IR4056-20
Economic model |
IR4057-20
Bar graph for visual judgments Economic model High-speed model |
|
|
Approx. 0.8 s |
Approx. 0.3 s |
|
|
- |
|
159W×177H×53D |
159W×177H×53D |
600 |
640 |
|
|
Rated output voltage (DC) |
Resistance measurement |
Comparator judgment result response time |
200 mA continuity |
Bar graph |
Dimensions(mm) |
Mass(g) |
Voltage measurement |
Rated output voltage (DC) |
50V |
125V |
250V |
500V |
1000V |
Effective maximum indicated value |
100 MΩ |
250 MΩ |
500 MΩ |
2000 MΩ |
4000 MΩ |
Effective medium value |
2 MΩ |
5 MΩ |
10 MΩ |
50 MΩ |
100 MΩ |
1st effective measuring range [MΩ] |
0.200 to 10.00 |
0.200 to 25.0 |
0.200 to 50.0 |
0.200 to 500 |
0.200 to 1000 |
Accuracy |
±4 % rdg |
2st effective measuring range [MΩ] |
10.1 to 100.0 |
25.1 to 250 |
50.1 to 500 |
501 to 2000 |
1010 to 4000 |
Accuracy |
±8 % rdg |
Ωther measuring range [MΩ] |
0 to 0.199 |
Accuracy |
±2 % rdg. ±6 dgt. |
Lower limit resistance value to maintain nominal output voltage |
0.05 MΩ |
0.125 MΩ |
0.25 MΩ |
0.5 MΩ |
1 MΩ |
Ωverload protection |
600 VAC (10 s) |
1200 VAC (10 s) |
Indicator |
Indicator: Semi-transmissive FSTN LCD, positive Backlight |
Functions |
Live circuit indicator, Automatic electric discharge, Automatic DC/AC detection, Comparator, Built-in battery power indicator etc. |
IR4057-20 Functions |
Bar graph, Displaying 1-min. Values |
Power source |
LR6 alkaline battery × 4 |
Continuous operating time |
Approx. 20 hours
(Comparator off, backlight off, 500 V range, no load) |
Auto Power Save |
The power will go off automatically 10 minutes after the last operation. |
Ωperating temperature and humidity |
0 to 40°C (32 to 104°F)
90% RH or lower (non-condensating) |
Storage temperature and humidity |
-10 to 50°C (14 to 122°F)
90% RH or lower (non-condensating) |
Maximum rated voltage to earth |
600 V AC/DC, Measurement Category,
Anticipated Transient Ωvervoltage: 6000 V |
Dielectric strength |
7060 V AC, 50/60 Hz, Measurement terminals - electrical enclosure, 1 min, current sensitivity 1 mA |
Degree of protection |
IP40 |
Standards |
EN61326 (EMC), EN61557-1/ -2/ -4*/ -10 |
Drop proof |
Ωn concrete: 1 m |
Dimensions |
Approx. 159W×177H×53D mm
(6.26"W×6.97"H×2.09"D) (excluding protrusions) |
Mass |
IR4056-20: Approx. 600g (21.2 oz)
IR4057-20: Approx. 640g (22.6 oz)
(including battery, excluding test lead) |
Accessories |
Test Lead L9787 × 1, Neck strap × 1, Instruction manual × 1, LR6 alkaline battery × 4 |
Display range
(Auto range) |
4.2 V |
42 V |
420 V |
600 V |
Maximum indicated value |
4.200 V |
42.00 V |
420.0 V |
750 V |
Resolution |
0.001 V |
0.01 V |
0.1 V |
1 V |
Accuracy |
±1.3 % rdg. ±4 dgt. * |
Display range
(Auto range) |
420 V
(Minimum indicated value: 30.0 V) |
600 V |
Maximum indicated value |
420.0 V |
750 V |
Resolution |
0.1 V |
1 V |
Accuracy |
±2.3 % rdg. ±8 dgt. * |
Measurement Principles |
Average responding type |
Frequency range |
50/60 Hz |
AC/DC automatic detection range |
AC detected at 30 V or greater (50/60 Hz) (Pulsating currents with an overlapping AC component of 30 V or greater are detected as AC) |
Effect of temperature |
Measurement accuracy per 1°C × 0.1 (Applicable to the operating temperature range other than 18 to 28°C) |
Display range
(Auto range) |
10 Ω |
100 Ω |
1000 Ω |
Maximum indicated value |
10.00 W |
100.0 W |
1000 W |
Resolution |
0.01 W |
0.1 W |
1 W |
Accuracy
(after zero adjustment) |
0 to 0.19 W: ±3dgt.
0.20 to 10.00 Ω: ±3%rdg. ±2dgt. |
±3%rdg. |
±2dgt. |
Measuring current |
200 mA or more (at 6 Ω or less )
(Display value before zero adjustment) |
Ωverload protection |
600 VAC (10s, using Fuse) |