AC WITHSTANDING |
Output-Voltage Range |
0.050kV~5.000kV |
Output-Voltage Resolution |
1V |
Output-Voltage Accuracy |
±(1% of setting + 5V) [no load] |
Maximum Rated Load |
200 VA (5kV/40mA) |
Maximum Rated Current |
40mA (0.5kV< V 5kV); 10mA (0.05kV V 0.5kV) |
Output-Voltage Waveform |
Sine wave |
Output-Voltage Frequency |
50 Hz / 60 Hz selectable |
Voltage Regulation |
±(1% + 5V) [maximum rated load -- no load] |
Voltmeter Accuracy |
±(1% of reading + 5V) |
Current Measurement Range |
1μA~40.00mA |
Current Best Resolution |
1μA / 10μA |
Current Measurement Accuracy |
±(1.5% of reading + 30 A) μ |
Window Comparator Method |
Yes |
ARC Detect |
Yes |
RAMP UP (Rise Time) |
0.1s~999.9s |
RAMP DOWN (Fall Time) |
0.0s~999.9s |
TIMER (Test Time) |
OFF, 0.3s~999.9s |
WAIT TIME |
0.0s~999.9s |
GND |
ON/OFF |
DC WITHSTANDING |
Output-Voltage Range |
0.050kV~6.000kV |
Output-Voltage Resolution |
1V |
Output-Voltage Accuracy |
±(1% of setting + 5V) [no load] |
Maximum Rated Load |
50W (5kV/10mA) |
Maximum Rated Current |
10mA (0.5kV < V ≦6kV); 2mA (0.05kV ≦ V ≦ 0.5kV) |
Voltage Regulation |
±(1% + 5V) [maximum rated load -- no load] |
Voltmeter Accuracy |
±(1% of reading + 5V) |
Current Measurement Range |
1µA~10.00mA |
Current Best Resolution |
0.1µA/1µA/10µA |
Current Measurement Accuracy |
±(1.5% of reading + 3µA) when I Reading < 1mA ; ±(1.5% of reading + 30µA) when I Reading ≧1mA |
Window Comparator Method |
Yes |
ARC Detect |
Yes |
RAMP UP (Rise Time) |
0.1s~999.9s |
RAMP DOWN (Fall Time) |
0.0s~999.9s |
TIMER (Test Time) |
OFF, 0.3s~999.9s |
WAIT TIME |
0.0s~999.9s |
GND |
ON/OFF |
INSULATION RESISTANCE |
Output Voltage |
50V~1200V dc |
Output-Voltage Resolution |
50V |
Output-Voltage Accuracy |
±(1% of setting + 5V) [no load] |
Resistance Measurement |
Test Voltage |
Display Range |
Measurement Range / Accuracy |
50V≦V≦100V |
0.1MΩ~10.00GΩ |
0.1MΩ~1MΩ : ±(5% of reading + 3 count); 1 MΩ~50MΩ : ±(5% of reading + 1 count); |
150V≦V≦450V |
0.1MΩ~20.00GΩ |
51MΩ~2GΩ : ±(10% of reading + 1 count) |
500V≦V≦1200V |
0.1MΩ~50.00GΩ |
0.1MΩ~1MΩ : ±(5% of reading + 3 count); 1 MΩ~500MΩ : ±(5% of reading + 1 count);
501MΩ~9.999GΩ : ±(10% of reading + 1 count); 10GΩ~50GΩ : ±5(20% of reading + 1 count) |
Voltage Regulation |
±(1% + 5V) [maximum rated load -- no load] |
Voltmeter Accuracy |
±(1% of reading + 5V) |
Short-Circuit Current |
10mA max. |
Output Impedance |
2kΩ |
Window Comparator Method |
Yes |
RAMP UP (Rise Time) |
0.1s~999.9s |
RAMP DOWN (Fall Time) |
0.0s~999.9s |
TIMER (Test Time) |
0.3s~999.9s |
WAIT TIME |
0.0s~999.9s |
GND |
ON/OFF |
CONTINUITY TEST |
Output-Current |
100mA dc (fixed) |
Ohmmeter Measurement Range |
0.10Ω ~ 70.00Ω |
Ohmmeter Measurement Resolution |
0.01Ω |
Ohmmeter Measurement Accuracy |
±(10% of reading + 2Ω) |
Window Comparator Method |
Yes |
TIMER (Test Time) |
0.3s~999.9s |
MEMORY |
Single Step Memory |
MANU: 100 blocks |
Automatic Testing Memory |
AUTO: 100 blocks, Manu per auto: 10 |
INTERFACE |
Standard (Front) |
REMOTE, USB host |
Standard (Rear) |
Rear Output, RS-232C, USB device, Signal I/O, |
Option |
GPIB |
DISPLAY |
7" color LCD |
POWER SOURCE |
AC 100V~240V ± 10%, 50Hz/60Hz; Power consumption : Max. 400VA |
DIMENSIONS & WEIGHT |
380(W) x 148(H) x 454(D) mm; Approx. 15kg |