POWER |
Voltage |
90–134V and 180–250V, universal self-selecting |
Frequency |
50Hz, 60Hz, or 400Hz self-identifying at power-up. |
Consumption |
<55VA |
ENVIRONMENTAL |
Operating Temperature |
0° to 50°C |
Storage Temperature |
–40° to 70°C |
Humidity |
80% R.H., 0° to 35°C |
PHYSICAL |
Case Dimensions |
90mm high × 214mm wide × 369mm deep (3½ in. × 8½ in. × 14½ in.) |
Net Weight |
<4.2kg (<9.2 lbs.) |
Shipping Weight |
<9.1kg (<20 lbs.) |
STANDARDS |
EMI/RFI |
Conforms to European Union EMC directive |
Safety |
Conforms to European Union Low Voltage directive |
Note 1 |
For MIL-T-28800E, applies to Type III, Class 5, Style E |
DC VOLTS
DCV INPUT CHARACTERISTICS AND ACCURACY |
Enhanced Accuracy – 10PLC, DFILT 10 |
Range |
Full
Scale |
Resolution |
Input
Resistance |
Accuracy ±(ppm of reading + ppm of range) |
Transfer |
24
Hours |
90
Days |
1
Year |
2
Years |
200mV |
±210.00000mV |
1 nV |
>100 GΩ |
0.4+1.5 |
3.5+3 |
15+8 |
19+9 |
23+10 |
2V |
±2.1000000V |
10nV |
>100GΩ |
0.2+0.15 |
1.2+0.3 |
6+0.8 |
10+0.9 |
14+1 |
20V |
±21.000000V |
100nV |
>100GΩ |
0.1+0.05 |
1.2+0.1 |
6+0.15 |
10+0.15 |
14+0.15 |
200V |
±210.00000V |
1µV |
10MΩ±1% |
0.5+0.08 |
5+0.4 |
14+2 |
22+2 |
30+2 |
1000V |
±1100.0000V |
10µV |
10MΩ±1% |
1+0.05 |
5+0.08 |
14+0.4 |
22+0.4 |
30+0.4 |
Range |
Full
Scale |
Resolution |
Input
Resistance |
Accuracy ±(ppm of reading + ppm of range) |
24
Hours |
90
Days |
1
Year |
2
Years |
200mV |
±210.00000mV |
10 nV |
>100 GΩ |
3.5+6 |
15+11 |
19+12 |
23+13 |
2V |
±2.1000000V |
100nV |
>100GΩ |
1.2+0.6 |
6+1.1 |
10+1.2 |
14+1.3 |
20V |
±21.000000V |
100nV |
>100GΩ |
3.2+0.35 |
8+0.4 |
12+0.4 |
16+0.4 |
200V |
±210.00000V |
10µV |
10MΩ±1% |
5+1.2 |
14+2.8 |
22+2.8 |
30+2.8 |
1000V |
±1100.0000V |
100µV |
10MΩ±1% |
5+0.04 |
14+0.7 |
22+0.7 |
30+0.7 |
- Specifications are for 10 power line cycles, synchronous autozero, 10-reading repeat digital filter, auto range off, except as noted
- For TCAL ±1°C, following 4-hour warm-up. TCAL is ambient temperature at calibration (23°C at the factory). Add 0.5ppm of reading uncertainty if the unit is power cycled during this interval
- For TCAL ±5°C, following 4-hour warm-up
- Care must be taken to minimize thermal offsets due to operator cables
- Specifications apply for 20-reading repeat digital filter, TREF ±0.5°C (TREF is the initial ambient temperature), and for measurements within 10% of the initial measurement value and within 10 minutes of the initial measurement time.
- Add 20ppm × (VIN/1000V) 2 additional uncertainty for inputs above 200V, except in transfer accuracy specifications
- Specifications are for 1 power line cycle, normal autozero, digital filter off, autorange off.
|
Range |
Normal Mode RMS
90 Days, ±2°C from last AC self-cal for 1% to 100% of range
±(% of reading + % of range) |
20–50Hz |
50–100Hz |
0.1–2kHz |
2–10kHz |
10–30kHz |
200 mV |
0.25 + 0.015 |
0.07 + 0.015 |
0.02 + 0.02 |
0.02 + 0.02 |
0.025 + 0.02 |
2 V |
0.25 + 0.015 |
0.07 + 0.015 |
0.02 + 0.02 |
0.02 + 0.02 |
0.025 + 0.02 |
20 V |
0.25 + 0.015 |
0.07 + 0.015 |
0.03 + 0.015 |
0.04 + 0.015 |
0.05 + 0.015 |
200 V |
0.25 + 0.015 |
0.07 + 0.015 |
0.03 + 0.015 |
0.04 + 0.015 |
0.05 + 0.015 |
750 V |
0.25 + 0.015 |
0.1 + 0.015 |
0.05 + 0.015 |
0.06 + 0.015 |
0.08 + 0.015 |
- Specifications apply for sinewave input, AC + DC coupling, 1 power line cycle, autozero on, digital filter off, following 55-minute warm-up
- For 1% to 5% of range below 750V range, and for 1% to 7% of 750V range, add 0.01% of range uncertainty. For inputs from 200kHz to 2MHz, specifications apply above 10% of range.
- Add 0.001% of reading × (VIN/100V) 2 additional uncertainty for inputs above 100V rms.
- Typical values
|
Range |
Normal Mode RMS
90 Days, ±2°C from last AC self-cal for 1% to 100% of range
±(% of reading + % of range) |
30–50kHz |
50–100kHz |
100–200kHz |
0.2–1MHz |
1–2MHz |
200 mV |
0.05 + 0.01 |
0.3 + 0.015 |
0.75 + 0.025 |
2 + 0.1 |
5 + 0.2 |
2 V |
0.05 + 0.01 |
0.3 + 0.015 |
0.75 + 0.025 |
2 + 0.1 |
5 + 0.2 |
20 V |
0.07 + 0.015 |
0.3 + 0.015 |
0.75 + 0.025 |
4 + 0.2 |
7 + 0.2 |
200 V |
0.07 + 0.015 |
0.3 + 0.015 |
0.75 + 0.025 |
4 + 0.2 |
|
750 V |
0.1 + 0.015 |
0.5 + 0.015 |
|
|
|
- Specifications apply for sinewave input, AC + DC coupling, 1 power line cycle, autozero on, digital filter off, following 55-minute warm-up
- For 1% to 5% of range below 750V range, and for 1% to 7% of 750V range, add 0.01% of range uncertainty. For inputs from 200kHz to 2MHz, specifications apply above 10% of range.
- Add 0.001% of reading × (VIN/100V) 2 additional uncertainty for inputs above 100V rms.
- Typical values
|
Range |
ACI ACCURACY
90 Days, 1 Year or 2 Years, TCAL ±5°C, for 5% to 100% of range, ±(% of reading + % of range) |
20Hz–
50Hz |
50Hz–
200Hz |
200Hz–
1kHz |
1kHz–
10kHz |
10kHz–
30kHz |
30kHz–
50kHz |
50kHz–
100kHz |
200 µA |
0.35+0.015 |
0.2+0.015 |
0.4+0.015 |
0.5+0.015 |
|
|
|
2 mA |
0.3+0.015 |
0.15+0.015 |
0.12+0.015 |
0.12+0.015 |
0.25+0.015 |
0.3+0.015 |
0.5+0.015 |
20 mA |
0.3+0.015 |
0.15+0.015 |
0.12+0.015 |
0.12+0.015 |
0.25+0.015 |
0.3+0.015 |
0.5+0.015 |
200 mA |
0.3+0.015 |
0.15+0.015 |
0.12+0.015 |
0.15+0.015 |
0.5+0.015 |
1+0.015 |
3+0.015 |
2 A |
0.35+0.015 |
0.2+0.015 |
0.3+0.015 |
0.45+0.015 |
1.5+0.015 |
4+0.015 |
|
- Specifications apply for sinewave input, AC+DC coupling, 1 power line cycle, autozero on, digital filter off, following 55-minute warm-up
- Add 0.005% of range uncertainty for current above 0.5A rms for self-heating
- Typical values
|
AC VOLTAGE INPUT |
1Hz–15MHz |
ACCURACY |
±(0.03% of reading) |
TYPICAL RANGES |
Current |
100µA to 12A |
Trace Resistance |
1mΩ to 10Ω typical |
ACCURACY |
±(5% + 500µA). For 1 power line cycle, autozero on, 10-reading digital filter, TCAL ±5°C, 90 days, 1 year or 2 years |
Built-in linearization for J, K, N, T, E, R, S, B thermocouple types to ITS-90 and 100O platinum RTDs DIN 43760, IPTS-68, and ITS-90. |