Specifications |
Guaranteed accuracy period |
1 year |
Number of input channels |
2 channels |
Input terminals |
R05-R5F by TAJIMI |
Appropriate adapter |
Strain gauge adapter |
Bridge voltage |
2, 5, 10 ± 0.05 V |
Bridge resistance |
120 Ω to 1 kΩ (Bridge voltage 2 V)
350 Ω to 1 kΩ (Bridge voltage 5, 10 V) |
Balance adjustment range |
±10000 με max |
Balancing |
Electronic auto-balancing |
Measurement ranges |
20, 50, 100, 200, 500, 1000 με/div |
DC amplitude accuracy |
±(0.4%f.s.+ 2 με) (filter 5 Hz ON)
(When using the conversion cables supplied.
When using other cables, cable resistance is 1.5 Ω max. at bridge resistance 350 Ω) |
Zero position accuracy |
±(0.1%f.s.+ 2 με) (filter 5 Hz ON, after auto-balancing) |
Temperature characteristic |
Gain: ±0.05%f.s./°C Zero position (after auto-balancing):±2.5 με/°C |
Frequency characteristic |
DC to 20 kHz dB |
Low-pass filter |
OFF, 5±30%, 10±30%, 100±30%, 1k±30% (Hz) -3 dB |
Anti-aliasing filter |
Cutoff frequency (fc) 20, 40, 80, 200, 400, 800,
2 k, 4 k, 8 k, 20 k, 40 k (Hz) (These frequencies
are automatically set when the anti-aliasing filter is set to ON)
Attenuation; -66 dB min at 1.5 fc A/D resolution 16 bits |
Maximum sampling rate |
200 kS/s |
Maximum input voltage |
10 VDC |
Maximum rated voltage to earth |
33 Vrms or 70 VDC (between each input channel
and main unit, and between input channels) |
Dielectric strength |
350 VAC for 15 seconds (between input module
and main unit, and between input modules) |
Operating temperature and humidity ranges |
Same as the Memory HiCorder in which the 8960 is installed |
Operating environment |
Same as the Memory HiCorder in which the 8960 is installed |
Storage temperature and humidity ranges |
Same as the Memory HiCorder in which the 8960 is installed |