Voltage Version |
220V |
Input Bias Current |
<30pA at 25°C |
Input Protection |
1000V all ranges (2W input) |
AC CMRR |
1000V all ranges (2W input) |
Power Supply |
100V/120V/220V/240V |
Power Line Frequency |
50/60Hz auto detected |
Power Consumption |
25VA max |
Digital I/O Interface |
USB-compatible Type B connection, GPIB (option) |
Environment |
For indoor use only |
Operating Temperature |
0° to 40°C |
Operating Humidity |
Maximum relative humidity 80% for temperature up to 31°C |
Storage Temperature |
–40° to 70°C |
Operating Altitude up To 2000 M above sea level |
Bench Dimensions (with handles and bumpers) |
107 mm high × 252.8 mm wide × 305 mm deep (3.49 in. × 9.95 in. × 12.00 in.) |
Weight |
2.23 kg ( 4.92 lbs.) |
Safety |
Conforms to European Union Low Voltage Directive, EN61010-1. Measurement Cat 1 1000V and CAT II 600V |
EMC |
Conforms to European Union Directive 89/336/EEC, EN61326-1 |
Warranty |
One year |
DC Voltage |
Range |
Resolution |
Input Resistance |
Accuracy
±(% of reading + % of range)
1 Year, 23° ±5°C |
Temperature
Coefficient
0°–18°C & 28°–40°C |
100.000 mV |
1 µV |
|
0.012 + 0.004 |
0.001 + 0.0005 |
1.00000 V |
10 µV |
|
0.012 + 0.001 |
0.0009 + 0.0005 |
10.0000 V |
0.1 mV |
10 MO |
0.012 + 0.002 |
0.0012 + 0.0005 |
100.000 V |
1 mV |
|
0.012 + 0.002 |
0.0012 + 0.0005 |
1000.00 V |
10 mV |
|
0.02 + 0.003 |
0.002 + 0.0015 |
DCI (DC Current) |
Range |
Resolution |
Shunt Resistance |
Accuracy
±(% of reading + % of range)
1 Year, 23° ±5°C |
Temperature
Coefficient
0°–18°C & 28°–40°C |
10.0000 mA |
0.1 µA |
5.1 Ω |
0.05 + 0.020 |
0.005 + 0.002 |
100.000 mA |
1 µA |
5.1 Ω |
0.05 + 0.010 |
0.005 + 0.001 |
1.00000 A |
10 µA |
0.1 Ω |
0.150 + 0.020 |
0.008 + 0.001 |
3.0000 A |
100 µA |
0.1 Ω |
0.200 + 0.030 |
0.008 + 0.001 |
10.0000 A |
100 µA |
5 Ω |
0.250 + 0.050 |
0.008 + 0.001 |
Resistance |
Range |
Resolution |
Test Current |
Accuracy
±(% of reading + % of range)
1 Year, 23° ±5°C |
Temperature
Coefficient
0°–18°C & 28°–40°C |
100.000 Ω |
1 mΩ |
1 mA |
0.020 + 0.020 |
0.003 + 0.0005 |
1.00000 kΩ |
10 mΩ |
1 mA |
0.020 + 0.003 |
0.003 + 0.0005 |
10.0000 kΩ |
100 mΩ |
100 µA |
0.020 + 0.002 |
0.003 + 0.0005 |
100.000 kΩ |
1 Ω |
10 µA |
0.020 + 0.002 |
0.003 + 0.0005 |
1.00000 MΩ |
10 Ω |
1 µA |
0.030 + 0.004 |
0.005 + 0.0005 |
10.0000 MΩ |
100 Ω |
0.1 µA |
0.200 + 0.004 |
0.05 + 0.0005 |
100.000 MΩ |
1 kΩ |
0.1 µA |
2.000 + 0.005 |
0.5 + 0.0005 |
Diode Test |
Range |
Resolution |
Test Current |
Accuracy
±(% of reading + % of range)
1 Year, 23° ±5°C |
Temperature
Coefficient
0°–18°C & 28°–40°C |
1.0000V |
10 µV |
1 mA |
0.020 + 0.030 |
0.002 + 0.0005 |
Continuity |
Range |
Resolution |
Test Current |
Accuracy
±(% of reading + % of range)
1 Year, 23° ±5°C |
Temperature
Coefficient
0°–18°C & 28°–40°C |
1000Ω |
10 mΩ |
1 mA |
0.020 + 0.020 |
0.002 + 0.0005 |
CMRR |
120dB for 1kΩ unbalance in LO lead |
NMRR |
60dB for line frequency ±0.1% |
Thermocouple Type |
Range |
Accuracy 1 ±°C
1 Year, exclusive of lead accuracy |
B |
600 to 1800°C |
1.5 |
C |
0 to 2300°C |
1.5 |
E |
–250 to 1000°C |
1.5 |
J |
–200 to 1200°C |
1.0 |
K |
–200 to 1350°C |
1.0 |
N |
–200 to 1300°C |
1.0 |
R |
0 to 1750°C |
1.5 |
S |
0 to 1750°C |
1.5 |
T |
–250 to 400°C |
1.5 |
RTD and NTC Thermistor Measurements |
Accuracy ±0.8°C, 1 year, exclusive of lead accuracy. PT100, D100, F100, PT385, PT3916, SPRTD (R-Zero, A4, B4, Ax, Bx, Cx, and Dx), NTCT (A, B, and C), and user-definable RTD. |
Range |
Test Current |
Accuracy
±(% of reading + % of range)
1 Year, 23° ±5°C |
1.000 nF |
10 µA |
2.0 + 0.80 |
10.00 nF |
10 µA |
1.0 + 0.50 |
100.0 nF |
100 µA |
1.0 + 0.50 |
1.000 nF |
100 µA |
1.0 + 0.50 |
10.00 nF |
100 µA |
1.0 + 0.50 |
100.0 nF |
1 mA |
1.0 + 0.50 |
DC Voltage |
Range |
Frequency (Hz) |
Accuracy
±(% of reading + % of range)
1 Year, 23° ±5°C |
Temperature
Coefficient
0°–18°C & 28°–40°C |
100.000 mV to
750.000 V |
10–40 |
0.03 |
0.002 |
40–300k |
0.02 |
0.002 |
ACV (AC TRMs Voltage) |
Range |
Resolution |
Frequency |
Accuracy
±(% of reading + % of range)
1 Year, 23° ±5°C |
Temperature
Coefficient
0°–18°C & 28°–40°C |
100.000 mV
to
750.000 V |
1 µV
to
10 mV |
10 Hz–20 kHz |
0.12 + 0.05 |
0.01 + 0.01 |
20 kHz–50 kHz |
0.25 + 0.05 |
0.02 + 0.02 |
50 kHz–100 kHz |
0.65 + 0.08 |
0.04 + 0.02 |
100 kHz–300 kHz |
5.00 + 0.50 |
0.2 + 0.02 |
ACI (AC TRMs Current) |
Range |
Resolution |
Frequency |
Accuracy
±(% of reading + % of range)
1 Year, 23° ±5°C |
Temperature
Coefficient
0°–18°C & 28°–40°C |
1.0000 A to
3.00000 A |
10 µA to
100 µA |
10 Hz–900 Hz |
0.30 + 0.06 |
0.02 + 0.01 |
900 Hz–5 kHz |
1.50 + 0.15 |
0.02 + 0.01 |
10.0000 A |
100 µA |
10 Hz–900 Hz |
0.50 + 0.12 |
0.02 + 0.01 |
900 Hz–5 kHz |
2.50 + 0.20 |
0.02 + 0.01 |