Measurement functions | Number of channels: 1 voltage or temperature measurement channel |
Input | Voltage input: isolated BNC connector, thermocouple input: Plug-in terminal, Max. rated voltage to earth: 370 V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channels and chassis, or across input pins, without causing damage) |
Voltage Measurement range | 500 ?V to 2 V/division, 12 settings, full-scale = 20 divisions, low-pass filter: 1/5/50/500/20 kHz, the measurement resolution: 1/1600 of range (using 16-bit AD) |
Thermocouple range | K: -200 to 1350 ?C, E: -200 to 800 ?C, J: -200 to 1100 ?C, T: -200 to 400 ?C, N: -200 to 1300 ?C, R: 0 to 1700 ?C, S: 0 to 1700 ?C, B: 300 to 1800 ?C, W: 0 to 2000 ?C Reference junction compensation: internal/external (interchangeable) |
Maximum sampling rate | Voltage input: 100 kS/s, temperature measurement: 4 kS/s (data updates every 250 ms) |
Accuracy | Voltage input: DC amplitude: ?0.2% f.s. Zero-position: ?0.2% f.s. Temperature measurement (K, E, J, T, N): ?0.1% f.s. ?1 ?C, ?0.1% f.s. ?2 ?C (-200 ?C to 0 ?C), (R, S, W): ?0.1% f.s. ?3 ?C, (B): ?0.1% f.s. ?4 ?C (400 to 1800 ?C) Reference contact compensation accuracy: ?0.1% f.s. ?1.5 ?C (for reference junction internal compensation) |
Frequency characteristics | Voltage input: DC to 20 kHz +1/-3 dB Temperature measurement: DC to 1 kHz +1/-3 dB |
Input resistance and capacitance | Voltage input: 1 M-ohm, 60 pF (when C is 10 kHz) Temperature measurement: 4.8 M-ohm or more |
Max. Allowable Input | 30 V rms or 60 V DC (the maximum voltage that can be applied between input pins without causing damage) |
Accessories | None * The input cord and conversion cable are optional |