Measurement Functions |
Number of channels: 2, for voltage measurement, DC/RMS selectable |
Input Connectors |
Isolated BNC connector (input impedance 1 MΩ, input capacitance 30 pF)
Max. rated voltage to earth: 370 V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) |
Measurement Range |
5 mV to 20 V/division, 12 ranges, full scale: 20 divisions, AC voltage for possible measurement/display using the memory function: 280 V rms, low-pass filter: 5/500/5k/100k Hz |
Measurement Resolution |
1/100 of measurement range (using 12-bit A/D conversion; installed in 8860 series) |
Highest Sampling Rate |
1 MS/s (simultaneous sampling in 2 channels) |
Accuracy |
DC amplitude: ±0.4% of full scale (with filter 5 Hz)
zero position: ±0.1% of full scale (with filter 5 Hz, after zero adjustment) |
RMS Measurement |
RMS amplitude accuracy: ±1% of full scale (DC, 20 Hz - 1 kHz), ±3% of full scale (1 kHz - 100 kHz),
response time: SLOW 5 s (rise time from 0 to 90% of full scale), MID 800 ms (rise time from 0 to 90% of full scale), FAST 100 ms (rise time from 0 to 90% of full scale),
crest factor: 2 |
Frequency Characteristics |
DC to 400 kHz ±3 dB, with AC coupling: 7 Hz to 400 kHz ±3 dB |
Max. Allowable Input |
400 V DC (the maximum voltage that can be applied across input pins without damage) |
Dimensions and Mass |
approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm, approx. 290 g (10.2 oz) |