Julabo PRESTO W80t General Specifications |
Category |
Temperature Control PRESTO |
Working temperature range (°C) |
-80 ... +250 |
Temperature control |
ICC |
Temperature stability (°C) |
±0.01 ... ±0.05 |
Setting / display resolution |
0.01 °C |
Integrated programmer |
8x60 steps |
Temperature Display |
TFT Touchscreen |
Heating capacity (kW) |
2.8 |
Cooling capacity (Medium: JULABO Thermal | Ethanol) |
°C |
200 |
20 |
0 |
-20 |
-40 |
-60 |
-80 |
kW |
1.2 |
1.2 |
1.2 |
1.1 |
1.1 |
0.65 |
0.1 |
|
Pump capacity flow rate (l/min) |
16 ... 40 |
Pump capacity flow pressure (psi) |
1.45 ... 18.85 |
Pump connections |
M24x1.5 |
Refrigerant stage 1 |
R507 |
Filling volume refrigerant stage 1 (g) |
720 |
Global Warming Potential for |
3985 |
Carbon dioxide equivalent stage 1 (t) |
2.869 |
Refrigerant stage 2 |
R23 |
Filling volume refrigerant stage 2 (g) |
500 |
Global Warming Potential for |
14800 |
Carbon dioxide equivalent stage 2 (t) |
7.4 |
External Pt100 sensor connection |
integrated |
Digital interfaces |
RS232
SD memory card
USB
Ethernet
Modbus
Alarm-out
Optional: RS485, Profibus |
Analog connection input / output |
Optional |
Ambient temperature |
5 ... 40 °C |
Dimensions W x L x H (inch) |
16.9 x 25.6 x 49.5 |
Weight (LBS) |
357 |
Sound pressure level (distance 1 m) max. (dBA) |
64 |
Process volume min. (active heat exchanger volume) liters |
3.9 (1.7) |
Internal usable expansion vol. (liters) |
5.6 |
Classification according to DIN12876-1 |
Classification III (FL) |
Cooling of compressor |
2-stage Water |
Cooling water connection |
G ¾‘‘ male with barbed fittings for tubing ½’’ lD |
Cooling water consumption (l/min) |
2 |
Cooling water temperature (°C) |
<30 |
Cooling water differential pressure (bar) |
0.5 |
Power requirement V / Hz / A |
3x 208-220/60/15 Additional voltage versions available |
Available voltage versions |
3 x 208-220V/60Hz (+/- 5%) / 18A @ 208V, 20A @ 220V / Without Plug
3 x 230V/50Hz (+/- 10%) / 20A / Plug 32A CEE
3 x 400V/50Hz (+/- 10%) / 13A / Plug 16A CEE |