Model |
|
PSU |
6-200 |
12.5-120 |
20-76 |
40-38 |
60-25 |
100-15 |
150-10 |
300-5 |
400-3.8 |
600-2.6 |
Output ratings |
Rated output voltage (*1) |
|
V |
6 |
12.5 |
20 |
40 |
60 |
100 |
150 |
300 |
400 |
600 |
Rated output current (*2) |
|
A |
200 |
120 |
76 |
38 |
25 |
15 |
10 |
5 |
3.8 |
2.6 |
Rated output power |
|
W |
1200 |
1500 |
1520 |
1520 |
1500 |
1500 |
1500 |
1500 |
1520 |
1560 |
Ripple and noise(*5) |
CVp-p( 10 ~ 20MHz) |
p-p (*6) |
mV |
60 |
60 |
60 |
60 |
60 |
80 |
100 |
150 |
200 |
300 |
CVrms(5Hz ~ 1MHz) |
r.m.s. (*7) |
mV |
8 |
8 |
8 |
8 |
8 |
8 |
10 |
25 |
40 |
60 |
CCrms(5Hz ~ 1MHz) |
r.m.s.(*12) |
mA |
400 |
240 |
152 |
95 |
75 |
45 |
35 |
25 |
17 |
12 |
Load regulation |
Voltage(*4) |
|
mV |
2.6 |
3.25 |
4 |
6 |
8 |
12 |
17 |
32 |
42 |
62 |
Current(*11) |
|
mA |
45 |
29 |
20.2 |
12.6 |
10 |
8 |
7 |
6 |
5.76 |
5.52 |
Line regulation |
Voltage(*3) |
|
mV |
2.6 |
3.25 |
4 |
6 |
8 |
12 |
17 |
32 |
42 |
62 |
Current(*3) |
|
mA |
22 |
14 |
9.6 |
5.8 |
4.5 |
3.5 |
3 |
2.5 |
2.38 |
2.26 |
Analog Programming and Monitoring |
External voltage control output voltage |
|
|
Accuracy and linearity: ±0.5% of rated output voltage. |
External voltage control output current |
|
|
Accuracy and linearity: ±1% of rated output current. |
External resistor control output voltage |
|
|
Accuracy and linearity: ±1% of rated output voltage. |
External resistor control output current |
|
|
Accuracy and linearity: ±1.5% of rated output current. |
Output voltage monitor |
|
|
Accuracy: ±1% |
Output current monitor |
|
|
Accuracy: ±1% |
Shutdown control |
|
|
Turns the output off with a LOW (0V to 0.5V) or short-circuit. |
Output on/off control |
|
|
Possible logic selections: |
Turn the output on using a LOW (0V to 0.5V) or short-circuit, turn the output off using a HIGH (4.5V to 5V) or open-circuit. |
Turn the output on using a HIGH (4.5V to 5V) or open-circuit, turn the output off using a LOW (0V to |
Alarm clear control |
|
|
Clear alarms with a LOW (0V to 0.5V) or short-circuit. |
CV/CC/ALM/PWR ON/OUT ON indicator |
|
|
Photocoupler open collector output; Maximum voltage 30V, maximum sink current 8mA. |
Trigger out |
|
|
Maximum low level output = 0.8V; minimum high level output = 2V; Maximum source current = 8mA. |
Trigger in |
|
|
Maximum low level input voltage = 0.8V; minimum high level input votage = 2V, Maximum sink current = 8mA. |
Front Panel |
Display, 4 digits, Voltage accuracy |
0.1% + |
mV |
12 |
25 |
40 |
80 |
120 |
200 |
300 |
600 |
800 |
1200 |
Current accuracy |
0.2% + |
mA |
600 |
360 |
228 |
114 |
75 |
45 |
30 |
15 |
11.4 |
7.8 |
Indications |
|
|
GREEN LED's: CV, CC, V, A, VSR, ISR, DLY, RMT, LAN, M1, M2, M3, RUN, Output ON |
|
|
|
RED LED's: ALM, ERR |
Buttons |
|
|
Lock/Local(Unlock), PROT(ALM_CLR), Function(M1), Test(M2), Set(M3), Shift, Output |
Knobs |
|
|
Voltage, Current |
USB port |
|
|
Type A USB connector |
Transient response time (*10) |
Transient response time |
|
ms |
1.5 |
1 |
1 |
1 |
1 |
1 |
2 |
2 |
2 |
2 |
Output response time |
Rise time (*8) |
Rated load |
ms |
80 |
80 |
80 |
80 |
80 |
150 |
150 |
150 |
200 |
250 |
|
No load |
ms |
80 |
80 |
80 |
80 |
80 |
150 |
150 |
150 |
200 |
250 |
Fall time(*9) |
Rated load |
ms |
10 |
50 |
50 |
80 |
80 |
150 |
150 |
150 |
200 |
250 |
|
No load |
ms |
500 |
700 |
800 |
1000 |
1100 |
1500 |
2000 |
2500 |
3000 |
4000 |
Programming and Measurements (RS-232/485, USB, LAN, GPIB) |
Output voltage programming accuracy |
0.05% + |
mV |
3 |
6.25 |
10 |
20 |
30 |
50 |
75 |
150 |
200 |
300 |
Output current programming accuracy |
0.2% + |
mA |
200 |
120 |
76 |
38 |
25 |
15 |
10 |
5 |
3.8 |
2.6 |
Output voltage programming resolution |
|
mV |
0.2 |
0.4 |
0.7 |
1.3 |
2 |
3.4 |
5.2 |
10.2 |
13.6 |
20.4 |
Output current programming resolution |
|
mA |
6 |
4 |
2.5 |
1.2 |
0.8 |
0.5 |
0.34 |
0.19 |
0.13 |
0.09 |
Output voltage measurement accuracy |
0.1% + |
mV |
6 |
12.5 |
20 |
40 |
60 |
100 |
150 |
300 |
400 |
600 |
Output current measurement accuracy |
0.2% + |
mA |
400 |
240 |
152 |
76 |
50 |
30 |
20 |
10 |
7.6 |
5.2 |
Output voltage measurement resolution |
|
mV |
0.2 |
0.4 |
0.7 |
1.3 |
2 |
3.4 |
5.2 |
10.2 |
13.6 |
20.4 |
Output current measurement resolution |
|
mA |
6 |
4 |
2.5 |
1.2 |
0.8 |
0.5 |
0.34 |
0.19 |
0.13 |
0.09 |
Temperature coefficience |
Voltage |
|
ppm/℃ |
100ppm/℃ after a 30 minute warm-up |
Current |
|
ppm/℃ |
100ppm/℃ after a 30 minute warm-up |
Remote sense compensation voltage(single wire) |
voltage |
|
V |
1 |
1 |
1 |
2 |
3 |
5 |
5 |
5 |
5 |
5 |
Protection Function |
Over voltage protection(OVP) |
Setting range |
V |
0.6 - 6.6 |
1.25 - 13.75 |
42788 |
16163 |
24228 |
5 - 110 |
5 - 165 |
5 - 330 |
5 - 440 |
5 - 660 |
|
Setting accuracy |
mV |
60 |
125 |
200 |
400 |
600 |
1000 |
1500 |
3000 |
4000 |
6000 |
Over current protection(OCP) |
Setting range |
A |
5 - 220 |
5 - 132 |
5 - 83.6 |
3.8 - 41.8 |
2.5 - 27.5 |
1.5 - 16.5 |
42746 |
0.5 - 5.5 |
0.38 - 4.18 |
0.26 - 2.86 |
|
Setting accuracy |
mA |
4000 |
2400 |
1520 |
760 |
500 |
300 |
200 |
100 |
76 |
52 |
Under voltage limit(UVL) |
Setting range |
|
0 - 6.3 |
0 - 13.12 |
0 - 21 |
0 - 42 |
0 - 63 |
0 - 105 |
0 - 157.5 |
0 - 315 |
0 - 420 |
0 - 630 |
Over temperature protection(OHP) |
Operation |
|
Turn the output off. |
Incorrect sensing connection protection (SENSE) |
Operation |
|
Turn the output off. |
Low AC input protection (AC-FAIL) |
Operation |
|
Turn the output off. |
Shutdown (SD) |
Operation |
|
Turn the output off. |
Power limit (POWER LIMIT) |
Operation |
|
Over power limit. |
|
Value (fixed) |
|
Approx. 105% of rated output power |
Interface Capabilities |
USB |
|
|
TypeA: Host, TypeB: Slave, Speed: 1.1/2.0, USB Class: CDC(Communications Device Class) |
LAN |
|
|
MAC Address, DNS IP Address, User Password, Gateway IP Address, Instrument IP Address, Subnet Mask |
RS-232 / RS-485 |
|
|
Complies with the EIA232D / EIA485 Specifications |
GPIB (Factory Option) |
|
|
SCPI - 1993, IEEE 488.2 compliant interface |
Isolated Analog Control Interface (Factory Option) |
Voltage Control |
|
|
Using 0-5V or 0-10V signals for programming and measurement |
Current Control |
|
|
Using 4-20mA current signals for programming and measurement |
Environmental Conditions |
Operating temperature |
|
|
0˚C to 50˚C |
Storage temperature |
|
|
-25˚C to 70˚C |
Operating humidity |
|
|
20% to 85% RH; No condensation |
Storage humidity |
|
|
90% RH or less; No condensation |
Altitude |
|
|
Maximum 2000m |
Input Characteristics |
Nominal input rating |
|
|
100Vac to 240Vac, 50Hz to 60Hz, single phase |
Input voltage range |
|
|
85Vac ~ 265Vac |
Input frequency range |
|
|
47Hz ~ 63Hz |
Maximum input current |
100Vac / 200Vac |
A |
21 / 11 |
Inrush current |
|
|
Less than 50A. |
Maximum input power |
|
VA |
2000 |
Power factor |
100Vac / 200Vac |
|
0.99 / 0.98 |
Efficiency (*13) |
100Vac / 200Vac |
% |
77 / 79 |
82 / 85 |
83 / 86 |
84 / 87 |
84 / 87 |
84 / 87 |
84 / 87 |
84 / 87 |
84 / 87 |
84 / 87 |
Hold-up time |
|
|
20ms or greater |
General |
Weight |
main unit only |
kg |
Less than 8.7kg |
Dimensions |
(W×H×D) |
mm^3 |
423×43.6×447.2 |
Cooling |
|
|
Forced air cooling by internal fan. |
EMC |
|
|
Complies with the European EMC directive 89/336/EEC for Class A test and measurement products. |
Safety |
|
|
Complies with the European Low Voltage Directive 73/23/EEC and carries the CE-marking. |
Withstand voltage |
|
|
AC to Chassis : 1500Vac / 1min |
AC to Output terminal : 3000Vac / 1min |
Vout≦150V |
Output terminal to Chassis : 1000Vdc / 1min |
150 |
Output terminal to Chassis : 1500Vdc / 1min |
Insulation resistances |
|
|
Chassis and output terminal; chassis and AC input; AC input and output terminal: 100MΩ or more (DC 1000V) |