Current Pulse Termination |
The current pulse will terminate within 3 μs after terminal voltages exceed a bipolar, programmable abort threshold. Separate abort thresholds can be programmed for the sense terminals and the source terminals. |
Range |
Programming Resolution |
Accuracy (1 Year) ±(% rdg. + amps) |
Typical Noise (RMS) 10 kHz–1 MHz |
1 A |
100 μA |
0.17% + 2.0 mA |
380 μA |
5 A |
100 μA |
0.17% + 2.5 mA |
1.4 mA |
10 A |
100 μA |
0.22% + 3.0 mA |
3.1 mA3 |
Temperature Coefficient (0°–18 °C and 28°–50 °C) |
± (0.15 × accuracy specification)/°C. |
Pulsing Limits 4, 5 |
±10 A @ ± 10 V pulse, 3% duty cycle
±3 A @ ± 10 V pulse, 10% duty cycle
±1.0 A @ ± 10 V pulse, 30% duty cycle
±500 mA @ ± 10 V pulse, 60% duty cycle6
±250 mA @ ± 10 V continuous |
Current Regulation |
Line: 0.01% of range. Load: ±100 μA. |
Overshoot |
< ±0.5% of step size (typical). |
Rise Time (10% to 90%) |
<1.7 μs for a full-scale step of current into any load voltage (10 V maximum). |
Current Range |
Settling Time |
1 A |
< 9 μs (Vload ≤ 10 V) |
5 A |
< 9 μs (Vload ≤ 10 V) |
10 A |
< 9 μs (Vload ≤ 10 V) |
Output Off Normal State |
Electrical short (< 1 Ω) between HI and LO.
Maximum DC current from external sources during OUTPUT OFF state must be limited to <1 A. |
Remote Voltage Sense |
Maximum voltage between HI and SENSE HI = ±30 V.
Maximum voltage between LO and SENSE LO = ±30 V. |
Over Temperature Protection |
Internally sensed temperature overload puts unit in standby mode. |
Safety Interlock |
Hardware interlock (available, optional). |
Range |
Programming Resolution |
Accuracy (1 Year) ±(% rdg. + amps) |
Typical Noise (RMS) 0.1 Hz–100 kHz |
250 mA |
10 μA |
0.17% + 1 mA |
200 μA |
Temperature Coefficient (0°–18 °C and 28°–50 °C) |
± (0.15 × accuracy specification)/°C. |
Pulse Width Programming Resolution |
1 μs. |
Pulse Width Programming Maximum |
500 μs. |
Pulse WidthProgramming Minimum |
10 μs. |
Pulse Width Programming Accuracy |
±200 ns. |
Pulse Width Jitter |
110 ns (typical). |
Pulse Period Jitter |
2 μs (typical). |
Pulse Width Programming, Minimum toff |
16 μs. |
Range |
Display Resolution |
Accuracy (1 Year) 23 °C ± 5 °C ± (% rdg. + volts)1 |
5 V |
1 μV |
0.05% + 2.5 mV |
10 V |
10 μV |
0.05% + 4 mV |
Voltage Range |
Settling Time (typical) |
5 V and 10 V |
< 9 μs |
Temperature Coefficient (0°–18 °C and 28°–50 °C) |
± (0.15 × accuracy specification)/C. |
Range |
Display Resolution |
Accuracy (1 Year) 23 °C ± 5 °C ± (% rdg. + amps)1 |
1 A |
1 μA |
0.12% + 0.5 mA |
5 A |
1 μA |
0.12% + 1 mA |
10 A |
10 μA |
0.12% + 1 mA |
Current Measure Settling Time |
Time required to reach specified accuracy of after source level command is processed on a fixed range.
Current Range: 1 A – 10 A.
Settling Time: < 9 μs (typical). |
Temperature Coefficient (0°–18 °C and 28°–50 °C) |
± (0.15 × accuracy specification)/°C. |
Maximum Load Inductance |
3 μH (cable plus device under test (DUT)), measured at 100 kHz. |
Common Mode Isolation |
>1 GΩ, <4500 pF. |
Overrange |
100% of bias range, 101% of source range, 102% of measure range. |
Maximum Source/Sense Lead Resistance |
0.5 Ω / 100 Ω per lead. |
Sense High/Low Input Impedance |
2 MΩ (typical). |
SMU-to-Pulser Transition Time |
<7 ms. |
Range |
Programming Resolution |
Accuracy (1 Year) ±(% rdg. + volts) |
Typical Noise (peak-peak) 0.1 Hz–10 Hz |
100 mV |
5 μV |
0.02% + 250 μV |
20 μV |
1 V |
50 μV |
0.02% + 400 μV |
50 μV |
6 V |
50 μV |
0.02% + 1.8 mV |
100 μV |
40 V |
500 μV |
0.02% + 12 mV |
500 μV |
Temperature Coefficient (0°–18 °C and 28°–50 °C) 2 |
±(0.15 × accuracy specification)/ °C. Applicable for normal mode only. Not applicable for High Capacitance Mode. |
Maximum Output Power and Source/Sink Limits 3 |
40.4 W maximum. ±40.4 V @ ±1.0 A, ±6.06 V @ ±3.0 A, four quadrant source or sink operation. |
Range |
Programming Resolution |
Accuracy (1 Year)a ±(% rdg. + amps) |
Typical Noise (peak-peak) 0.1Hz–10Hz |
100 nA |
2 pA |
0.1% + 100 pA |
5 pA |
1 μA |
20 pA |
0.03% + 800 pA |
25 pA |
10 μA |
200 pA |
0.03% + 5 nA |
60 pA |
100 μA |
2 nA |
0.03% + 60 nA |
3 nA |
1 mA |
20 nA |
0.03% + 300 nA |
6 nA |
10 mA |
200 nA |
0.03% + 6 μA |
200 nA |
100 mA |
2 μA |
0.03% + 30 μA |
600 nA |
1 A |
20 μA |
0.05% + 1.8 mA |
70 μA |
3 A |
20 μA |
0.06% + 4 mA |
150 μA |
10 A 2 |
200 μA |
0.5% + 40 mA |
N/A |
Temperature Coefficient (0°–18 °C and 28°–50 °C) 3 |
±(0.15 × accuracy specification)/ °C. |
Maximum Output Power and Source/Sink Limits 1 |
40.4 W maximum. ±1.01 A @ ±40.0 V, ±3.03 A @ ±6.0 V, four quadrant source or sink operation. |
Noise, 10 Hz to 20 MHz |
<20 mV peak-peak, < 3 mVRMS. 6 V range. |
Transient Response Time |
< 70 μs for the output to recover to within 0.1% for a 10% to 90% step change in load. |
Voltage |
<±0.1% + 10 mV. Step size = 10% to 90% of range, resistive load, maximum current limit/compliance. |
Current |
<±0.1%. Step size = 10% to 90% of range, resistive load.
See Current Source output settling time for additional test conditions. |
Voltage1 |
<300 mV + 0.1% of larger range. Overshoot into a 100 kΩ load, 20 MHz bandwidth. |
Current2 |
<300 mV/RLOAD + 5% of larger range. |
Guard Offset Voltage |
< 4 mV. Current <10 mA. |
Remote Sense Operating Range3 |
Maximum voltage between HI and SENSE HI = 3 V.
Maximum voltage between LO and SENSE LO = 3 V. |
40 V Range |
Maximum output voltage = 42 V – (total voltage drop across source leads). Maximum 1 Ω per source lead. |
6 V Range |
Maximum output voltage = 8 V – (total voltage drop across source leads). Maximum 1 Ω per source lead. |
Overtemperature Protection |
Internally sensed overtemperature condition puts the instrument in standby mode. |
Limit/Compliance |
Bipolar limit (compliance) set with a single value. |
Voltage4 |
Minimum value is 10 mV; accuracy is the same as voltage source. |
Current5 |
Minimum value is 10 nA; accuracy is the same as current source. |
Voltage Range |
Settling Time |
100 mV |
< 50 μs |
1 V |
< 50 μs |
6 V |
< 110 μs |
40 V6 |
< 150 μs |
Current Range |
Settling Time |
100 nA |
< 20 ms |
1 μA |
< 2 ms |
10 μA |
< 500 μs |
100 μA |
< 150 μs |
1 mA |
< 100 μs |
10 mA to 1 A |
< 80 μs (RLOAD > 6 Ω) |
3 A |
< 80 μs (Current < 2.5 A, RLOAD > 2 Ω) |
Normal Mode |
10 nF. |
High Capacitance Mode |
50 μF. |
Overrange |
101% of source range, 102% of measure range. |
Maximum Sense Lead Resistance |
1 kW for rated accuracy. |
Sense High Input Impedance |
>10 GW. |
IEEE-488 |
IEEE-488.1 compliant. Supports IEEE Std 488.2 common commands and status model topology. |
USB Control (rear) |
USB 2.0 device, USB-TMC488 protocol. |
RS-232 |
Baud rates from 300 bps to 115200 bps. |
Ethernet |
RJ-45 connector, 10/100BaseT, Auto-MDIX. |
Expansion Interface |
The TSP-Link expansion interface allows TSP-enabled instruments to trigger and communicate with each other. Cable Type: header 5e or higher LAN crossover cable. 3 meters (9.84 ft) maximum between each TSP-enabled instrument. A maximum of 32 TSP-Link nodes can be interconnected. Each source-measure instrument uses one TSP-Link node. |
LXI Compliance |
Version 1.5 LXI Device Specification 2016 compliant. |
LXI Timing |
Total Output Trigger Response Time: 245 μs minimum, 280 μs typical, (not specified) maximum. Receive LAN[0-7] Event Delay: Unknown. Generate LAN[0-7] Event Delay: Unknown. |
Connector |
25-pin female D. |
Input/Output Pins |
14 open drain I/O bits. |
Absolute Maximum Input Voltage |
5.25 V. |
Absolute Minimum Input Voltage |
–0.25 V. |
Maximum Logic Low Input Voltage |
0.7 V, +850 μA max. |
Minimum Logic High Input Voltage |
2.1 V, +570 μA. |
Maximum Source Current (flowing out of Digital I/O bit) |
+960 μA. |
Maximum Sink Current @ Maximum Logic Low Voltage (0.7V) |
–5.0 mA. |
Absolute Maximum Sink Current (flowing into Digital I/O pin) |
–11 mA. |
5V Power Supply Pins |
Limited to 250 mA total for all three pins, solid-state fuse protected. |
USB File System (Front) |
USB 2.0 Host: Mass storage class device. |
Power Supply |
100 V to 240 VAC, 50–60 Hz (auto sensing), 240 VA maximum. |
Cooling |
Forced air. Side intake and rear exhaust. One side must be unobstructed when rack mounted. |
EMC |
Conforms to European Union EMC Directive. |
Safety |
NRTL listed to UL61010-1:2008 and CSA C22.2 No. 61010-1. Conforms to European Union Low Voltage Directive. |
2601B-PULSE only |
105 mm × 235 mm × 445 mm (4.1 in. high × 9.25 in. wide × 17.5 in. deep). |
2601B-PULSE with 2601B-P-INT attached |
105 mm × 235 mm × 503 mm (4.1 in. high × 9.25 in. wide × 19.82 in. deep). |
2601B-PULSE only |
5.9 kg (13 lb). |
2601B-PULSE with 2601B-P-INT attached |
6.4 kg (14 lb). |
Altitude |
Maximum 2000 meters above sea level. |
Operating |
0 °C to 35 °C at up to 70% relative humidity; at 35 °C to 50 °C, derate 3% relative humidity per °C. |
Storage |
–25 °C to 65 °C. |