| Main Unit Specifications : |
| Model |
6332 / 6332D |
| Measuring Object |
Clean air and compatible, non-combustible gases |
| Measuring Range |
Maximum range 0.1 ~ 50 m/s (Range dependant on probe type*) |
| Measuring Accuracy |
± (3% of reading + 0.1) m/s |
| Temperature Compensation Range |
|
0.1-4.99m/s |
5.00-9.99m/s |
10.0-24.9m/s |
25.0-50.0m/s |
| 5~40°C |
±0.25m/s |
±0.25m/s |
±1.25m/s |
±1.25m/s |
| 40~80°C |
±0.35m/s |
±0.70m/s |
±1.75m/s |
±3.50m/s |
| Selectable Output Option |
(Set via onboard DIPswitch).
Current output: DC 4~20mA (Max. load resistance: 250Ω)
Voltage output: DC 0~5V |
| Selectable Range Option |
(Rotary Switch selectable).
0~2, 0~5, 0~10, 0~25, or 0~50 m/s |
| Display Resolution (Model 6332D only) |
Velocity Range: 0~2, 0~5, 0~10 m/s ? Display Resolution: 0.01m/s
Velocity Range: 0~25, 0~50 m/s ? Display Resolution: 0.1m/s |
| Power Supply |
DC 12~24V |
| Power Consumption (Reference Value) |
Approx. 2.0W (Under the following conditions - power supply: 12V, air velocity: approx. 10m/s, using a unit with a display with probe model 0965-03.) Power consumption rate is subject to change according to conditions such as air velocity, probe type and use of display. |
| Connection Wire Size |
0.5 ~ 1.5 mm2 |
| Operating Temperature Range |
5 ~ 40°C |
| Storage Temperature Range |
-10 ~ 50°C |
| Dimensions |
Approx. 128(W)×78(D)×30(H) mm |
| Weight |
Approx. 320g (11oz.) |
| Standard Accessory |
Operation Manual : 1, Main Unit Case: 1 |
| Optional Accessory |