| Specifications |
| Detector |
Halogen-quenched Geiger-Mueller tube. Effective diameter 1.75″ (45 mm). Mica window density 1.5-2.0 mg/cm²; Detects Alpha, Beta, Gamma, and X-radiation |
| Display |
128×128 color OLED (Organic Light Emitting Diode) |
| Operating Range |
µSv/hr: .000 to 1,000::: µR/hr: .000 to 100,000 ::: CPM: 0 to 350,000 |
| Accuracy |
± 10% typical; ±15% maximum |
| Alert Range |
0 to 99,999 CPM; Beeper sounds the alert when warning level feature is activated |
| Anti-Saturation |
Readout holds at full scale in fields up to 100 times the maximum reading |
| Audio |
Beeper chirps with each count when Geiger Beep function is activated |
| Averaging |
Accumulate/Average feature builds average over time when activated |
| Certifications |
CE Certified: Emissions: EN 55011:2009 + A1:2010 (Class B emissions limits); EN 61000-4-2:1995 (ESD); EN61000-4-3:1997. RoHS Compliant, Meets WEEE standards |
| Count Light |
Red LED flashes with each count |
| Calibration |
Cesium-137 (gamma from daughter metastable Barium) |
| Gamma Sensitivity |
334 CPM per µSv/hr (3340 CPM per mR/hr) referenced to Cs-137 |
| Efficiency |
For 4 pi at contact |
| Beta C-14 (49 keV avg. 156 keV max.) |
5.3% Bi-210 (390 keV avg. 1.2 MeV max.): 32% Sr-90 (546 keV and 2.3 MeV): 38% P-32 (693 keV avg. 1.7 MeV max.): 33% |
| Alpha Am-241 (5.5 MeV) |
18% |
| I-125 |
Smallest detectable level for I-125 is .02 mCi at contact |
| Manual |
ONYX Operation Manual |
| Options |
USB cable, charger and carrying case available |
| Ports: Output |
Stereo 3.5 mm jack sends pulses to headset or external data collection devices |
| Power |
One 3.7 V Lithium Ion battery |
| Size |
130 x 660 x 230 mm (5.1 x 2.6 x 0.9 inches) |
| Temperature Range |
-20º to +50º C , -4º to +122º F |
| Timer |
Can set sampling periods of 1 minute to 40 hours. |
| Weight |
200 grams (7 oz) including battery |