AC WITHSTANDING (all models) |
Output-Voltage Range |
0.100kV~5.000kV ac |
Output-Voltage Resolution |
2V/step |
Output-Voltage Accuracy |
± (1% of setting + 5V) [no load] |
Maximum Rated Load |
500 VA (5kV/100mA) |
Maximum Rated Current |
100mA (0.5kV< V 5kV)
10mA (0.1kV V 0.5kV) |
Output-Voltage Waveform |
Sine wave |
Output-Voltage Frequency |
50 Hz / 60 Hz selectable |
Voltage Regulation |
± (1% of reading + 5V) [full load → no load] |
Voltmeter Accuracy |
± (1% of reading + 5V) |
Current Measurement Range |
0.001mA~100.0mA |
Current Best Resolution |
0.001mA / 0.01mA /0.1mA |
Current Measurement Accuracy |
±(1.5% of reading + 30 counts) when HI SET < 1.11mA
±(1.5% of reading + 3 counts) when HI SET 1.11mA |
Window Comparator Method |
Yes |
ARC Detect |
Yes |
Sweep Function* |
Yes |
RAMP (Ramp-Up Time) |
0.1s~999.9s |
TIMER (Test Time)* |
OFF, 0.5s~999.9s |
GND |
ON/OFF |
DC WITHSTANDING (GPT-9902A/9903/9904) |
Output-Voltage Range |
0.100kV~6.000kV dc |
Output-Voltage Resolution |
2V/step |
Output-Voltage Accuracy |
± (1% of setting + 5V) [no load] |
Maximum Rated Load |
100W (5kV/20mA) |
Maximum Rated Current |
20mA (0.5kV< V 6kV)
2mA (0.1kV V 0.5kV) |
Voltage Regulation |
± (1% of reading + 5V) [full load → no load] |
Voltmeter Accuracy |
± (1% of reading + 5V) |
Current Measurement Range |
0.001mA~20.0mA |
Current Best Resolution |
0.001mA /0.01mA /0.1mA |
Current Measurement Accuracy |
±(1.5% of reading + 30 counts) when HI SET < 1.11mA:
±(1.5% of reading + 3 counts) when HI SET 1.11mA |
Window Comparator Method |
Yes |
ARC Detect |
Yes |
Sweep Function* |
Yes |
RAMP (Ramp-Up Time) |
0.1s~999.9s |
TIMER (Test Time)* |
OFF, 0.5s~999.9s |
GND |
ON/OFF |
INSULATION RESISTANCE (GPT-9903/9904) |
Output Voltage |
50V~1000V dc |
Output-Voltage Resolution |
50V/step |
Output-Voltage Accuracy |
± (1% of setting + 5V) [no load] |
Resistance Measurement Range |
0.001GO~ 50.00GO |
Test Voltage & Measurement Accuracy |
50VV 450V
0.001~0.050GO:±(5% of reading + 1 count)
0.051~2.000GO:±(10% of reading + 1 count) |
500VV 1000V
0.001~0.500GO:±(5% of reading + 1 count)
0.501~9.999GO:±(10% of reading + 1 count)
10.00~50.00GO:±(15% of reading + 1 count)
|
Output Impedance |
600kΩ |
Window Comparator Method |
Yes |
Sweep Function* |
Yes |
RAMP (Ramp-Up Time) |
0.1s~999.9s |
TIMER (Test Time) |
1s~999.9s |
GND |
OFF (fixed) |
Ground Bond (GPT-9904 only) |
Output-Current |
03.00A~32.00A ac |
Output-Current Resolution |
0.01A |
Output-Current Accuracy |
3AI8A:±(1% of reading + 0.2A)
8A |
Test-Voltage |
6Vac max (open circuit) |
Test-Voltage Frequency |
50Hz/60Hz selectable |
Resistance Measurement Range |
10.0mO~ 650.0mO |
Resistance Measurement Resolution |
0.1mO |
Resistance Measurement Accuracy |
±(1% of reading + 2 mO) |
Window Comparator Method |
Yes |
Sweep Function* |
Yes |
TIMER (Test Time) |
0.5s~999.9s |
Test Method |
Four Terminal |
GND |
OFF (fixed) |
MEMORY |
Single Step Memory |
MANU:100 blocks |
Automatic Testing Memory |
AUTO:100 blocks, manu per auto:16 |
INTERFACE |
Rear Output |
Standard |
RS-232C |
Standard |
USB |
Standard |
GPIB |
Option |
Remote terminal (Front) |
Standard |
Signal I/O |
Standard |
DISPLAY |
240 x 64 Ice Blue Dot matrix LCD |
POWER SOURCE |
AC 100V/120V/220V/230V ± 10%, 50/60Hz |
DIMENSION & WEIGHT |
330(W) x 148(H) x 587(D) mm for GPT-9904/9903
330(W) x 148(H) x 482(D) mm for GPT-9902A/9901A
Approx. 27kg max. |