X-Y MODE |
X-Axis Input |
Channel 1 |
Y-Axis Input |
Channel 2 |
Phase Shift |
±3° at 100kHz |
CURSORS AND MEASUREMENT |
Cursors |
Amplitude, Time, Gating Available; Unit : Seconds(S), Hz(1/S), Phase (Degrees), Ratio(%) |
Automatic Measurement |
38 sets : Pk-Pk, Max, Min, Amplitude, High, Low, Mean, Cycle Mean, RMS, Cycle RMS, Area, Cycle Area, ROVShoot, FOVShoot, RPREShoot, FPREShoot, Frequency, Period, RiseTime, FallTime, +Width, -Width, Duty Cycle, +Pulses, -Pulses, +Edges, -Edges, %Flicker, Flicker Idx., FRR, FRF, FFR, FFF, LRR, LRF, LFR, LFF, Phase |
CONTROL PANEL FUNCTION |
Auto Counter |
6 digits, range from 2Hz minimum to the rated bandwidth |
Autoset |
Single-button, automatic setup of all channels for vertical, horizontal and trigger systems, with undo Autoset |
Save Setup |
20 sets |
Save Waveform |
24 sets |
DISPLAY SYSTEM |
TFT LCD Type |
8" TFT LCD WVGA color display |
Display Resolution |
800 horizontal x 480 vertical pixels (WVGA) |
Interpolation |
Sin(x)/x |
Waveform Display |
Dots, Vectors, Variable persistence(16ms~4s), Infinite persistence |
Waveform Update Rate |
120,000 waveforms per second, maximum |
Display Mode |
YT ; XY |
Display Graticule |
8 x 10 divisions |
INTERFACE |
USB Port |
USB 2.0 High-speed host port x 1, USB 2.0 High-speed device port x 1 |
Ethernet Port (LAN) |
RJ-45 connector, 10/100Mbps with HP Auto-MDIX |
Go/NoGo BNC |
5V Max/10mA TTL open collector output |
Kensington Style Lock |
Rear-panel security slot connects to standard Kensington-style lock |
SPECTRUM ANALYZER SPECIFICATIONS |
Frequency Range |
DC~1GHz(Max.) (Max. bandwidth ~1GHz uncalibrated) |
Span |
1kHz ~ 1GHz(Max.) |
Resolution Bandwidth |
1Hz ~ 1MHz(Max.) |
Reference Level |
-50 dBm to +40dBm in steps of 5dBm |
Vertical Units |
dBV RMS; Linear RMS; dBm |
Vertical Position |
-12divs to +12divs |
Vertical Scale |
1dB/div to 20dB/div in a 1-2-5 Sequence |
Display Average Noise Level |
1V/div < -50dBm, Avg : 16 ; 100mV/div < -70dBm, Avg : 16 ; 10mV/div < -90dBm, Avg : 16 |
Spurious Response |
2nd harmonic distortion< 40dBc ; 3rd harmonic distortion< 45dBc |
Frequency Domain |
Normal ; Max Hold ; Min Hold ; Average (2 ~ 256) |
Trace Types |
Detection Methods |
Sample ; +Peak ; -Peak ; Average |
FFT Windows |
FFT Factor : Hanning 1.44 ; Rectangular 0.89 ; Hamming 1.30 ; Blackman 1.68 |
FREQUENCY RESPONSE ANALYSIS (MDO-2000AG only) |
Dynamic Range |
> 80 dB (typical) |
Input and Output Sources |
Channel 1 or 2 |
Frequency Range |
20 Hz to 25 MHz |
Number of Test Points |
10 to 90 points per decade |
Test Amplitude |
20 mVpp to 5 Vpp into High-Z ; Fixed test amplitude or custom amplitude for each decade |
Test Results |
Logarithmic overlaid gain and phase plot |
Manual Measurements |
Two pairs of tracking gain and phase markers |
Plot Scaling |
Auto-scaled during test |
MISCELLANEOUS |
Line Voltage Range |
AC 100V ~ 240V, 50Hz ~ 60Hz, auto selection |
Multi-Language Menu |
Available |
On-Line Help |
Available |
Time Clock |
Time and date, provide the date/time for saved data |
Operation Environment |
Temperature: 0°C to 50°C. Relative Humidity: 80% at 40°C or below; 45%, 41°C ~ 50°C ≤ ≤ |
Dimensions & Weight |