2.1000000MΩ |
100 mΩ |
1 Ω |
770 nA |
50+4.5 |
110+4.5 |
160+4.5 |
230+5 |
20 MΩ |
21.000000MΩ |
1 Ω |
10 Ω |
70 nA |
160+4.5 |
560+4.5 |
900+4.5 |
1100+5 |
200 MΩ |
210.00000MΩ |
10 Ω |
100 Ω |
4.4 nA |
3000 +
100 |
10000 +
100 |
20000 +
100 |
30000 +
100 |
1 GΩ |
1.0500000GΩ |
100 Ω |
1 kΩ |
4.4 nA |
9000 +
100 |
20000 +
100 |
40000 +
100 |
60000 +
100 |
OHMS NOTES |
- Current source is typically ±9% absolute accuracy
- For 2-wire mode
- Specifications are for 1 power line cycle, 10 reading digital filter, Auto Zero on, 4-wire mode, offset compensation on (for 20Ω to 20kΩ ranges)
- For TCAL ±1°C, following 55 minute warm-up. TCAL is ambient temperature at calibration (23°C at the factory)
- For TCAL ±5°C, following 55-minute warm-up. Specifications include traceability to US NIST
- When measuring resistance of inductive loads, the inductance of that load must be 10mH or less
|
Range |
Full
Scale |
Resolution |
Default
Resolution |
Maximum
Burden
Voltage |
Accuracy ±(ppm of reading + ppm of range) |
24
Hours |
90
Days |
1
Year |
2
Years |
200 µA |
210.00000µA |
10 pA |
100 pA |
0.25 V |
63+25 |
300+25 |
500+25 |
1350+25 |
2 mA |
2.1000000mA |
100 pA |
1 nA |
0.31 V |
64+20 |
300+20 |
400+20 |
750+20 |
20 mA |
21.000000mA |
1 nA |
10 nA |
0.4 V |
65+20 |
300+20 |
400+20 |
750+20 |
200 mA |
210.00000mA |
10 nA |
100 nA |
0.5 V |
96+20 |
300+20 |
500+20 |
750+20 |
2 A |
2.1000000A |
100 nA |
1 µA |
1.5 V |
500+20 |
600+20 |
900+20 |
1350+20 |
DC AMPS NOTES |
- Specifications are for 1 power line cycle, Auto Zero on, 10 reading digital filter
- For TCAL ±1°C, following 55 minute warm-up
- For TCAL ±5°C, following 55 minute warm-up. Specifications include traceability to US NIST
- Add 50 ppm of range for current above 0.5A for self heating
- Actual maximum voltage burden = (maximum voltage burden) × (IMEASURED/IFULL SCALE)
|
Range |
ACI ACCURACY
90 Days, 1 Year or 2 Years, TCAL ±5°C, for 5% to 100% of range, ±(% of reading + % of range) |
20Hz–
50Hz |
50Hz–
200Hz |
200Hz–
1kHz |
1kHz–
10kHz |
10kHz–
30kHz |
30kHz–
50kHz |
50kHz–
100kHz |
200 µA |
0.35+0.015 |
0.2+0.015 |
0.4+0.015 |
0.5+0.015 |
|
|
|
2 mA |
0.3+0.015 |
0.15+0.015 |
0.12+0.015 |
0.12+0.015 |
0.25+0.015 |
0.3+0.015 |
0.5+0.015 |
20 mA |
0.3+0.015 |
0.15+0.015 |
0.12+0.015 |
0.12+0.015 |
0.25+0.015 |
0.3+0.015 |
0.5+0.015 |
200 mA |
0.3+0.015 |
0.15+0.015 |
0.12+0.015 |
0.15+0.015 |
0.5+0.015 |
1+0.015 |
3+0.015 |
2 A |
0.35+0.015 |
0.2+0.015 |
0.3+0.015 |
0.45+0.015 |
1.5+0.015 |
4+0.015 |
|
AC AMPS NOTES |
|
- Specifications apply for sinewave input, AC+DC coupling, 1 power line cycle, digital filter off, following 55 minute warm-up
- Add 0.005% of range uncertainty for current above 0.5A rms for self-heating
- Typical values
|
AC VOLTAGE INPUT |
1Hz–15MHz |
ACCURACY |
±(0.03% of reading) |
TYPICAL RANGES |
Current |
100µA to 12A |
Trace Resistance |
1mΩ to 10Ω typical |
Built-in linearization for J, K, N, T, E, R, S, B thermocouple types to ITS-90 and 100Ω platinum RTDs DIN 43 760 or IPTS-68 |